MICROSOUND SYSTEMS NURSE CALL eight channels of music & ward TV input

Similar documents
IPiN SYSTEM DESCRIPTION.

CERTIFICATES OF ELECTRICAL SAFETY RESPONSIBILITY OF RECs AND ELECTRICIANS

CAP IP. ip intercom REFUGE CALL POINTS IP NETWORK SALES DEPARTMENT SECURITY DISPATCH SMARTPHONE SUPERVISION IP MAYLIS RANGE

SECTION NURSE CALL/CODE BLUE SYSTEM (JERON)

GIVES YOU A CHOICE OF NURSE CALL.

Siemens TEC Unit Vent 0-10V Output Controller. Accessories. Control Applications. Product Description. Product Numbers. Warning/Caution Notation

Element D Services Electrical

E-BOOK // DOUBLE SWITCH DIAGRAM LIGHTS

Parts List. Commercial Series Refrigerated Kitchen Equipment

Easylink to Square D Module

Item No. Description Part No.

Nurse Call System. A Voice over IP Based Solution for Streamlined Communication, Alerting and Workflow

TABLE OF CONTENTS FINAL ASSEMBLY (3 WIRE)... 1 FINAL ASSEMBLY (2 WIRE)... 2 DOOR ASSEMBLY (3 WIRE)... 4 DOOR ASSEMBLY (2 WIRE)...

ARCHITECTS' AND ENGINEERS' SPECIFICATIONS FOR NC300 MICROPROCESSOR NURSE CALL SYSTEM

PC-1600 Mech 230/460v PC v BL PC v BL PC-1300 Mech 230/460v

WIRING DIAGRAM KITCHEN OUTLETS

[FILE] SERIES OUTLET WIRING DIAGRAM EBOOK

Full IP. nursecall and notification

Continuing Education Credits

Clinical Pendants. starkstrom. Enhancing Safety at the Bedside. Powering Healthcare. a Progility company

User Manual Lobby Camera Phone

EASYLINE SYSTEM OVERVIEW

Easylink to Liebert Sitelink Module

700R4 LOCKUP WIRING TOGGLE SWITCH DOCUMENT

CHUK Hospital Statistics 1

ELECTRICAL INSPECTION ACT CANADIAN ELECTRICAL CODE REGULATIONS

UNIT Music: Technical Support (SCQF level 6)

Element D Services Electrical

Specifications, Performance & Submittal

Cabin Crew B787 SOP Cabin Layout Aircraft Configuration Special Seating Requirements Cabin Crew Stations Interphone Dimmable Window

Telligence Nurse Call Communication System. Total solutions for patient-staff communications

NCS Nurse Call System. Bridging Communications & Care Delivery Services

A. General provisions and other electrical systems are specified in other Sections of Division 26.

Smoke Control BACnet PTEC Unit Vent Controller. Control Applications. Product Description. Product Numbers

ATTACHMENT B. 1. Intent to Bid/Bidder Contact Information Form. 2. Executive Summary. 3. Proposal Characteristics and Term Sheet

medical and commercial f ields.

Isolation Precaution (Part 2) Protective Environment (PE) Room. Combined AII/PE Rooms. Contact Isolation 5/22/2017

Feel at home. No matter where you stay. TECHNICAL MANUAL

PATIENT SATISFACTION: THE LITTLE THINGS ADD UP

ENTERPRISE with people in mind

Introducing the. Yesterday s challenge Today s solution. innova.uk.com

B. All of the documents in this RFP form the Basis-of-Design for this project.

Medical Grade Power Strips

AIRCRAFT WEAPONS SYSTEMS TEST EQUIPMENT

TELECOMMUTING AGREEMENT

LESTER SENIOR HOUSING JOB DESCRIPTION ANNUAL EVALUATION AND COMPETENCY Concierge

MAXI SKY 2 THE NEXT GENERATION CEILING LIFT SOLUTION

6-3 [D5A0] WIRING DIAGRAM. 5. Wiring Diagram. 5. Wiring Diagram A: POWER SUPPLY ROUTING

Microprocessor Nurse Call System In-Service Training

Leaders' Guide 2018 MERIT BADGE MIDWAY. Firelands Scout Reservation: May 4-6, 2018 Lorain County Joint Vocational School: May 5, 2018

The University of Texas at San Antonio

Kimberly Shea, Administrator, Technical Projects. SUBJECT: NFPA 70 Second Draft TC Ballot Final Results (A2016)

Rulemaking Hearing Rules of Department of Commerce and Insurance Division of Fire Prevention. Chapter Electrical Installations.

DUTIES OF TM, TTA & SR TOA S AS PER DOT NO.29-1/96 TE II DATED

Introduction. Welcome to Human Care.

OPERATOR'S, ORGANIZATIONAL, DIRECT SUPPORT, AND GENERAL SUPPORT MAINTENANCE MANUAL FOR MULTIPLEXER SET AN/FCC-97 (NSN )

Designed and manufactured in the UK - specified and installed throughout the world.

Overview. Airespace AP a Radio Card Quick Installation

Audio. Distribution. Network ADN C1/ADN D1. Quick guide

TELLIGENCE. Ascom. Nurse Call System. The complete patient-staff communication solution. Ascom

Indications for Calling A Code Blue or Pediatric Medical Emergency

INTERNAL FRENCHFOLD DOOR SYSTEM

Le Moyne College. BEICKERT AUDIO 219 Arnold Ave, Syracuse, NY Thomas J. Niland Jr. Athletic Complex SOUND STUDY QUOTATION.

LESSON ASSIGNMENT. Environmental Health and the Practical Nurse. After completing this lesson, you should be able to:

More comfortable childbirth

Declaration of Classification ValiSpore Incubator Manual

Comprehensive Staff and Patient Communications: Emphasizing Patient-Centered Care

User-Service Manual Joerns Assist Handle Model F025

UK Universities Student Sensor Competition

MediCom. MediCom AUSTCO. State of the art Hospital Emergency Call System. The very best in nurse call and. care management solutions

Guidelines on Postanaesthetic Recovery Care

Physical Security Workshop TRENDS ACROSS MEMBER UNITS. Dave Hearn

P A N E L B O A R D S C H E D U L E

Integrated Nurse Call Solutions

Isolette TI500 Neonatal Transport

Rauland Responder 4000

Spare Part Catalogue. Star Cool refrigeration unit Model SC MCI40, SCU and SCI

Recommended Physical Plant Improvements to Existing Nursing Homes for Disaster Preparedness. Report to the Governor and to the Legislature

TECHNICAL MANUAL OPERATOR'S, UNIT, DIRECT SUPPORT, AND GENERAL SUPPORT MAINTENANCE MANUAL

(800) Copyright TekTone Sound & Signal Mfg., Inc. All Rights Reserved.

Safe & Sound Video and audio baby monitor User guide Models: BM3000/ BM3000 Twin

econsult econsult Introduction & Training Manual

CDRL A006 Training Manual User's Guide for STAT! TM EMEDS ICU Serious Medical Game. Release v November 26, 2014

Caution! triangles are used to warn of situations that demand extra care and attention.

ISABELLA COUNTY REQUEST FOR PROPOSALS COMMISSION ON AGING CATV AND HEADEND EQUIPMENT

CARESCAPE Central Station Powerful insight. Streamlined design.

Giving Intravenous (IV) Nutrition Through a Central Line with a CADD Pump

EZTest Incubator OPERATION MANUAL. 10 Evergreen Drive Bozeman, MT USA (303) Model

SAN DIEGO UNIFIED SCHOOL DISTRICT. Site Emergency Response Plan. La Jolla Elementary Marine St., PUBLIC DOCUMENT

Building elder friendly elements into acute hospital care a pilot project

SAN DIEGO UNIFIED SCHOOL DISTRICT. Site Emergency Response Plan. Pacific Beach Elementary Tourmaline St., PUBLIC DOCUMENT

Choosing Regulatory-Compliant Power Strips for Every Area of a Healthcare Facility

RFP for 2015 E-Rate Funded Wireless Upgrade Project at Four Locations In Thomasville City Schools RFP Reference# 2015Wireless Cabling Contact Mike

Eleganza 3XC. Designed for ICU

REQUEST FOR PROPOSAL INSTRUCTOR AUDIO-VISUAL PODIUMS FOR ILLINOIS VALLEY COMMUNITY COLLEGE PROPOSAL #RFP2013-P06

1510 Self-Contained Biological Indicator Incubator OPERATION MANUAL

Thank you for purchasing the Actuator Systems NextBolt fingerprint deadbolt.

L1-CHE-SPE-305 SIGNAL IDENTIFICATION AND GENERAL INFORMATION

OSHPD 3 CLINIC CERTIFICATION INSTRUCTIONS

MEMORANDUM. According to the final ballot results, all ballot items received the necessary affirmative votes to pass ballot.

Transcription:

MICROSOUND SYSTEMS NURSE CALL eight channels of music & ward TV input MANUAL

Contents 1) Nurse Call Equipment...3 a. General...3 1) Central Equipment...3 a. Nurses Station Equipment...3 b. Bed Head Unit...3 i) Functions on all bed head unit...3 2) Installation Requirements...4 a. Central Music Cabinet...4 1) Multiplexer....4 a. Power Supply...4 b. Lamp Panel...4 c. Bed Head Unit with Music connections...4 d. T.V. Sets...5 e. Hand held unit with music HO5...5 f. Hand held unit with remote TV control HO7...5 g. Reading Light...5 h. Power requirements...5 2) Planning the Installation...6 a. Wards...6 1) Nurses Stations...6 2) Cable Layout...7 a. Wards...7 Terminations...7 At the Bed Head Unit...7 At the Ward Distribution Box...8 At the Nurses Station...8 Intercom Terminations...8 Labeling...8 Radio Distribution...8 Co-ax Cable Network...8 Chimes and Alarms....8 Guarantee...9 Nurse Call with TV.doc Page 2 of 9 29-Oct-02

1) Nurse Call Equipment a. General The Microsound range of nurse call equipment has been designed to cover all the requirements for Hospitals, Clinics, Old Age Homes and Retirement Villages. The systems provide the facilities for patients to call nurses for assistance plus the distribution of music to the patients. Various units cover requirements for different areas in a hospital. There are units for beds in general and private wards, children s wards, doctors and nurses rest rooms, nurse's stations and toilets. Mechanically the units are designed to clip in to 100 x 100 mm or 100 x 50 mm outlet boxes or flush mount on ducting systems. Mechanically the units are designed to clip into the hospital ducting systems. The Nurse Call system can be coupled to the Microsound SLSI or M20 intercom systems to provide audio communications between the nurse's station and patients bed unit, theatres, doctors, nurse's rest rooms and I.C.U. The range has been designed to cover as many different requirements as possible so as to provide an off the shelf, reliable, high quality product. The product is locally designed, manufactured and serviced. 1) Central Equipment The only equipment required is the Music Cabinet, which consists of radio tuners, a tape deck (if required) and monitor speaker. The radio tuners are connected to a FM/AM aerial mounted externally. The music cabinet distributes all the required music channels over twisted pairs of wire to the different sections or areas of the hospital. The music cabinet can distribute from four to eight channels of music to the whole hospital complex. a. Nurses Station Equipment There are three main units that are situated at the nurse's station, the power supply, multiplex and the indication panel. The power supply supplies D.C. to all bed units, toilets units, over door lights and indication panel in the wards associated with the particular nurses station. The power supply requires a 220v AC 15 amp outlet. The multiplexer combines the individual music channels from the central music cabinet, and distributes the music on a common 75 ohm coax cable which is looped though each bed unit. The multiplexer has a maximum of 80 db input channels and two co-ax outputs. Each output can drive a maximum of 20 bed units to a maximum distance of 200 meters. A system that has no music does not need a multiplexer. The Indicator Panel consists of a number of a number of LED s to indicate from which bed/ward a call has be initiated plus a audible tone to draw the attention of the duty sister to the call. The visual, audible indication can only be cancelled from the point where the call has been made. The indication panel can also include a crash call intercom point, whereby the sister can call for assistance to a central point in the hospital. If the system has audio communication to each bed the nurse station will require a separate master station to select any required bed. b. Bed Head Unit i) Functions on all bed head unit 1. Call/Emergency button: when pushed it initiates a call to the nurses station, lights the over door lamp and lights the reassurance lamp on the unit. Nurse Call with TV.doc Page 3 of 9 29-Oct-02

2. Cancel this button is pressed by the nurse to reset the call. 3. Pearpush socket: A pearpush (1,5 metre cable) with a remote call button can be plugged into it. This allows the patient to initiate a call by pressing the remote call button. 2) Installation Requirements a. Central Music Cabinet. A 220v 15-amp mains socket outlet is required. External FM/AM aerial input. A Co-ax cable must be installed between an external aerial and Music Cabinet. One pair 0,5 mm screened telephone cable pair per music channel to each multiplexer in the hospital. The music cable must be earthed at music cabinet the screen must extend to all multiplexers. Up to eight channels of music are available. 1) Multiplexer. Requires 24v DC regulated from P.S.U. Eight music input pairs from central cabinet. Provides two 75 ohm co-ax output. (Each end must be terminated). The music input cable must not be looped (jointed) in multiplex but in a separate DP Box spare multiplexer music input must not be left unused but connected to used inputs. The switch on the multiplexer is to switch off the music to the beds in the event of a problem in the music system. This should be pointed out to Hospital Maintenance. Each output must supply only half the beds in a ward, i.e. each output must be equally loaded. The Co-ax screen and center must be completely isolated from the building electrical earth, ducting or anything else. Good quality double screened 75 ohm flexible co-ax must be used, the co-ax must be connected into the BHU, loose wires must not be used to interconnect between the Co-ax and the BHU. All joints in the duct must be insulated. The Co-ax must loop from the multiplexer to the termination at the end via all BHU, no T off s can be made. The co-ax must run the shortest distance possible, where wards are back to back allow for a conduit connection between the ducts at the far end. The Multiplexer must have through flow ventilation for cooling. a. Power Supply. A 220vac 15 amp mains socket outlet supplied by a separate circuit breaker is required. Supplies 24v via red and black power cable to all bed, toilet, bathroom, rest rooms and overdoor units. The P.S.U's are 100va maximum, but must only be loaded to 50va continuously. Where more than one P.S.U. is used the positive must not be connected to each other, the negative must be connected via the mains earth. The P.S.U. needs through flow ventilation for cooling. Regulated P.S.U. must be used for systems with music. b. Lamp Panel A 24vdc power supply is required. The lamp panel indicates calls from the wards, individual beds, toilets and bathrooms. If equipment has crash call intercom system, one twisted pair is required to intercom central. Wall or desktop mountable. c. Bed Head Unit with Music connections A Molex crimping tool must be used to crimp wires into the molex sockets, the conductor and the isolation must be crimped (ribbon cable must not be used) Pin 1 P/S Positive 24v Dc input (red) Pin 2 P/S GND input (black) Pin 3 + 4 0/D (zone over door) input (1 core 0,5mm telephone cable) common to all beds in one ward. Nurse Call with TV.doc Page 4 of 9 29-Oct-02

Pin 5 L/P (lamp panel) input (1 core 0,5mm telephone cable) from each bed in a zone/ward wired back to indication panel. Pin 6 EME (Optional) ALL BHU in a zone are communed up via the input to give a EMERGENCY call priority over a normal nurse call. (Only used if the system has emergency call). Pin 7 reading lamp relay input TV. GND. COAX. GND. TV audio from ward TV transformer TV transformer ground Music co-ax center Co-ax ground Bed unit with TV control connections Pins 1 to 7 same as above. Blue wire. Yellow wire. TV audio from bed TV transformer. TV audio from bed TV transformer. d. T.V. Sets The ward or bed TV audio must be isolated from the BHU with a double wound transformer mounted at the TV. e. Hand held unit with music HO5 The HHU is wired directly to the BHU. It has hook to hook onto the duct, a speaker for music and TV sound and 5-button keypad for the following controls. 1. Nurse Call button with reassurance LED. 2. Reading light 3. Radio selection button for 8 channels and ward TV sound 4. Volume up 5. Volume down and off. f. Hand held unit with remote TV control HO7 The unit is wired directly to the BHU. It has a hook to hook onto the duct, a speaker for the TV sound and 7 button keypad for the following controls. 1. Nurse call button with reassurance LED. 2. Reading light. 3. TV on off 4. Program up 5. Program down 6. Volume up 7. Volume down The HHU must be factory programmed for a specific make of TV. g. Reading Light The HHU operates the reading light via a separate latching relay PCB. Not directly. h. Power requirements All measured at 24v DC. Nurse Call with TV.doc Page 5 of 9 29-Oct-02

(a) Music Unit Nominally -65ma Max -75 ma With call active 125 ma (b) Call Cancel & Toilet Units Nominally - 30 ma With call active -80 ma (c) O/D Units when on - 125 ma (d) Indication Panel No lamps on - 50 ma With one LED on - 70 ma 2) Planning the Installation a. Wards In most cases ducting is used to bring cables into the room and from the ducting a conduit makes the connection to the bed head trunking. There might be 2 rows of beds in which case 2 or more 25mm conduits must connect the overhead ducting with each bed head trunking. In some convenient place, near the main arterial ducting, place must be found for the Ward Distribution Box This is a 300 x 300mm by 50mm deep electrical box (bigger if more than 6 beds). Cables, from each piece of equipment in that ward, must go directly to the Ward Distribution Box via a dedicated conduit or via the common ducting/trunking system. These units include all the following with recommended conduit sizes in brackets: - Bed Head Units Bathroom/Toilets Units Overdoor Lights 1 x 25mm for each bed 1 x 20mm for each unit 1 x 20mm for each unit The Ward Distribution Box must also be connected to cables going to the nurses station distribution board, so connection to the corridor ducts must also be possible again via a conduit or via the common ducting system. At least a 25mm must be used here. The most economical position for this Ward Distribution Box is somewhere in the ceiling next to or on the ward ducting near the arterial ducts, near the bathroom/overdoor lights. The most convenient position is at +- 1600mm AFL in the all outside in the passage or inside the ward near the door. 1) Nurses Stations It is difficult to give precise instruction for this area because the design and the aesthetics of this area vary so much, however thought must be given to the type of equipment to be installed here. Whatever type of system is to be installed, a central wire connection box must be provided for, preferably about 800 x 600mm wide x 250mm deep mounted on or in a wall at 450mm AFL in a position close to the desk of the station. Vertical ducts 50mm x 50mm should connect the this distribution board with the corridor ducts in the passages. A 50mm duct should connect the distribution box to the position of the lamp panel/intercom unit. This could be scaled down depending upon the size of the unit and the number of cables required. It might be possible to mount the control equipment in the distribution cabinet. If so, a power point is required in the cabinet. If not then some other space must be allocated to house thes equipment with a conduit connection to the distribution cabinet. Equipment that could be fitted in this area is as follows: - 1. Power Supply Unit. 2. Music Multiplexer. Nurse Call with TV.doc Page 6 of 9 29-Oct-02

3. Intercom Controller. 4. Intercom Control Panel. 5. Mimic Panel. 6. Chimes Unit. 7. Day Night Switch. 8. Distribution Panel. 2) Cable Layout a. Wards i) Bed Head Unit. Only one cable per bed should be installed from the Ward Distribution Box to the Bed Head Unit. This cable must have 6 flexible cores of.22mm or 5mm for systems without intercom and 8 cores if intercom is required. In addition the music co-ax cable must pass each bed and a SHORT T in connection must be made using a piece of thin 75E co-ax. The place where this connection is made must be insulated using tape or some large sleeving. It is important that no part of the co-ax is in contact with the trunking. The connection must be of the soldered type. 1) Bathroom/Toilet Unit. Only one cable of 5 cores is necessary to connect this unit to the Ward Distribution Box. Corridor. 2) Overdoor Lights. A single pair (ripcord) is needed back to the Ward Distribution Box. Separate cables must be used for: 1. Power 2.5mm Pair 2. Nurse Call 1 x 10 pair cable for 2 x 5 bed wards 3. Intercom 1 x 10 pair cable for 2 x 5 bed wards 4. Music 75E co-ax cable RG 59 It is our suggestion that only 10 pair cables be used here to simplify connections at the distribution Box at the nurse's station. Terminations At the Bed Head Unit The cable should arrive at the bed head unit directly from the Ward Distribution Box. The cables for the nurse call and power must be connected to the connecting plug for nurse call, while the intercom wires must be connected to the separate plug in connector on the Interco board if any. In addition the co-ax connection for the Radio must also be connected to the connecting plug in the following way. As the main co-ax passes the bed position the cable must be cut and rejoined with a soldered connection and a short length of thin 2.5mm flexible co-ax cable must be added to this connection before the joint is covered with tape or sleeving. The other end of the thin co-ax must then be connected to the correct pins of the in-line connector mentioned above. Nurse Call with TV.doc Page 7 of 9 29-Oct-02

At the Ward Distribution Box All cables are terminated here on screw terminals. The cables can be screwed directly into the terminal blocks or preferably the individual wires will be fitted with crimped lugs and then the lugs will connect to the terminal blocks. The only cable, which is not connected via the Distribution box, will be the Radio co-ax cable. At the Nurses Station All the cables that need termination at the Nurses position should arrive at the Nurses Termination Box where all the wires must be terminated including all the spare wires. The terminations must be done on 10 way Krone strips in such a way that all the wires of one 10 pair cable are connected, in order, on a single strip. Stranded and solid core cable can be connected via the Krone strip. Intercom Terminations Intercom connections must be made on a separate group of Krone strips. Interconnect wires must be introduced to connect the intercom connections to the field cable which, as mentioned above, should be terminated on its own strip elsewhere. Labeling All cables arriving at the Nurses Station must be clearly labeled so that: 1. Installation can proceed without confusion. 2. The service engineer that has to fix the system in 5 years time will have no difficulty in understanding which cable goes where. Radio Distribution Co-ax Cable Network The Co-ax cable which carries up to 8 channels of music must be routed past every Bed Head Unit. This cable should be kept as short as possible to reduce losses in the cable. Where there is a possibility of passing the co-ax cable through the wall from one bed head trunking to the next this is preferable compared to doubling up on the cable to take it out of the ward where it came in. The routing of the co-ax cable must be planned in such a way that optimum use is made of both outputs of the Multiplexer. Both legs should be equal in length if possible although this requirement is not essential. Radio Cabinet and distribution to the Multiplexer. The Radio Cabinet should be installed a some convenient place in charge of some responsible person. The cable from the Radio cabinet to the multiplexer could be routed directly between the 2 units or be taken to some central point before distribution to several multiplexers. Chimes and Alarms. The box containing the Chime and Alarm unit must be positioned so that it is near the Lamp indicator panels and at the same time it should not be too far from the Nurses Station Distribution Box. It has to be mounted in such a way that the sound is not too muffled. Be sure to order the Mimic Panel with sufficient cable to reach from the correct position to the Nurses Station Distribution box. The cable should be terminated directly onto the field cable Krone strip. Nurse Call with TV.doc Page 8 of 9 29-Oct-02

Drawings for: 1. Production of the Mimic Panel. 2. Positioning and 3. Housing of the Mimic/Lamp panels Must be obtained from the consultant as soon as possible after receiving the order so that these items do not hold up the installation. Guarantee Equipment is guaranteed by the manufacture for a period of six months from the date of delivery against faulty components and workmanship (ex Factory). Manufactured by: Microsound Systems P.O. Box 38220 Booysens South Africa 2016 Phone 011 493-2424 Fax # 011 493-0347 Web: http://www.microsound.co.za Email: sales@microsound.co.za Nurse Call with TV.doc Page 9 of 9 29-Oct-02