A Team-Based Approach to Reducing Cardiac Monitor Alarms

Similar documents
Alarm Fatigue In The Emergency Room Setting

Leveraging Technology to Reduce Inactionable Alarms from Bedside Physiologic Monitors

Addressing the Problem of Alarm Fatigue: Enhancing Patient Safety through Cardiac Alarm Customization

Using Quality Improvement to Optimize Pediatric Discharge Efficiency

New Monitoring Alarm Challenges and Opportunities for Collaborative Progress

Follow this and additional works at: Part of the Nursing Commons

The Problem of Alarm Fatigue

Addressing vital sign alarm fatigue using personalized alarm thresholds*

Using Quality Improvement to Introduce and Standardize the National Early Warning Score (NEWS) for Adult Inpatients at a Children s Hospital

ASCO s Quality Training Program

Keywords: Traditional Medical Monitoring, Questionnaire, Weighted Average, Remote Medical Monitoring, Vital Signs.

Performance-Based Assessment of Radiology Practitioners: Promoting Improvement in Accordance with the 2007 Joint Commission Standards

Decreasing Time to Antibiotics in Febrile Patients With Central Lines in the Emergency Department

Acute Care Workflow Solutions

Early Recognition of In-Hospital Patient Deterioration Outside of The Intensive Care Unit: The Case For Continuous Monitoring

Keep watch and intervene early

The Institute of Medicine concluded that medical

Matching Capacity and Demand:

Using Quality Improvement to Reduce Continuous Pulse Oximetry Use in Children With Wheezing

AUSTRALIAN AND NEW ZEALAND COLLEGE OF ANAESTHETISTS ABN RECOMMENDATIONS ON MONITORING DURING ANAESTHESIA

Place hospital logo here

Alarm Safety in a Regional Neonatal Intensive Care Unit

Monday, August 15, :00 p.m. Eastern

The Effect of an Electronic SBAR Communication Tool on Documentation of Acute Events in the Pediatric Intensive Care Unit

abstract QUALITY REPORT

Development of a Policy and Procedure to Decrease Alarm Fatigue

SITE APPLICABILITY This practice applies to all pediatric patient care areas that have been designated by your health authority.

Description and Evaluation of an Educational Intervention on Health Care Costs and Value

SPC Case Studies Answers

Alarm Fatigue: A Risk Assessment

Patients Being Weaned From the Ventilator: Positive Effects of Guided Imagery. Authors McVay, Frank; Spiva, Elizabeth; Hart, Patricia L.

Using Predictive Analytics to Improve Sepsis Outcomes 4/23/2014

ASCO s Quality Training Program

Safety Innovations FOUNDATIONHTSI. Clinical Practice Changes Associated with Alarm Standardization. The Boston Medical Center Experience

Alarm Management: From Confusion, to Information, to Wisdom. August 3, 2015

Digistat Patient Watch

Policy and Procedures. RNSP: RN Procedure. I.D. Number: 1142

National Early Warning Score (ViEWS) System. Recommendations for Audit. February 2012

1. Create a heightened awareness of clinical risks and enterprise-wide challenges associated with misuse of copy and paste.

Lessons from Chicago

Are National Indicators Useful for Improvement Work? Exercises & Worksheets

SARASOTA MEMORIAL HOSPITAL NURSING DEPARTMENT POLICY

EMR Surveillance Intervenes to Reduce Risk Adjusted Mortality March 2, 2016 Katherine Walsh, MS, DrPH, RN, NEA-BC Vice President of Operations,

Medicare: This subset aligns with the requirements defined by CMS and is for the review of Medicare and Medicare Advantage beneficiaries

Using Data to Inform Quality Improvement

SARASOTA MEMORIAL HOSPITAL STANDARDS OF CARE STANDARDS OF PRACTICE CARDIAC ACUTE CARE AND CARDIAC PROGRESSIVE UNITS

Scaling Up and Validating a Nursing Acuity Tool to Ensure Synergy in Pediatric Critical Care

Approximately 180,000 patients die annually in the

Advanced Measurement for Improvement Prework

The Effects of an Electronic Hourly Rounding Tool on Nurses Steps

Reducing Waiting-time of Preterm Babies at a Retinopathy of Prematurity Clinic: A Quality Improvement Project

Guidelines for Pediatric Cardiology Diagnostic and Treatment Centers

We See You When You're Sleeping

Research Design: Other Examples. Lynda Burton, ScD Johns Hopkins University

Robert J. Welsh, MD Vice Chief of Surgical Services for Patient Safety, Quality, and Outcomes Chief of Thoracic Surgery William Beaumont Hospital

Patient-Centered Fall Prevention Toolkit Paper Fall TIPS Instruction Sheet for Nurses

Shalmon SC 1 (Department of Nursing, BLDEA s Shri BM Patil institute of Nursing science, Bijapur/ Rajiv Gandhi university of Health sciences, India)

Improving Nursing Workflow Efficiency & Nurses Knowledge & Attitude Toward Computers. WellStar Health System. Background

Title & Subtitle can. accc-cancer.org March April 2017 OI

Saving Lives: EWS & CODE SEPSIS. Kim McDonough RN and Margaret Currie-Coyoy MBA Last Revision: August 2013

Department of Anesthesiology and Pediatrics, Duke University School of Medicine, Durham, NC, USA

Ruchika D. Husa, MD, MS

of the respiratory checklist from July1, April 30, Measures were evaluated monthly. Primary measures:

Basic Skills for CAH Quality Managers

RUNNING HEAD: HANDOVER 1

CARESCAPE Central Station Powerful insight. Streamlined design.

TASCS 2017 Annual Conference 3/2/2017

Care of Patients Receiving Analgesia by Catheter Techniques Position Statement and Policy Considerations

Use of Monitor Watchers in Hospitals: Characteristics, Training, and Practices

On the CUSP: Stop BSI

UNIVERSITY OF PENNSYLVANIA HEALTH SYSTEM

Objectives. Key Elements. ICAHN Targeted Focus Areas: Staff Competency and Education Quality Processes and Risk Management 5/20/2014

Financial Disclosure. Learning Objectives. Reducing GI Surgery Re-Admissions, While Increasing Patient Satisfaction

Effectiveness of Video Assisted Teaching Regarding Knowledge and Practice of Intra-Venous Cannulation for Under-five Children

October 2, 2015 Page 1

Theradex Audit 2013: Findings & Corrective Action

Changes in practice and organisation surrounding blood transfusion in NHS trusts in England

Clinical Safety & Effectiveness Cohort # 8

ALARM MANAGEMENT ON AN INPATIENT SURGICAL UNIT. Rhyana Rose Whiteley

X Series Extensive Capabilities

The Solution to Medical Device Security Also Could Save Tens of Thousands of Lives and Millions of Dollars

Recognising a Deteriorating Patient. Study guide

Writing Manuscripts About Quality Improvement: SQUIRE 2.0 and Beyond

GE Healthcare. B40 Patient Monitor Connecting intelligence and care

The Impact of CPOE and CDS on the Medication Use Process and Pharmacist Workflow

Tools & Resources for QI Success

Application of QSEN competencies in a practice setting: The case of Cincinnati Children s Hospital Medical Center

December 16, Thoracostomy Tube Removal Procedural Pain Practice Guideline Implementation Lisa M. Ring, DNP, CPNP, AC-PC

TWH ED ACUTE & SUBACUTE BEDS UTILIZATION PROJECT

Improving Patient Surveillance: Instituting a Respiratory Risk Screening Tool

Experience of inpatients with ulcerative colitis throughout

Evaluating Quality of Anesthesiologists Supervision

Chapter 01: Professional Nursing Practice Lewis: Medical-Surgical Nursing, 10th Edition

Can Improvement Cause Harm: Ethical Issues in QI. William Nelson, PhD Greg Ogrinc, MD, MS Daisy Goodman, CNM. DNP, MPH

NHS LOTHIAN Standard Operating Procedure: EHSCP Physiological Observations of Patients in the Community Setting

Provincial Pediatric Early Warning System (PEWS) Clinical Decision Support Tool. Guideline Purpose. Practice Level / Competencies.

Ruchika D. Husa, MD, MS Assistant Professor of Medicine Division of Cardiovascular Medicine The Ohio State University Wexner Medical Center

Provincial Pediatric Early Warning System (PEWS) Clinical Decision Support Tool. Guideline Purpose. Practice Level / Competencies.

Position Number(s) Community Division/Region(s) Yellowknife

Policies and Procedures. I.D. Number: 1145

Transcription:

QUALITY REPORT A Team-Based Approach to Reducing Cardiac Monitor Alarms AUTHORS: Christopher E. Dandoy, MD, MSc, a,b Stella M. Davies, MBBS, PhD, a Laura Flesch, CRNP, a Melissa Hayward, BSN, a Connie Koons, BSN, a Kristen Coleman, BSN, a Jodi Jacobs, CPNP, a Lori Ann McKenna, CNP, a Alero Olomajeye, BS, c Chad Olson, CBET, d Jessica Powers, a Kimberly Shoemaker, a Sonata Jodele, MD, a Evaline Alessandrini, MD, b and Brian Weiss, MD e Divisions of a Bone Marrow Transplantation and Immunodeficiency, and e Oncology, Cancer and Blood Disease Institute, b James M. Anderson Center for Health Systems Excellence, c Clinical Engineering, and d Clinical Integration, Cincinnati Children s Hospital Medical Center, Cincinnati, Ohio KEY WORDS alarm fatigue, cardiac monitor, Model for Improvement, Plan-Do- Study-Act, quality improvement ABBREVIATIONS BMT bone marrow transplant CCHMC Cincinnati Children s Hospital Medical Center CMCP cardiac monitor care process NP nurse practitioner PCA patient care assistant PDSA Plan-Do-Study-Act RN registered nurse SpO 2 pulse oxygen saturation Dr Dandoy conceptualized and designed the study and also drafted the initial manuscript; Ms Flesch, Ms Hayward, Ms Koons, Ms Coleman, Ms Jacobs, Ms McKenna, Ms Olomajeye, Mr Olson, Ms Powers, and Ms Shoemaker designed, conducted, and tested various components of the cardiac monitor care process; assisted in the collection of data; and reviewed and revised the manuscript; Drs Davies, Jodele, Alessandrini, and Weiss directly supervised this improvement project with critical direction throughout implementation and critically reviewed the manuscript. All authors approved the final manuscript as submitted. www.pediatrics.org/cgi/doi/10.1542/peds.2014-1162 doi:10.1542/peds.2014-1162 Accepted for publication Jun 10, 2014 Address correspondence to Christopher Dandoy, MD, Division of Bone Marrow Transplantation and Immunodeficiency, Cincinnati Children s Hospital Medical Center, 3333 Burnet Ave, MLC 11027, Cincinnati, OH 45229. E-mail: christopher.dandoy@cchmc.org PEDIATRICS (ISSN Numbers: Print, 0031-4005; Online, 1098-4275). Copyright 2014 by the American Academy of Pediatrics FINANCIAL DISCLOSURE: The authors have indicated they have no financial relationships relevant to this article to disclose. FUNDING: No external funding. POTENTIAL CONFLICT OF INTEREST: The authors have indicated they have no potential conflicts of interest to disclose. abstract BACKGROUND AND OBJECTIVES: Excessive cardiac monitor alarms lead to desensitization and alarm fatigue. We created and implemented a standardized cardiac monitor care process (CMCP) on a 24-bed pediatric bone marrow transplant unit. The aim of this project was to decrease monitor alarms through the use of team-based standardized care and processes. METHODS: Using small tests of change, we developed and implemented a standardized CMCP that included: (1) a process for initial ordering of monitor parameters based on age-appropriate standards; (2) pain-free daily replacement of electrodes; (3) daily individualized assessment of cardiac monitor parameters; and (4) a reliable method for appropriate discontinuation of monitor. The Model for Improvement was used to design, test, and implement changes. The changes that were implemented after testing and adaptation were: family/patient engagement in the CMCP; creation of a monitor care log to address parameters, lead changes, and discontinuation; development of a painfree process for electrode removal; and customized monitor delay and customized threshold parameters. RESULTS: From January to November 2013, percent compliance with each of the 4 components of the CMCP increased. Overall compliance with the CMCP increased from a median of 38% to 95%. During this time, the median number of alarms per patient-day decreased from 180 to 40. CONCLUSIONS: Implementation of the standardized CMCP resulted in a significant decrease in cardiac monitor alarms per patient day. We recommend a team-based approach to monitor care, including individualized assessment of monitor parameters, daily lead change, and proper discontinuation of the monitors. Pediatrics 2014;134:e1 e9 PEDIATRICS Volume 134, Number 6, December 2014 e1

Physiologic monitoring systems generate visual and audible alarm signals to alert staff to changes in the patient s clinical status. 1 However, each patient can generate hundreds of alarms per day, resulting in thousands of alarms overall for the unit. In fact, some highacuity units experience up to 350 physiologic monitor alarms per patient per day. 2 Providers feel overwhelmed as they differentiate between the large amount of alarms, and they may become desensitized. 3 Desensitization to alarms, or alarm fatigue, leadstoalackofresponse to the alarms due to sensory overload. In January 2010, excessive alarms reached national headlines when a patient s death was directly related to alarm fatigue. 4,5 However, the dangers from excessive alarms are more than an isolated case. From 2005 to 2010, the Emergency Care Research Institute reported 216 physiologic monitor-related deaths. 6 The institute publishes an annual top 10 technology list, and Alarm Hazards has been at the top of the list for the last several years. 6 8 In April 2013, the Joint Commission announced a Sentinel Event Alert to all hospitals based on alarm fatigue and cardiac monitor device care. 9 They reported 80 alarm-related deaths between January 2009 and June 2012, all traced back to alarm-related issues. The major factors reported in these deaths were from alarm fatigue, alarm parameters not customized to the patient, and inadequate staff training on the functioning of the monitors. The present improvement project was performed in the bone marrow transplant (BMT) unit at Cincinnati Children s Hospital Medical Center (CCHMC). Hospital guidelines recommend cardiopulmonary monitoring when patients receive patient-controlled analgesia, infusions, transfusions, and/or when clinically indicated. During a 3-month period (before implementation of our improvement process), the BMT unit had 478 143 cardiac monitor alarms. This total averaged to 5300 alarms per day or 250 alarms per monitor per day. Family and staff commonly complained of the frequency of alarms, and personal communication with the nursing staff revealed that they spent on average 20 to 25 minutes per shift attending to the alarms of patients requiring monitoring. In addition, 95% of the alarms on the floor were false alarms. Despite the high frequency of alarms, there were no acute patient decompensations missed because of alarm fatigue, but the high frequency of alarms raised important safety concerns. The aim of the present project was to create and implement a standardized process for cardiac monitor care to decrease the number of nuisance alarms on the BMT floor. We hypothesized that this process would result in a decrease in the total number of nuisance alarms without missing true patient decompensation. METHODS In January 2012, we established a multidisciplinary alarm oversight task force consisting of key stakeholders, including physicians, nurse practitioners (NPs), nursing leadership, registered nurses (RNs), patient care assistants (PCAs), clinical engineering, and patient family representatives. The team reviewed the current cardiac monitor care practice, published recommendations, identified gaps between practice and evidence, and identified areas of improvement. Using the Model of Improvement, we identified a reliable mechanism to implement the specific processes and tested hypotheses by using Plan-Do- Study-Act (PDSA) measures. Our focus was to create a standardized, team-centered, cardiac monitor care process (CMCP) that could be implemented throughout the BMT unit for patients undergoing cardiopulmonary monitoring. The CMCP included: (1) ageappropriate parameters for patients upon placement on a cardiac monitor; (2) daily electrode changes; (3) daily evaluation of cardiopulmonary monitor parameters; and (4) timely discontinuation of the monitor once the patient was off patient-controlled analgesia or clinically stable. Context CCHMC is a large, urban pediatric medical center, and the BMT team performs 100 to 110 transplants per year. The BMT unit contains 24 beds, and 60% to 70% of patients on the floor are on cardiac monitors, which include pulse oxygen saturation (SpO 2 ) monitoring as well as cardiopulmonary monitoring. Patients admitted to the BMT unit are often hospitalized for up to 30 to 40 days, and they can be on monitors for 50% to 60% of their hospitalization. The clinical providers caring for the patients include 14 BMT attending physicians, 15 fellows, 7 NPs, and 6 hospitalists. Residents do not rotate on the BMT unit. NPs and hospitalists place the majority of the orders on the BMT unit, and hospitalists provide 24-hour coverage for all patients on the floor. The BMT unit employs 130 bedside RNs and 30 PCAs. Family members take an active role in the care of patients in the BMT unit as participants in the Family Advisory Council and through daily family-centered rounds. 10 Improvement Before any intervention, we analyzed our monitor care delivery over a 2-week period of time. This investigation revealed several deficits in cardiac monitor care: (1) only 20% of the monitors were programmed to age-appropriate parameters upon initiation of cardiopulmonary monitoring; (2) the electronic medical record did not contain monitor parameter orders to allow communication betweenthephysicianteamand thenursing staff; (3) there was no standard process for electrode replacement; (4) random e2 DANDOY et al

QUALITY REPORT sampling revealed that electrodes were placed in the wrong location 30% of the time; (5) electrodes were replaced infrequently (up to every 2 weeks); and (6) the monitors were not set according to the patient sspecific needs (ie, a patient with persistent fever would have a high heart rate parameter of 120 beats per minute despite persistent tachycardia [130 beats per minute]). The CMCP was based primarily on our observations of monitor care delivery and PDSA testing. A cardiac monitor log was created to track compliance. Monitor parameters, subjective issues with the alarms, electrode change during the previous 24 hours, and anticipated discontinuation of the monitor were documented daily in the monitor log by the nursing staff and discussed with the hospitalist service each afternoon. Monitor parameters were scrutinized against current vital signs and appropriate clinical need. Family Engagement Family members helped identify barriers to implementation of the CMCP as participants of the multidisciplinary team. The lead change protocol was discussed with family members on initiation of the cardiac monitoring. Family members and patients communicated problems with alarms to the nursing staff, and these issues were addressed directly and documented in the monitor log. Family perception of timely attendance to monitorsbystaffwasmeasuredinthebmt patient satisfaction questionnaire. Standardized Reliable Process for Age-Appropriate Ordering of Monitor Parameters An age-appropriate order set with baseline parameters was created in the electronic medical record. With the use of this order set, the staff was able to change the monitor orders if needed. Initial monitor parameters were documented on the cardiac monitor log, and accuracy was verified on the day of admission by the NP and the hospitalist team. Daily Lead Change Daily electrode change was based on manufacturer guidelines, published findings, and PDSA testing results. Prolonged electrode duration can result in signal impedance and increased signal noise due to decreased conductivity, which lead to increased false alarms. Cvach et al 11 demonstrated a 46% decrease in the number of alarms per day with daily electrode replacement. Our PDSA testing showed similar findings, with a 25% to 30% increase in alarms for each additional day the patient s leads were not changed. Finally, we estimated the hospital cost for lead replacement to be,$1.00 per day. Daily Assessment of Monitor Parameters The cardiac monitor log was completed daily by the nursing staff and reviewed daily by the medical team. Excessive false alarms were investigated and corrected by the nursing staff with the use of monitor troubleshooting algorithm. Clearly Defined Roles and Responsibilities All staff received education through a mandatory computer module and with direct interaction with the various CMCP team members. PCAs were responsible for daily lead changes; RNs documented and verified lead changes, monitored settings, and anticipated discontinuation of the monitor on the monitor care log; and the charge nurses reviewed the log with the providers daily. Intermittently, throughout the study, we asked nurses how much time they spent addressing alarms and performing alarm care. These data were used to test the effectiveness of the interventions. Standardized Reliable Process for Monitor Discontinuation We created intentional redundancy in the assessment of the discontinuation of the monitors. Both providers and nurses assessed the need for the cardiac monitors, and this information was documented in the cardiac monitor log. For patients on a monitor secondary to patient-controlled analgesia use, the pain team created a reliable system to verify discontinuation of the monitor once it was no longer needed. Customized Monitor Delay and Increased Threshold Settings Customized monitor delays and increased parameter threshold settings have been showntobeaneffectivemethodforreducing false alarms. 12,13 We increased the SpO 2 alarm delay from 5 to 10 seconds because the majority of the SpO 2 alarms self-correct within the delay period set. High respiratory rate, without other vital sign abnormalities, is nearly always a false alarm. The safety of increasing the respiratory rate on monitors was evaluated by the hospital oversight committee, and in our unit, the high respiratory rate was determined to not be an actionable alarmalone;therefore,thehighrespiratory rate limit on all monitors on the floor was increased. Study of the Improvement The process was analyzed through a quantitative time series study design. We measured each component of the CMCP individually and cumulatively. The measured CMCP components included: (1) standardized process of initial ordering and setting of monitor parameters based on age-appropriate standards; (2) pain-free daily replacement of electrodes; (3) daily individualized assessment of cardiac monitor parameters; and (4) appropriate discontinuation of the monitor when it was no longer indicated (Fig 1). On a weekly basis, each component of the process was measured independently as a percentage of patients who had the intervention (numerator) over the number of opportunities. PEDIATRICS Volume 134, Number 6, December 2014 e3

We measured the CMCP in whole as a percentile of the number of completed components of the CMCP over the number of opportunities. For example, if apatient wasadmitted onday 1 andthe correct parameters were applied to the monitor, electrodes were replaced daily for 5 days, the patient received daily evaluation of proper parameters for 5 days, and the monitor was discontinued appropriately, then this method would equate to 12 opportunities. We measured the numberof alarms per monitored day on monitored patients. This amount was calculated as a total sum of the numberof alarms,divided by the number of monitor days on the unit during the preceding week. Finally, as a balancing measure, we examined each code blue and staff emergency to determine if a change in the cardiac monitor care could have recognized the clinical status change at an earlier time. Data Collection Data to generate the measures and weekly reports were collected daily on the monitor log, which was a communication tool between the nursing staff and the physician team. The log included current monitor parameters, family and nursing concerns regarding monitors, electrode change, and anticipated discontinuation of the cardiac monitor. The data from these sheets were collected and assessed weekly. Analysis The difference in family perception of timely response before and after the intervention was evaluated by using Fisher s exact test. Statistical process control methods were used to monitor changes in care processes and health outcomes. Annotated run charts and U charts were developed and updated weekly. We established a median, illustrated as the centerline on all control charts with upper and lower control limits in the U charts. Standard industry criteria were used to determine if observed changes in measures were chance random variation (common cause variation) or due to a specific assignable cause, in this case the intervention (special cause variation). 14,15 Human Subjects Protection The present initiative fell within the CCHMC institutional review board s guidance for quality improvement projects that did not constitute human subjects research. FIGURE 1 Key driver diagram of the CMCP implementation. CMCL, cardiac monitor care log; EMR, electronic medical record. e4 DANDOY et al

QUALITY REPORT FIGURE 2 Timeline of steps to decrease false alarms. January through February: initial PDSA testing on lead replacement, timing of lead replacement, and timing of daily assessment of monitor parameters. March: all providers received education on the CMCP with appropriate assignment of roles and responsibilities. CMCL, cardiac monitor care log. RESULTS The study period was from January 2013 through November 2013 (Fig 2). Initial PDSA testing on electrode replacement, the timing of electrode replacement, and the timing of daily assessment of monitor parameters occurred in January and February.InMarch,formaleducationwas given to all nursing staff in the form of an online module. In June, we increased the high respiratory rate on all monitors, and in July we increased the SpO 2 alarm delay. Compliance With the CMCP Overall compliance with the CMCP remained stable at a median of 38% (range: 33% 43%) through February 2013 as PDSA testing was accomplished. Once roles and responsibilities were determined and the process was clearly defined, full implementation continued on the unit. This process occurred primarily in March 2013. The unit s overall compliance with the CMCP increased to amedianof95%(range:73% 98%), which was sustained for 7 months from MaytoNovember2013(Fig3). With redundancy in the system, we maintainedahighdegreeofreliabilityinpatients having proper initial monitor parameters (Fig 4A). This reliability was accomplished with the creation of the age-specific monitor parameter settings in the electronic medical record order set, as well as daily review by the providing team. With the use of PDSA testing, we discovered improved sustainability of daily evaluation of the FIGURE 3 Overall compliance with the CMCP. Measured as percent compliance (number of completed interventions/number of opportunities). PEDIATRICS Volume 134, Number 6, December 2014 e5

parameters (Fig 4B) by having the cardiac monitor log completed by the nursing staff duringthedayandwithreviewbythe hospitalist service in the afternoon. Patient discomfort was the largest barrier to daily lead change. PDSA testing revealed that soaking the leads with water while taking a bath, forat least 1 minute, reduced patientpain.dailyleadchangecompliance improved from a median of 55% (range: 41% 69%) to 85% (range: 63% 97%) during the 11-month period (Fig 4C). Multiple interventions enabled a high percentage of patients being taken off monitors appropriately (Fig 4D). Nursing staff documented and discussed the anticipated monitor discontinuation date daily. Number of Alarms As CMCP compliance improved, there was a decrease in the number of alarms, FIGURE 4 Compliance with the individual components of the CMCP. A, Percent compliance with age-appropriate alarm parameters upon placement on the monitor. B, Percentage of patients with daily assessment of alarm parameters. C, Percentage of leads replaced daily. D, Percentage of monitors discontinued appropriately. e6 DANDOY et al

QUALITY REPORT FIGURE 4 Continued. from a median of 180 to 40 during the study period (Fig 5). We maintained a rate of 50 to 70 alarms per day with utilization of the CMCP alone and were able to further decrease the number of alarms by increasing the high respiratory rate parameter as well as increasing the SpO 2 delay. During the implementation, the median number of false alarms on the floor fell from 95% to 50%. The median time that individual nurses spent addressing frequent alarms decreased from 20 to 25 minutespershift to 10 minutes per shift, including the time it took each nurse to complete the monitor log. Ninety-five percent(20of 21) of families reported that alarms were addressed in a timely manner after the PEDIATRICS Volume 134, Number 6, December 2014 e7

interventionversus 87%(53 of 61) before the intervention (P =.43). Finally, no acute decompensations, code, or staff emergency occurred or were missed because of the CMCP. DISCUSSION We describe a standardized, team-based CMCP to reduce the number of alarms in a single unit. Implementation of the CMCP resulted in an 80% decrease in the number of alarms per patient day. We sustained a 55% reduction in the number of alarms through human factor dependent processes such as changing leads, and an additional 25% reduction was accomplished through customization of the monitor settings. This standardized, patient-centered approach decreased the number of false alarms and the time nurses spent addressing monitor issues. The ageappropriate initial setting of the monitor parameters allowed for a set starting point for all patients, and having the order set in the electronic medical record allowed for proper communication between the nurses and the physician staff. Review of the appropriateness of the parameters on the first day of monitoring allowed for early patientspecific adjustments. Daily review of the monitor parameters, with family and nursing input, allowed for frequent evaluationoftheappropriatenessandsafetyof the current parameters. This daily assessment, with anticipation of discontinuation of monitoring, allowed for the timely discontinuation of the monitor. Daily lead changes can be cumbersome and painful for patients. We discovered that by replacing leads during daily baths, the unpleasantness of the lead change was manageable. By taking advantage of existing habits of the providers, and decreasing the pain experienced by the patient, we maintained an 85% compliance rate with daily lead changes. Interestingly, in September 2013, our compliance with the CMCP decreased with increased variation. Although we stayed within the control limits of our system, our number of alarms increased at the same time. When the variation improved, so did the number of alarms per monitored day. To improve the sustainability of the CMCP, we review compliance weekly and analyze failures with the nursing and clinical staff. Accountability, with clearly defined roles and responsibilities, helps components of thecmcptobeapartofthestaff s daily responsibilities. We believe that the roles and responsibilities entailed in the CMCP can be applied to most units with cardiac monitor care. The present study was limited by the lack of a comprehensive patient satisfaction survey before and after our intervention. Our patient questions were limited to the BMT patient satisfaction survey, which asks if nursing staff responded to alarms in a timely manner. Only a small percentage of patients completed the questionnaire, and the results did not show a statistically significant difference. In addition, because implementation of all these processes occurred at nearly the same time, it is difficult to know exactly what intervention made the largest change. FIGURE 5 Number of alarms per monitored patient-day. e8 DANDOY et al

QUALITY REPORT We found a significant decrease in the number of alarms per monitored patientday with the implementation of the standardized CMCP. We believe these efforts have improved safety on the BMT unit by decreasingthenumberofalarmsandthe risk of alarm fatigue. With fewer false alarms, the staff can address alarms more promptly. The patients and their caregiver s daily routines are less interrupted by alarms because of the CMCP. Finally, these efforts have decreased the amount of time the nursing staff spend addressing false alarms on the unit. CONCLUSIONS We recommend a team-based approach to cardiac monitor care. Initial parameters should be specific for the patient and evaluated frequently, electrodes should be replaced daily, and monitors should be discontinued when no longer clinically indicated. ACKNOWLEDGMENTS The authors thank the physicians, nurses, PCAs, NPs, hospitalists, fellows, and staff at CCHMC, and especially the patients and their families. We thank Brenda Lee, Christopher Jordan, and the CCHMC Anderson Center staff for their support. REFERENCES 1. Graham KC, Cvach M. Monitor alarm fatigue: standardizing use of physiological monitors and decreasing nuisance alarms. Am J Crit Care. 2010;19(1):28 34, quiz 35 2. Mitka M. Joint Commission warns of alarm fatigue: multitude of alarms from monitoring devices problematic. JAMA. 2013;309 (22):2315 2316 3. Bell L. Monitor alarm fatigue. Am J Crit Care. 2010;19(1):38 4. Alarm-notification problem spotlighted in Boston globe is all too common. Health Devices. 2010;39(4):139 5. McKinney M. Alarm fatigue sets off bells. Mass. incident highlights need for protocols check. Mod Healthc. 2010;40 (15):14 6. Keller JP Jr. Clinical alarm hazards: a top ten health technology safety concern. J Electrocardiol. 2012;45(6):588 591 7. McKinney M. High-tech hazards. Alarm hazards tops ECRI list of health dangers. Mod Healthc. 2011;41(45):18 8. Alarms #1 tech risk despite focus, according to ECRI. Hosp Peer Rev. 2013;38(12):133 135 9. Joint Commission. Medical Device Alarm Safety in Hospitals. Sentinel Event Alert. 2013; (50):1 3 Available at: www.jointcommission. org/assets/1/18/sea_50_alarms_4_5_13_fi- NAL1.PDF. Accessed April 30, 2013 10. Muething SE, Kotagal UR, Schoettker PJ, Gonzalez del Rey J, DeWitt TG. Familycentered bedside rounds: a new approach to patient care and teaching. Pediatrics. 2007;119(4):829 832 11. Cvach MM, Biggs M, Rothwell KJ, Charles- Hudson C. Daily electrode change and effect on cardiac monitor alarms: an evidencebased practice approach. J Nurs Care Qual. 2013;28(3):265 271 12. Sendelbach S, Funk M. Alarm fatigue: a patient safety concern. AACN Adv Crit Care. 2013;24(4):378 386; quiz 387 378 13. Welch J. An evidence-based approach to reduce nuisance alarms and alarm fatigue. Biomed Instrum Technol. 2011;(suppl):46 52 14. Benneyan JC, Lloyd RC, Plsek PE. Statistical process control as a tool for research and healthcare improvement. Qual Saf Health Care. 2003;12(6):458 464 15. Amin SG. Control charts 101: a guide to health care applications. Qual Manag Health Care. 2001;9(3):1 27 PEDIATRICS Volume 134, Number 6, December 2014 e9

A Team-Based Approach to Reducing Cardiac Monitor Alarms Christopher E. Dandoy, Stella M. Davies, Laura Flesch, Melissa Hayward, Connie Koons, Kristen Coleman, Jodi Jacobs, Lori Ann McKenna, Alero Olomajeye, Chad Olson, Jessica Powers, Kimberly Shoemaker, Sonata Jodele, Evaline Alessandrini and Brian Weiss Pediatrics originally published online November 10, 2014; Updated Information & Services Permissions & Licensing Reprints including high resolution figures, can be found at: http://pediatrics.aappublications.org/content/early/2014/11/05/peds.2 014-1162 Information about reproducing this article in parts (figures, tables) or in its entirety can be found online at: http://www.aappublications.org/site/misc/permissions.xhtml Information about ordering reprints can be found online: http://www.aappublications.org/site/misc/reprints.xhtml

A Team-Based Approach to Reducing Cardiac Monitor Alarms Christopher E. Dandoy, Stella M. Davies, Laura Flesch, Melissa Hayward, Connie Koons, Kristen Coleman, Jodi Jacobs, Lori Ann McKenna, Alero Olomajeye, Chad Olson, Jessica Powers, Kimberly Shoemaker, Sonata Jodele, Evaline Alessandrini and Brian Weiss Pediatrics originally published online November 10, 2014; The online version of this article, along with updated information and services, is located on the World Wide Web at: http://pediatrics.aappublications.org/content/early/2014/11/05/peds.2014-1162 Pediatrics is the official journal of the American Academy of Pediatrics. A monthly publication, it has been published continuously since 1948. Pediatrics is owned, published, and trademarked by the American Academy of Pediatrics, 141 Northwest Point Boulevard, Elk Grove Village, Illinois, 60007. Copyright 2014 by the American Academy of Pediatrics. All rights reserved. Print ISSN: 1073-0397.