EDUCATIONAL COMMENTARY KEY COMPONENTS OF AN INDIVIDUALIZED QUALITY CONTROL PLAN

Similar documents
3/14/2016. The Joint Commission and IQCP. Objectives. Before Getting Started

IQCP. Ensuring Your Laboratory s Compliance With Individualized Quality Control Plans. November/December 2016

CLIA S NEW IQCP SEABB. March 19, Linda Sigg, MT(ASCP)SBB,CQA(ASQ) Staff Lead Assessor, Accreditation, AABB

Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 08/21/2017)

Laboratory Risk Assessment: IQCP and Beyond. Ron S. Quicho, MS Associate Project Director Standards and Survey Methods, Laboratory July 18, 2017

Learning Objectives. Individualized Quality Control Plans. Agenda. Another Way To Determine QC? Hooray!!!! What is QC?

IQCP January Is Coming Fast What Do I Do?!? Jean Ball Bold, MBA, MT(HHS), MLT(ASCP

CLIA s New IQCP Requirements Are in Effect, or Are They?: Implementing Laboratory Risk Management Now to Ensure Success

5/8/2015. Individualized Quality Control Plans (IQCP) Changes to the CMS Quality Requirements. CLIA Quality Control Evolution of the Process

Pro-QCP SAMPLE REPORT

Policy Subject Index Number Section Subsection Category Contact Last Revised References Applicable To Detail MISSION STATEMENT: OVERVIEW:

Plan for Quality to Improve Patient Safety at the POC

CAP Most Frequent Deficiencies and How to Avoid Them. March 11, 2015

CLIA & Individualized Quality Control Plan (IQCP) Judith Yost Director Division of Laboratory Services

Point of Care Quality Management. Procedure. Approving Authority: President and CEO, Keith Dewar

CLIA & Individualized Quality Control Plan (IQCP) Karen W. Dyer MT(ASCP), DLM Director (Acting) Division of Laboratory Services

Heart of America POC Group Quality Management Making it Meaningful

The CLIA regulations..


CE Update [generalist compliance/regulation management/administration and training] COLA Accreditation An Educational Experience

CAP Accreditation and Checklists Update. Lyn Wielgos, MT(ASCP) Checklist Editor, CAP Accreditation Programs

COMMISSION ON LABORATORY ACCREDITATION. Laboratory Accreditation Program TEAM LEADER ASSESSMENT OF DIRECTOR & QUALITY CHECKLIST

Standards for Laboratory Accreditation

Point of Care Testing

insights INTO Quality Control MAY/JUNE 2013 COLA S

QMP-LS: A Canadian Regional EQA Program How Labs Get In and Out of Trouble in Ontario

Standards for Forensic Drug Testing Accreditation

Personnel. From RLM, COM, GEN and TLC Checklists

US ): [42CFR ]:

Karen W. Dyer MT(ASCP), DLM Director, Division of Laboratory Services Centers for Medicare & Medicaid Services CLIA

PROGRAM GUIDE - UNIVERSITY CLIA REGISTERED LABORATORIES COMPLIANCE COMMITTEE

Improving Your POC Program: An Upside Down Map. Sheila K. Coffman MT(ASCP)

TESTIMONY OF THOMAS HAMILTON DIRECTOR SURVEY & CERTIFICATION GROUP CENTER FOR MEDICAID AND STATE OPERATIONS CENTERS FOR MEDICARE & MEDICAID SERVICES

Quality Assurance Program For Hospital Based Point of Care Testing. Presented by: Jeanne Mumford, MT(ASCP) Pathology Supervisor, QA Specialist

Master. Point-of-Care-Testing Checklist. CAP Accreditation Program

SPECIMEN REQUIREMENTS

SUCCESSION PLANNING: FILLING A LABORATORY DIRECTOR S SHOES PART ONE

CAP Forensic Drug Testing Accreditation Program Standards for Accreditation

Preanalytical Errors in Laboratory - Their Consequences and Measures to Reduce Them

Reducing Risk in the Laboratory: IQCP In Action! 9/24/2015

Hannah Poczter, AVP, Laboratories, Cari Gusman, Administrative Director, Ed Giugliano, PhD, Project Manager, Certified Six Sigma Black Belt

Master. Point-of-Care-Testing Checklist. CAP Accreditation Program

STANDARDS Point-of-Care Testing

Quality Medical and Laboratory Practice in Cellular Therapy

CAP Companion Society Meeting at USCAP 2009 Quality Assurance, Error Reduction, and Patient Safety in Anatomic Pathology

SAMPLE. Statistical Quality Control for Quantitative Measurement Procedures: Principles and Definitions

Subject: Quality Management for Origin date: 3/06 Point of Care and Waived Testing Reviewed: 2009 /2010 Revised: 2/2009

Pre-Analytical Laboratory Procedures for Medical Office Staff CLP 011a-001 Revised March, 2007

insights INTO Preparing Your Laboratory For Its Next Survey COLA S FALL 2016 ALSO IN THIS ISSUE: Letter from the Chair... 2

Compounded Sterile Preparations Pharmacy Content Outline May 2018

TITLE: POINT OF CARE TESTING

Tutorial: Basic California State Laboratory Law

What s New and Improved for the Laboratory Program in 2013 April 23, 2013

PURPOSE: This policy provides an overview of SHANDS Jacksonville Laboratory s commitment to the care and safety of the patients we serve.

ASSEMBLY BILL No. 940

Center for Medicaid and State Operations/Survey and Certification Group

Clinical and Laboratory Standards Institute: Addressing POCT Needs; The Good, The Bad, and The Risky

Disclosures. Relevant Financial Relationship(s): Nothing to Disclose. Off Label Usage: Nothing to Disclose 6/1/2017. Quality Indicators

QC Explained Quality Control for Point of Care Testing

Performance of Point-of-Care Testing in Unaccredited Settings:

Best Practices for Equipment Calibration and Analytical Controls in the Diagnostics Laboratory

The Lab General Checklist

Clinical Laboratory Standards of Practice

Bureau of Clinical Laboratories Quality Assessment Plan

Laboratory Services. Specimen Collection & Rejection Procedure

POSITION DESCRIPTION

Catholic Health Initiatives

6/28/2016. Questions? Workshop 6 CAP Inspection Preparation Thursday, June 23, 2016

A COLA White Paper: FEDERAL GOVERNMENT QUESTIONS QUALITY IN WAIVED TESTING.

Martin Health System Stuart, Florida Laboratory Services. Laboratory Services and Policies

insights COLA Updates Into MAY / JUNE 12 COLA s

Massachusetts General Hospital Point of Care Testing Program

QUALITY POLICY MANUAL. Revision: 05 Author: T. Joseph Issue Date: 6/6/2010 Approved By: Dr S. King

Specimen Collection Regulations and Standards

Allen D. Leman Swine Conference

Centers for Medicare and Medicaid Services (CMS) Survey and Certification Group (SCG) Mission:

Fulton County Medical Center. Position Description. Pathologist, Laboratory Manager, and Medical Technologist

Regulatory,Quality & Emergency Preparedness. MaryBeth Parache Director, Quality Affairs New York Blood Center

Plan for Quality to Improve Patient Safety at the POC

Point of Care Testing. BOPCC May 31, 2011 Beatrice O Keefe, Chief Laboratory Field Services California Department of Public Health

Master. Point-of-Care-Testing Checklist. Every patient deserves the GOLD STANDARD... CAP Accreditation Program

Adverse Events: Thorough Analysis

Internal Lab Inspections: Are You Inspection Ready? Presented by: Jeanne Mumford, MT(ASCP) Manager, Point of Care Testing, JHM

SOP WP6-QUAL-04, Version 1.0, 23 February 2014 Page 1 of 8. SOP Title: Laboratory (GCLP) supervision visits

What s New in Statistical Quality Control Guidance: CLSI s C24 Updates. August 2, 2016

PERSONNEL REQUIREMENTS. March 9, 2018

VUMC Office of Research Research Core Facilities/Shared Resources 2015 Professional Development Track. Core Research Assistant I

AMERICAN BOARD OF HISTOCOMPATIBILITY AND IMMUNOGENETICS Laboratory Director. Content Outline

THE VALUE OF CAP S Q-PROBES & Q-TRACKS

College of American Pathologists 325 Waukegan Road, Northfield, Illinois Advancing Excellence

Medicare Program; Announcement of the Reapproval of the Joint Commission as an

Standards, Guidelines, and Regulations

Daralyn Hassan, MS, MT(ASCP) April 3rd, 2014 CLIA

Patient Care Coordination Variance Reporting

TITLE 252. DEPARTMENT OF ENVIRONMENTAL QUALITY CHAPTER 302. FIELD LABORATORY ACCREDITATION

Standards for Biorepository Accreditation

Organization for Economic Co-operation and Development

POINT OF CARE TESTING MED Laboratory Branch Kim DeGroat, RMLS - Frankfurt Region Wilfred Lovelock, RMLS - Dakar Region

Technical Job Family: Technician Progression

International Organization for Standardization (ISO) 15189

Transcription:

Commentary provided by: E Susan Cease MT(ASCP) Laboratory Manager Three Rivers Medical Center Grants Pass, OR EDUCATIONAL COMMENTARY KEY COMPONENTS OF AN INDIVIDUALIZED QUALITY CONTROL PLAN Educational commentary is provided through our affiliation with the American Society for Clinical Pathology (ASCP). To obtain FREE CME/CMLE credits, click on Earn CE Credits under Continuing Education on the left side of the screen Learning Objectives On completion of this exercise, the participant should be able to identify the mandatory elements of an individualized quality control plan (IQCP); understand how to identify situations in which a risk assessment must be completed; and determine which tests qualify for an IQCP. IQCP: Introduction The end of 2015 brought a flurry of activity as laboratories evaluated how the Individualized Quality Control Plan (IQCP) option from the Centers for Medicare & Medicaid Services (CMS) would affect their operations. Resources available from CMS, and then from accrediting agencies and vendors, provided guidelines and forms to assist labs in determining which test systems qualified for IQCP and how to develop a viable plan. Fast forward two years, and many labs have moved into the next phase of IQCP the continual Risk Assessment (RA) and evaluation of the Quality Assurance cycle. The focus on defining how frequently external IQCP = Individualized controls will be performed is just one component of an IQCP. This Quality Control Plan article will provide an overview of the components and a few of the tools Permits laboratories to that can be used in RA and evaluation of the IQCP. develop site-specific quality plans for non-waived tests IQCP: A Historical Review The Centers for Medicare & Medicaid Services revised the Clinical Laboratory Improvement Amendments of 1988 (CLIA) interpretive guidelines ( 493.1250) to implement IQCP and eliminate Equivalent Quality Control (EQC), effective January 1, 2016. Labs now have the option to implement IQCP or maintain the CMS-defined default frequency for external QC for qualifying non-waived tests. EQC is no longer an option for test systems with internal controls. EQC allowed external controls to be run with each new lot number, each new shipment, or once every 30 days, whichever was most frequent. For most tests, default QC frequency for external controls is two levels once per day of patient testing. Using RA, the lab may determine that less frequent QC provides acceptable QA, and it can develop and use an IQCP to manage that test. A laboratory cannot reduce the default frequency of external control American Proficiency Institute 2018 1 st Test Event 1

If the manufacturer s QC protocol is less stringent than the regulatory requirement, then you will need to do an IQCP to follow the manufacturer s QC protocol. testing without using an IQCP, and under no circumstances can the lab reduce frequency below the manufacturer s recommendation. Not all non-waived tests are eligible for IQCP. Specifically excluded are any tests performed and billed under Anatomic Pathology, Oral Pathology, Histology, or Cytology. Test media and reagents used in Microbiology for identification and susceptibility testing are eligible for IQCP. In many laboratories, the Microbiology section uses IQCP to a greater extent than any other department. Developing an IQCP: A Story in Three Parts Risk Assessement (RA) Quality Control Plan (QCP) Quality Assurance (QA) IQCP IQCP 1 is an all-inclusive approach to assuring quality. It includes many practices that a laboratory already uses to ensure quality testing, beyond requiring that a certain number of QC materials be tested at a designated frequency. IQCP uses a Risk Assessment (RA) to develop a Quality Control Plan (QCP) and monitors the effectiveness with Quality Assurance (QA). RA is a detailed review of all three phases of testing (preanalytic, analytic, and postanalytic) using at least five components (specimen, environment, testing personnel, reagents, and test system). Risk Assessment Quality Conrol Plan QCP is a document that describes the practices, resources, and procedures to control the quality of a particular test process. QA is the review system for the ongoing monitoring of the effectiveness of QCP. The monitoring should include at least the following: specimen, environment, testing personnel, reagents, and test system. IQCP Quality Assurance Risk Assessment The RA will identify and evaluate any possible risks that may occur in your testing process. Risks are potential failures and sources of error that can affect the accuracy and precision of test results. American Proficiency Institute 2018 1 st Test Event 2

The RA requires review of the entire testing process (preanalytic, analytic, and postanalytic) across a minimum RA = 5 Components + 3 Phases of five components: Specimen, Environment, Testing Personnel, Reagents, and Test System. Risk Assessment (RA) Specimen Environment Testing Personnel Reagents Test Systems Preanalytic Preanalytic Preanalytic Preanalytic Preanalytic Analytic Analytic Analytic Analytic Analytic Postanalytic Postanalytic Postanalytic Postanalytic Postanalytic Step 1. Conducting the RA begins with compiling and reviewing data. Determine what you are already doing, collect data, and make sure you have a detailed process map. These data can be new or historical, but they must include the laboratory s own data; see Table 1 for common data sources. It is important to have a multidisciplinary team working on the IQCP to ensure that all phases and components are evaluated during the assessment. See the list of resources at the end for links to governmental and accreditation organizations that offer tool kits, forms, and instructions for all elements of the IQCP. Table 1. Common data sources. Sources for Data Laboratory procedures/standard operating procedures (SOPs) Historical QC data, including data from a previously conducted EQC study Manufacturer s instructions/package inserts Instrument and troubleshooting manuals Manufacturer s alerts and bulletins, FDA Alerts Data obtained through verification or establishment of performance specifications Testing personnel training and competency records Specimen-rejection logs Calibration data Instrument correlation data Proficiency test results and data Records of complaints and corrected reports Regulatory and accreditation requirements Scientific publications Test process flowcharts or maps Turnaround time reports American Proficiency Institute 2018 1 st Test Event 3

Step 2. After compiling the information, analyze all five components in each of the three phases of testing. A process map is commonly used to ensure that all steps of the testing process are covered; see Figure 1. Use tools like the Fishbone and the 5 Whys to evaluate each step involved in the process map. Figure 2 provides an example Fishbone diagram. Many accreditation agencies and vendors, as well as CMS, supply forms that you can adapt to your lab. Start Process Map Be sure to include all steps in the process Specimen Process QC/QA Passes Result Fails Repair Analysis Figure 1 Questions You Should Ask as You Evaluate the Data: What are the chances of an error happening at this point? Is there a step in the process that helps reduce the chance of an error? If not, how do I minimize or reduce the likelihood that this error could occur anywhere in the testing process? What is the impact to the patient? Try using the 5 Whys. Why was the test delayed? Why did the sample have to be redrawn? Why was the first specimen rejected? Why did the phlebotomist draw the incorrect tube? Why is the training guide incorrect? Why was the SOP not updated when the kit was changed? Continue until you find the root cause. Table 2 offers suggestions for steps/points to evaluate. In laboratories with multiple identical devices (same make and model), an RA will need to be completed for each individual location and/or device, as well as when different groups of personnel perform the testing. All RA documentation must be maintained for at least two years after the corresponding QCP has been discontinued. American Proficiency Institute 2018 1 st Test Event 4

1-Specimen 2-Environment 3-Personnel 4-Reagents 5-Test System 1 2 3 1 2 3 1 2 3 Identify potential hazards Test Result 4 5 4 5 4 5 Preanalytic Analytic Postanalytic 1= Specimen Sample Integrity Lipemia Hemolysis Interfering substances Clotted Incorrect tube Sample Presentation Bubbles Inadequate volume 1. List all possible steps, complications, or requirements for each component. 2. Repeat for each component for preanalytic, analytic, postanalytic phases. Preanalytic Analytic Postanalytic Test Result Figure 2. Example of a Fishbone diagram American Proficiency Institute 2018 1 st Test Event 5

Table 2. Components of a Fishbone diagram and points to consider for each. Component SPECIMEN ENVIRONMENT TESTING PERSONNEL REAGENT TEST SYSTEM Points to Consider Must include preanalytic, analytic, postanalytic for each component Patient preparation Specimen collection Specimen labeling Specimen storage, preservation, and stability Specimen transportation Specimen processing Specimen acceptability and rejection Specimen referral Temperature Airflow/ventilation Light intensity Noise and vibration Humidity Altitude Dust Water Space Utilities (electrical failure/power supply variance or surge) Training Competency Education and experience Staffing Shipping/receiving Storage condition requirements Expiration date (may differ based on storage requirements) Preparation Inadequate sampling Clot-detection capabilities Capabilities for detection of interfering substances (e.g., hemolysis, lipemia, icterus, turbidity) Calibration-associated issues Mechanical/electronic failure of test system Optics Pipettes or pipettors Bar code readers Failure of system controls and function checks Built-in procedural and electronic controls (internal controls) External or internal liquid quality control (assayed vs unassayed) Temperature monitors and controllers Software/hardware Transmission of data to LIS Result reporting Step 3. Rate the risks identified in the assessment by severity and likelihood of occurrence. One common approach is to use an RA tool. An example is the Risk Matrix. 2 Establish the frequency of occurrence and severity of harm using the criteria listed. Complete a Risk Matrix and use the frequency and harm values to assess whether the risk is acceptable or unacceptable. Frequency of Occurrence Unlikely (once every2-3 yrs) Occasional (1/yr) Probable (1/mo) Frequent (1/wk) Severity of Harm Negligible (temporary discomfort) Minor (temporary injury; not requiring medical intervention) Serious (impairment requiring medical intervention) Critical (permanent impairment requiring medical intervention) American Proficiency Institute 2018 1 st Test Event 6

For example, an error that can cause serious harm, but is unlikely to occur, may be an acceptable risk. The RA for any given test system may look very different in different laboratories. For example, the same risk may be assigned different frequency or severity ratings by different sites, based on testing personnel, intended medical use, or specific environmental factors. Frequency of Harm Severity of Harm Negligible Minor Serious Critical Frequent Not Acceptable Not Acceptable Not Acceptable Not Acceptable Probable Acceptable Not Acceptable Not Acceptable Not Acceptable Occasional Acceptable Acceptable Acceptable Not Acceptable Unlikely Acceptable Acceptable Acceptable Acceptable After the lab has identified the sources of potential failures and errors for a testing process, the findings are used to develop the QCP. Any factors deemed Not Acceptable should be monitored. Quality Control Plan The QCP is a written document that will define how you monitor the accuracy and precision of the performance over time. It must include the number, type, and frequency of QC testing with criteria for acceptable QC results The QCP will describe the practices, resources, and procedures to control the quality of a particular test process. It should be able to provide for the immediate detection of errors that occur and identify changes that happen over time in the test system, environmental conditions, or variance in operator performance. CLIA does not require the use of any specific tools or format in the development of an IQCP. The new IQCP option gives laboratories the flexibility to determine the appropriate control frequency and required monitors for their unique environment. Although an annual review by the Medical Director or his/her designee is required, an additional monthly or quarterly review may also be required, depending on your accrediting body. See Figure 3. In other words, include What you monitor or audit How often you monitor/audit (frequency) How you know it is acceptable And have documentation available to support compliance with the plan. American Proficiency Institute 2018 1 st Test Event 7

Example: Quality Control Plan Laboratory ABC Test/Instrument XYZ Type of Control Frequency Criteria for Acceptability Internal QC every time a test is run Every use Within manufacturer s limits using automated lockout Two levels of EQC Weekly Within defined QC limits Monitor temp of Daily, monthly Within 2 o -8 o C reagent storage area supervisor review Training and Competence At initial training, 6 months later, Successful test performance including external QC and PT testing Assessment then annually Etc. Monthly review of QMP data Monthly by Supervisor All elements meet criteria for acceptability. Review by Medical Director Annually All elements meet criteria for acceptability. Etc. Reviewed and Approved Medical Director Date MONTHLY QC REVIEW/IQCP ASSESSMENT FORM TEST SYSTEM LOCATION(S) 1. Determine if the IQCP Quality Control, Instrument/Equipment Maintenance, and Function checks were performed as defined in the IQCP. 2. Evaluate errors relating to preanalytic, analytic and postanalytic phases of testing 3. Review any complaints from clinicians and/or other healthcare providers regarding the quality of testing. 4. Describe and evaluate corrective actions taken if problems were identified. 5. Did the IQCP mitigate potential risk for the test system? If the IQCP did not mitigate potential risk, list additional investigation activities and suggestions for modification of the IQCP.. Reviewed by: Date: Based on CAP and COLA documents. Figure 3. Example monthly or quarterly review documentation. American Proficiency Institute 2018 1 st Test Event 8

Based on CAP and COLA documents. Annual Review: Instrument XYZ Lab ABC Location: North Side Quality Control performed appropriately QC issues resolved? and reviewed monthly Temperature log sheets completed and reviewed monthly Out of range or missing temperatures resolved? Maintenance logs completed and reviewed monthly Incomplete data? Corrective actions recorded? Instrument issues resolved and recorded Instrument failures or downtime? Proficiency testing performed and reviewed Unsuccessful PT performance? Sampling of personnel training/competency Retraining needed? reviewed Sampling of patient results reviewed Reporting errors corrected? Relevant quality indicators reviewed Turnaround time, corrected reports, specimen rejection, etc.? Laboratory occurrence reports Corrective actions completed? Complaint reports Physician or caregiver concerns? IQCP reapproval by laboratory director or designee Questions or concerns identified in the review/reapproval? Have test process failures been identified? a. Assess the use (e.g. timely, effective) of the monthly review process of Quality Control, temperature, and maintenance logs to identify problems b. Record any corrective action for patient results affected by the testing process failure. c. Evaluate the effectiveness of the corrective action taken. Have any changes been made to the five elements of the Risk Assessment (i.e. specimen, reagents, environment, testing personnel, test system) requiring reevaluation of the Quality Control Plan? Have any changes been made to the Quality Control Plan? a. Specify any updates/modifications Have revisions to the Quality Control Plan been signed by the laboratory director (including signature and date)? Is the IQCP sufficient to mitigate risk in this laboratory? If no, explain actions to be taken. Reviewed by Date Quality Assurance QA is the review system for the ongoing monitoring of the effectiveness of the QCP. The monitoring should include at least the following: specimens, reagents, environment, testing personnel, and test American Proficiency Institute 2018 1 st Test Event 9

system. This should be reflected in the documentation required to comply with the Quality Management Plan (QMP). Whenever a step or process fails to meet the criteria set out, the cycle begins again. You also need to reassess whenever any component or step in the system changes. If the testing location is changed, the humidity changes as a result of a new cooling system, the manufacturer modifies the kit, or the job description of staff performing the testing changes, you will need to perform a new RA. Additional Risk Prevention/Control measures Life Cycle Risk Management Process Severity of Harm Hazard Identification Risk Estimation Risk Evaluation Risk Control Risk Monitoring Probability of Harm Failure Investigation New Hazard? Changed Severity? Changed Frequency? When the laboratory discovers a testing process failure, you will need to conduct an investigation to identify the cause of the failure, its impact on patient care, appropriate corrective action for affected patients, and appropriate modifications to the QCP to prevent recurrence. The investigation must include documentation of all corrections, corresponding corrective actions for all patients affected by the testing process failure, and evaluation of the effectiveness of the corrective action(s). The laboratory must implement the correction(s) necessary to resolve the failure and reduce the risk that the failure will recur in the future. 1 American Proficiency Institute 2018 1 st Test Event 10

Summary An Individualized Quality Control Plan is a cycle with three main components. There are required elements, but it also Quality Risk offers the flexibility to customize the plan to fit a specific Control Assessment Plan laboratory and location. Multidisciplinary teams ensure that IQCP the IQCP plan is robust and conforms to frequency guidelines, and the medical director review and integration into the Quality Management Plan ensures that all Quality Assurance departments are aware of the IQCP. An Individualized Quality Control Plan is about the patient and ensuring that we provide accurate, dependable results every time. Key Points One IQCP for each location/site of testing Cannot go below manufacturer s recommendation for QC frequency Cannot use manufacturer s data or template IQCP alone must contain your data and match your process Must have data to support frequency of external controls IQCP must be incorporated into the overall lab Quality Management Plan (QMP) Medical Director must review, approve, and sign the initial IQCP The Medical Director, or designee, must review and sign the IQCP annual review There must be evidence of review and evaluation whenever there is a change in any of the components of the plan Multidisciplinary teams are necessary to build a robust plan Laboratories implementing IQCP for new tests are encouraged to perform control procedures more frequently during initial implementation, allowing the laboratory to identify performance issues that could indicate a need to adjust the QCP American Proficiency Institute 2018 1 st Test Event 11

RA: Risk Assessment QCP: Quality Control Plan QA: Quality Assurance 5 Components 3 Phases Gather relevant information Perform the analysis Identify hazards and prioritize the importance Evaluate the risk, decide on and implement mitigation Use the prioritized RA information, develop criteria for the plan Identify frequency and critieria for each element of the plan Review and approval by Medical Director Monitor testing specimens, reagents, environment, personnel, and test system Evaluate after any failures or changes to any of the componenets of the test References 1. Centers for Medicare & Medicaid Services. Developing an IQCP A Step-By-Step Guide https://www.cms.gov/regulations-and-guidance/legislation/clia/downloads/iqcp-workbook.pdf. Accessed February 10, 2018. 2. Clinical and Laboratory Standards Institute. EP23-A Laboratory Quality Control Based on Risk Management: Approved Guideline. Wayne, PA: Clinical Laboratory Standards Institute; 2011. CLSI document EP23-A. 3. Centers for Disease Control and Prevention. Clinical Laboratory Improvement Amendments (CLIA): Individualized Quality Control Plan. https://wwwn.cdc.gov/clia/resources/iqcp/. Accessed February 10, 2018. 4. Joint Commission website. https://www.jointcommission.org/. Accessed February 10, 2018. 5. Centers for Medicare & Medicaid Services. Individualized Quality Control Plan. https://www.cms.gov/regulations-andguidance/legislation/clia/individualized_quality_control_plan_iqcp.html. Accessed February 10, 2018. 6. College of American Pathologists website. http://www.cap.org/web/home. Accessed February 10, 2018. American Proficiency Institute 2018 1 st Test Event 12

7. IQCP from COLA. http://www.cola.org/programs-services/iqcp-program/. Accessed February 10, 2018. 8. CLIA Brochure 12. https://www.cms.gov/regulations-and- Guidance/Legislation/CLIA/Downloads/CLIAbrochure12.pdf. Accessed February 10, 2018. 9. CLIA Brochure 13. https://www.cms.gov/regulations-and- Guidance/Legislation/CLIA/Downloads/CLIAbrochure13.pdf. Accessed February 10, 2018. 10. Tague NT. The Quality Toolbox. 2nd ed. Milwaukee, WI: ASQ Quality Press; 2005. ASCP 2018 American Proficiency Institute 2018 1 st Test Event 13