Guidance for Industry and FDA Staff Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use

Similar documents
The Mammography Quality Standards Act Final Regulations Motion of Tube-Image Receptor Assembly

Interactive Review for Medical Device Submissions: 510(k)s, Original PMAs, PMA Supplements, Original BLAs, and BLA Supplements

Contains Nonbinding Recommendations. Draft Not for Implementation

The Mammography Quality Standards Act Final Regulations Quality Assurance Documentation

DRAFT GUIDANCE This guidance document is being distributed for comment purposes only. Document issued on: August 5, 2008

Draft Guidance for Industry on Part 11, Electronic Records, Electronic. Signatures Scope and Application; Availability of Draft Guidance and

Document issued on: July 8, 2010

Chapter 4732 Modifications Summary SEPTEMBER 30, 2016

Standards and Medical Device Regulation Roundtable. Seoul South Korea. 21 October ASTM International

Guidance for Industry and Food and Drug Administration Staff

Review of Existing Center for Drug Evaluation and Research Regulatory and Information

RADIATION POLICY Page 1 of 5 Reviewed: August 2017

Contains Nonbinding Recommendations

RADIATION PROTECTION

Notification Form for. Veterinary Clinics having. X-Ray Equipment

Q11 Development and Manufacture of Drug Substances--Questions and Answers

Requests for Feedback and Meetings for Medical Device Submissions: The Q-Submission

King Abdulaziz University Faculty of Dentistry. Radiology Policy & Procedure Clinical Manual

Guidance for Industry

PLAN OF ACTION FOR IMPLEMENTATION OF 510(K) AND SCIENCE RECOMMENDATIONS

Radiologic technologists take x rays and administer nonradioactive materials into patients bloodstreams for diagnostic purposes.

May 8, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane, Room 1061 Rockville, MD

10/8/12. Radiation Reduction and Monitoring Program: What the FDA and Other Regulatory Agencies Want. Disclosures. Two Principles of Radiation Safety

Compliance Guidance for DENTAL CONE BEAM COMPUTED TOMOGRAPHY (CBCT) QUALITY ASSURANCE MANUAL (1st Edition)

DRAFT REGULATORY GUIDE

Guidance for Industry

RADIATION PROTECTION PROGRAM FOR USE OF RADIATION GENERATING MACHINES IN THE HEALING ARTS, RESEARCH AND EDUCATION

Peninsula Dental Social Enterprise (PDSE)

The Mammography Quality Standards Act Final Regulations Document #2

Order no. 173/2003 of16/10/2003

Submitted electronically via: May 20, 2015

November 7, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane, Room 1061 Rockville, MD 20852

Inspections, Compliance, Enforcement, and Criminal Investigations

United States. Guidance for Security Screening of Humans Utilizing Ionizing Radiation

Health Canada Safety Codes

105 CMR: DEPARTMENT OF PUBLIC HEALTH

National Radiation Safety Committee, HSE

Mandatory Licensure for Radiologic Personnel. Christopher Jason Tien

Organizations in Nuclear Medicine Part IV- Others of Importance. Bennett S. Greenspan, MD SNM MWM Orlando, FL January 27, 2012

University of Pennsylvania Environmental Health and Radiation Safety. Diagnostic Energized Equipment Radiation Safety Manual

(2) Ensure measures are established to control health and safety hazards from ionizing radiation sources and radioactive material.

National Council on Radiation Protection and Measurements Homeland Security Recommendations Related to Nuclear and Radiological Terrorism

Allied Health Department. Radiation Protection Program (RPP) Policies & Procedures

DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service 3%3&4

The Mammography Quality Standards Act Final Regulations - Preparing For MQSA Inspections

RADIATION PROTECTION PROGRAM GUIDANCE

Michigan Department of Licensing and Regulatory Affairs Part 15 Computed Tomography Installations Guidance for CT Rules

HEALTH P H Y S I C S SOCIETY

AAPM Responds to Follow up Questions from Congress after Hearing on Radiation in Medicine

REGULATORY GUIDE 4.3 TEXAS DEPARTMENT OF STATE HEALTH SERVICES RADIATION SAFETY LICENSING BRANCH (RSLB) P.O. Box Austin, Texas

Guidance for Industry ANDA Submissions Prior Approval Supplements Under GDUFA

REVISION: This revised Management Directive (MD) updates TSA MD , dated January 29, 2004.

Guidance for Institutional Review Boards, Clinical Investigators, and Sponsors. Exception from Informed Consent Requirements for Emergency Research

RADIATION EXPOSURE CONTROL PLAN

Medical Device Reporting for Manufacturers

Doing Business As name (if applicable): 2. Mailing Address: (Street Address/City/State/Zip) 3. Physical Location: (Street Address/City/State/Zip)

The Practice Standards for Medical Imaging and Radiation Therapy. Limited X-Ray Machine Operator Practice Standards

University of Maryland Baltimore. Radiation Safety Procedure

Specific Comments on Proposed Amendments

UNIVERSITY OF VICTORIA X-RAY EQUIPMENT SAFETY POLICIES AND PROCEDURES

September 16 th, Dockets Management Branch (HFA-305) Food and Drug Administration 5630 Fishers Lane, Rm Rockville, MD 20852

Changes to QSR. The table below provides a history of changes to FDA s Quality System Regulation (QSR)

Local Government Records Control Schedule

CENTRAL SERVICE (CS) TECHNICIANS PERFORM MANY IMPORTANT

Radiation Safety Audit Checklist

Complaint Handling and Medical Device Reporting (MDRs)

Training in radiological protection for interventional cardiologists. Initial European experience.

HEALTH P H Y S I C S SOCIETY

September 1, Comments on Draft Reportable Food Registry (RFR) Guidance; Docket No. FDA-2009-D-0260

The Practice Standards for Medical Imaging and Radiation Therapy. Radiography Practice Standards

WARNING LETTER. an both of which were sponsored by. (formerly ). The products

RADIATION PROTECTION PROGRAM GUIDANCE

Implementation of a radiation protection framework for medical and dental x-ray diagnostic services in Minas Gerais/Brazil

University of Cincinnati

University of Maryland Baltimore. Radiation Safety Procedure

Re: Docket No. FDA 2013-N-0500 Proposed Rule: Supplemental Applications Proposing Labeling Changes for Approved Drugs and Biological Products

Radiologic Technology Program. Radiation Safety and Protection Program

We are very excited to provide this update for your Radiation for Dental Safety Manual.

Department of Defense INSTRUCTION

The Practice Standards for Medical Imaging and Radiation Therapy. Quality Management Practice Standards

MONITORING HEALTH CARE INDUSTRY REPRESENTATIVES EXPOSED TO IONIZING RADIATION

STANDARD ITEM. 1.1 Title: Prevention of Radiographic-Inspection Ionizing-Radiation Hazard; accomplish

University of Cincinnati

NAVSEA STANDARD ITEM. 1.1 Title: Prevention of Radiographic-Inspection Ionizing-Radiation Hazards; accomplish

Safety Zone, Barrel Recovery, Lake Superior; Duluth, MN. SUMMARY: The Coast Guard is establishing a temporary safety zone

April 17, Edition of the Joint Commission International Accreditation. SUBJECT: MITA Feedback on the 5 th Standards for Hospitals

HERCA Position Paper. Justification of Individual Medical Exposures for Diagnosis

Fifty Shades Of Gray A Medical Physicists Guide as RSO. Kevin Nelson, Ph.D, CHP Mayo Clinic Florida

Standards of Practice, College of Medical Radiation Technologists of Ontario

University of Maryland Baltimore. Radiation Safety Procedure

QA offers significant economic benefits!

Massey University Radiation Safety Plan Version

Effective Date: 6/15/77. Date Reviewed:

THE UNIVERSITY OF AKRON

Drug Safety and Risk Management Advisory Committee and the Anesthetic and Analgesic Drug

The Mammography Quality Standards Act Final Regulations - Preparing For MQSA

DOE N /29/95 Expires: [1 year after initiation]

The Practice Standards for Medical Imaging and Radiation Therapy. Medical Dosimetry Practice Standards

MISSISSIPPI LEGISLATURE REGULAR SESSION 2013

The College of Radiographers

Transcription:

Guidance for Industry and FDA Staff Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use Document issued on December 24, 2008 For questions regarding this document contact CDR Sean Boyd at 240-276-3287. U.S. Department of Health and Human Services Food and Drug Administration Center for Devices and Radiological Health Diagnostic Devices Branch Division of Mammography Quality and Radiation Programs Office of Communication, Education, and Radiation Programs

Public Comment Contains Nonbinding Recommendations Preface Written comments and suggestions may be submitted at any time for Agency consideration to the Division of Dockets Management, Food and Drug Administration, 5630 Fishers Lane, Room 1061, (HFA-305), Rockville, MD, 20852. When submitting comments, please refer to the exact title of this guidance document. Comments may not be acted upon by the Agency until the document is next revised or updated. Additional Copies Additional copies are available from the Internet at: http://www.fda.gov/cdrh/ocer/guidance/1680.pdf. You may also send an e-mail request to dsmica@fda.hhs.gov to receive an electronic copy of the guidance or send a fax request to 240-276-3151 to receive a hard copy. Please use the document number (1680) to identify the guidance you are requesting.

Guidance for Industry and FDA Staff Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use This guidance represents the Food and Drug Administration's (FDA's) current thinking on this topic. It does not create or confer any rights for or on any person and does not operate to bind FDA or the public. You can use an alternative approach if the approach satisfies the requirements of the applicable statutes and regulations. If you want to discuss an alternative approach, contact the FDA staff responsible for implementing this guidance. If you cannot identify the appropriate FDA staff, call the appropriate number listed on the title page of this guidance. Introduction The intent of this guidance is to advise manufacturers and FDA staff about safety procedures and recommendations that should be provided to the end user to promote safe use of handheld x-ray equipment. In general, there are a number of regulations, consensus safety standards, and radiation protection guidelines governing the performance and use of diagnostic x-ray equipment. Diagnostic x-ray equipment manufactured for the US market must meet requirements described in Title 21 of the Code of Federal Regulations, sections 1020.30 through, which require manufacturers to provide users with operational and safety information as well as to meet standards for equipment performance. (1) Internationally, radiation safety standards are provided by the International Electrotechnical Commission (IEC) in its collateral standard 60601-1-3. (2) Voluntary guidelines, particularly relevant to handheld equipment, have been provided by the National Council on Radiation Protection and Measurement (NCRP) for radiation protection in dentistry. (3) Many states regulate the use of x-ray equipment based on the Suggested State Regulations for the Control of Radiation, published by the Conference of Radiation Control Program Directors (CRCPD). (4) FDA's guidance documents, including this guidance, do not establish legally enforceable responsibilities. Instead, guidances describe the Agency's current thinking on a topic and should be viewed only as recommendations, unless specific regulatory or statutory 1

requirements are cited. The use of the word should in Agency guidances means that something is suggested or recommended, but not required. The Least Burdensome Approach We believe we should consider the least burdensome approach in all areas of medical device and electronic product regulation. This guidance reflects our careful review of the relevant scientific and legal requirements and what we believe is the least burdensome way for you to comply with those requirements. However, if you believe that an alternative approach would be less burdensome, please contact us so we can consider your point of view. You may send your written comments to the contact person listed in the preface to this guidance or to the CDRH Ombudsman. Comprehensive information on CDRH's Ombudsman, including ways to contact him, can be found on the Internet at http://www.fda.gov/cdrh/ombudsman/. 1. Issue There are two basic sources of potential x-ray exposure to operators that should be addressed during the use of diagnostic hand-held x-ray equipment: Leakage radiation transmitted through the equipment housing and shielding that surrounds the source assembly Backscatter radiation from the patient and nearby structures. 2. Guidance A primary consideration in any radiographic procedure is to reduce the dose to the patient as well as to the operator as much as possible while still achieving the diagnostic goals. This guidance briefly discusses the hazards unique to hand-held medical x-ray equipment, clarifies applicable requirements, and makes recommendations for safe use of the equipment. What are the exposure concerns with handheld x-ray systems? The federal performance standard for diagnostic x-ray systems and their major components did not anticipate hand-held x-ray systems at the time it was written. It therefore does not address system performance attributes or protective measures in light of the operator s proximity to the source assembly and the patient. Because of such proximity, hand-held x-ray systems pose increased exposure concerns due to leakage radiation and backscatter radiation. The radiation exposure to the operator will be the sum of any radiation leaking from the x- ray tube source assembly (leakage radiation) and any radiation that scatters from the patient or any objects in the room that are in the x-ray field (backscattered radiation.) What limits are placed on leakage and backscatter radiation? 2

The applicable performance standard for a diagnostic source assembly, under 21 CFR 1020.30(k), limits radiation leakage from the x-ray source assembly to an air kerma of 0.88 mgy (corresponding to an exposure of 100 mr) in one hour at a distance of one meter from the x-ray source when measured as specified in the standard. How can leakage and backscatter radiation to the operator be reduced? Shield the unit housing as required by the federal standard. Leakage radiation is of particular concern because the operator may be holding the x-ray source assembly. Increasing the distance from the source may not be possible. Hand-held x-ray equipment should incorporate into its design physical means to protect the operator from leakage radiation with sufficient shielding surrounding the source assembly to ensure compliance with 21 CFR 1020.30(k). Provide external shielding or a means to increase distance between the operator and the unit. To reduce operator exposure to backscatter radiation, manufacturers of hand-held x-ray equipment should incorporate physical means to provide operator shielding or allow operation of the unit at a distance. For example, an external, physical shield may be supplied with the equipment to be placed between the operator and the patient to attenuate backscatter radiation. Additionally, you may provide a means to initiate x-ray exposure using an equipment stand and remote switch from a distance to increase distance between the operator and the unit during exposure. Any product features designed into the equipment to protect the operator from radiation exposure should be described in user information with instructions for their proper use. Measure typical exposures near and around the unit. Because operators will typically be closer than 1 m to the unit, exposure measurements at distances closer than those required by the Federal standard become important. Manufacturers of hand-held x-ray units should make and report exposure measurements at distances near the unit, especially representative of positioning where the operator will actually be in contact with or in close proximity to the unit during use. For example, exposures at the location of handgrips and control devices are of particular interest to the operator of the unit. These data may be used to estimate an operator s whole-body or partial-body radiation dose for a single exposure or over an extended period of typical use. The data may be compared to occupational exposure limits to ensure limits are not exceeded and allow users to implement appropriate control measures based on an actual exposure assessment. Identify necessary safety precautions as required by the Federal standard. Under 21 CFR 1020.30(h)(1)(i) manufacturers must provide adequate instructions concerning radiological safety procedures and precautions which may be necessary 3

because of unique features of the equipment. A unit that is hand-held constitutes a unique feature that requires manufacturers to describe any safety procedures and precautions necessary to reduce operator exposure. Safety precautions may include designating a significant zone of occupancy and providing exposure information near the unit as described in the IEC standard. (2) Information on exposures near and around a hand-held unit should be included with operator instructions to identify the unique exposure hazards posed by the equipment. Providing exposure information will support any manufacturer s precautions and allow the operator to develop additional safe working practices to limit unnecessary exposure during equipment use. Precautions may include wearing appropriate personnel protective equipment, such as a personnel exposure monitoring device or lead-lined gloves and gowns. Procedural controls may be recommended to limit the number of exposures initiated by a single operator over a period or time, or to limit use of the hand-held unit without an equipment stand or remote switch in such circumstances only as necessary. Manufacturers may also include information describing exposure hazards and radiation safety precautions appropriate for patients and other individuals present in the area during examination (e.g., a person holding a child or other personnel assisting with procedures). 3. Getting More Information For additional information on state requirements applicable to hand-held x-ray systems, refer to your state s radiation control program regulations. The CRCPD website has a list of contacts for each state (http://www.crcpd.org/links.asp). Please contact your state officials on this subject. Another resource to consider is the Suggested State Regulations for the Control of Radiation. Some state programs do not allow the use of hand-held x-ray equipment. If you have any questions about this guidance, contact CDR Sean Boyd in the Division of Mammography Quality and Radiation Programs, Office of Communication, Education and Radiation Programs, Center for Devices and Radiological Health (HFZ-240), Food and Drug Administration, 1350 Piccard Drive, Rockville, Maryland 20850 at 240-276-3332 or sean.boyd@fda.hhs.gov. 4. References 1. Code of Federal Regulations, Title 21 (Food and Drugs), Part 1020 (Performance Standards for Ionizing Radiation Emitting Products), Sections 1020.30-1020.33. 4

2. International Standard IEC 60601-1-3, Medical electrical equipment Part 1-3: General requirements for basic safety and essential performance Collateral Standard: Radiation protection in diagnostic X-ray equipment, Edition 2.0, (International Electrotechnical Commission, Geneva, 2008). 3. Radiation Protection in Dentistry, NCRP Report no. 145, (National Council on Radiation Protection and Measurements, Bethesda, MD, 2003). 4. Suggested State Regulations for the Control of Radiation, http://www.crcpd.org/publications.asp#ssrcr,(conference of Radiation Control Program Directors, 2008). 5