May 9, Leslie Kux Associate Commissioner for Policy U.S. Food and Drug Administration 5630 Fishers Lane, Rm Rockville, MD 20852

Similar documents
2016 Sterilization Standards Update

Endoscope Reprocessing

Reprocessing of Flexible Endoscopic Instruments

August 15, Dear Mr. Slavitt:

Infection Prevention. Fundamentals of. March 21-23, 2017 Oregon Medical Association Portland, OR. oregonpatientsafety.org

Implementing Technologies for the Reduction of Hospital Acquired Infections. AHCA 2015 October 12, 2015 / 1:00 2:00 pm

Objectives. IPC Open calls - bi-weekly series. Introduction to Infection Prevention & Control (IPC) Open Call Series

Re-use of Single Use Devices Implications for Hospitals

New SGNA Standards Call for Expanded Infection Prevention Efforts in GI Endoscopy

Speaker Declarations

10/18/2010. Disclosure. Learning Objectives. Components of an Effective Infection Control Program

Frequently Asked Questions NONTUBERCULOUS MYCOBACTERIUM (NTM) INFECTIONS IN PATIENTS WHO HAVE HAD OPEN HEART SURGERY WITH CARDIOPULMONARY BYPASS

This course presents the applications of sterile processing theory in the clinical setting.

1.2 billion ambulatory care visits in US: physician offices, outpatient hospital and ED

Sterile Processing in Healthcare Facilities

INFECTION C ONTROL CONTROL CONTROL EDUCATION PROGRAM

August 28, Dear Ms. Tavenner:

Infections Associated with Heater Cooler Units Used in Cardiopulmonary Bypass and ECMO

SBAR: Use of gloves for environmental cleaning

APIC Fellow. In May 2016, I was notified that I had been selected to be in the. Becoming an

Healthcare-Associated Infections

Executive Summary. IPC Annual Report to Alberta Health 1

4/30/2012. Disclosure. Housekeeping. The Role of the Infection Preventionist on the Value Analysis Committee. Boyd Wilson

CENTRAL SERVICE (CS) PROFESSIONALS REQUIRE SIGNIFICANT

AS/NZS 4187:2003 AS/NZS

IC CONTACT and CONTACT PLUS PRECAUTIONS REV. JULY 2017

QUESTIONS PERTINENT TO PRODUCT SELECTION:

Ambulatory Surgical Center (ASC) INFECTION CONTROL SURVEYOR WORKSHEET

2014 Partnership in Prevention Award. November 21, :00-1:00PM EST. Introduction

DIRTY SCOPES: What You Need to Know About the New Reprocessing Guidelines and Infection Risk

INFECTION CONTROL SURVEYOR WORKSHEET

Multi-Link. X2 ECG System. Choose flexibility and quality for your reusable and disposable single-patient-use leadwires

Key Scientific Publications

EVIDENCE FOR PRACTICE. Evidence Appraisal Score: II A

9/14/2017. Best Practices in Instrument Cleaning. Objectives. Healthcare-associated Infections

Worksheet: Friend, Foe or Both?

Commonwealth Nurses Federation. A Safe Patient. Jill ILIFFE Executive Secretary. Commonwealth Nurses Federation

ENDOSCOPY MICROBIOLOGY ALERTS PROCEDURE STANDARD OPERATING PROCEDURE

CENTRAL SERVICE (CS) IS A VITAL DEPARTMENT IN ANY HOSPITAL

Objectives. Hot Topics in Infection Prevention and Control in Post Acute Care Settings. NADONA Infection Prevention and Control Webinar Series

National Standards for the prevention and control of healthcare-associated infections in acute healthcare services.

Outbreak Management it takes a village. November 20, 2014 Linda Stein Marge Gribogiannis

INSTRUMENT CLEANING HAS BECOME A TOPIC OF INTEREST IN

UW HEALTH JOB DESCRIPTION

THE INFECTION CONTROL STAFF

Accreditation Program: Hospital

Surgical Instrumentation: Eliminating Chaos. The Complex Process of Surgical Instrument Maintenance and Improving the Healthcare Environment

How to Add an Annual Facility Survey

Recommendation II. Recommendation I. Who s on Your Team? Recommendation III

OR staffing supports the provision of safe perioperative patient care and promotes a safe perioperative environment

Part II Quality improvement in long-term care: Partnership of infection prevention and environmental services

Orientation Program for Infection Control Professionals

Infection Prevention Checklist Section I: Policies and Practices I.1 Administrative Measures

Duodenoscope Culture Methods Update

The Danger Within. Confronting the Challenge of Healthcare-Associated Infections

CENTRAL SERVICE (CS) TECHNICIANS PERFORM MANY IMPORTANT

HIQA s monitoring programme - National Standards for the Prevention and Control of Healthcare. theatre findings Katrina Sugrue Inspector HIQA

Of Critical Importance: Infection Prevention Strategies for Environmental Management of the CSSD. Study Points

Erlanger Infection Control Program. Resident Resident Orientation and. and

Joint Commission Resources Quality & Safety Network (JCRQSN) Resource Guide. Complying with the Most Challenging Joint Commission Standards

CMS REQUIREMENTS: ESSENTIAL ELEMENTS FOR ASCS

June 24, Dear Ms. Tavenner:

Infection Control in General Practice

Principles of Infection Prevention and Control

EAST CAROLINA UNIVERSITY INFECTION CONTROL POLICY

HOSPITAL EPIDEMIOLOGY AND INFECTION CONTROL: SURGICAL SITE INFECTION REPORTING TO CALIFORNIA DEPARTMENT OF PUBLIC HEALTH

EAST CAROLINA UNIVERSITY INFECTION CONTROL POLICY

STANDARD PRECAUTIONS POLICY Page 1 of 8 Reviewed: May 2017

Infection Control Manual. Table of Contents

Bloodborne Pathogens Cumru Township Fire Department 02/10/2011 Policy 10.5 Page: 1 of 7

Infection Control and Prevention On-site Review Tool Hospitals

Risk Assessment in the Sterile Processing Department: It s not what you know but what you don t

Infection Prevention and Control (IPC) Elements of an Effective Program

Hospital-Acquired Infections Prevention is in Your Hands. Rachel L. Stricof

APPLICATION. Thank you for your interest in applying for the APIC Program of Distinction.

Formal Interpretations Guidelines for Design and Construction of Hospitals and Outpatient Facilities, 2014 edition

The stand alone central sterilising supply

Performance improvement project in the Sterile Processing Department at a university hospital in Egypt: redesign to centralize the process

PROCESS IMPROVEMENT AND ENHANCED QUALITY CARE ARE THE

Prevention and Control of Carbapenem Resistant Enterobacteriaceae Infections

Association for Professionals in Infection Control and Epidemiology, Inc. (APIC) Vision: Healthcare without infection

An act to add Sections and to the Health and Safety Code, relating to health.

Lightning Overview: Infection Control

Infection Prevention Implementation and adherence to infection prevention practices are the keys to preventing the transmission of infectious diseases

Qmentum Program. Infection Prevention and Control Standards STANDARDS. For Surveys Starting After: January 01, Accredited by ISQua

CDPH HAI Program Overview

Call for Abstracts. The body of the abstract will be typed directly into the online submission form.

July 10, reduce the risk of staff or patient airborne exposure to communicable diseases during surgical procedures (See Appendix A) and

Kristi Felix RN, BSN, CRRN, CIC, FAPIC Infection Prevention Coordinator Madonna Rehabilitation Hospitals

September 16 th, Dockets Management Branch (HFA-305) Food and Drug Administration 5630 Fishers Lane, Rm Rockville, MD 20852

Standard of Infection Prevention in the Gastroenterology Setting

Cleaning and disinfecting patient care equipment is an important infection prevention strategy for patients receiving care in the home.

Endoscopy Center Design and Practice

42 CFR Infection Control

Serratia infections in NICUs

NSQHS Standard 3: How are we going?

NOSOCOMIAL INFECTION : NURSES ROLE IN MINIMIZING TRANSMISSION

Preventing Healthcare-Associated Infections. Luebbert Chinnes. Peggy Prinz Luebbert, MS, MT (ASCP), CIC, CHSP Libby F. Chinnes, RN, BSN, CIC

Number: Ratio of the airflow to the space volume per unit time, usually expressed as the number of air changes per hour.

Personal Hygiene & Protective Equipment. NEO111 M. Jorgenson, RN BSN

Transcription:

1275 K Street, NW, Suite 1000 Washington, DC 20005-4006 Phone: 202/789-1890 Fax: 202/789-1899 apicinfo@apic.org www.apic.org May 9, 2016 Leslie Kux Associate Commissioner for Policy U.S. Food and Drug Administration 5630 Fishers Lane, Rm. 1061 Rockville, MD 20852 Re: Docket No. FDA-2016-N-0436; Refurbishing, Reconditioning, Rebuilding, Remarketing, Remanufacturing, and Servicing of Medical Devices Performed by Third-Party Entities and Original Equipment Manufacturers: Request for Comments. Dear Ms. Kux: The Association for Professionals in Infection Control and Epidemiology (APIC) appreciates the opportunity to provide information and comments on Refurbishing, Reconditioning, Rebuilding, Remarketing, Remanufacturing, and Servicing of Medical Devices Performed by Third-Party Entities and Original Equipment Manufacturers. APIC is a nonprofit, multidisciplinary organization representing over 15,000 infection preventionists whose mission is to create a safer world through prevention of infection. Our members are charged with the prevention of healthcare-associated infections and providing guidance and recommendations to individuals in our facilities regarding third party reprocessing. In order to protect patients from harm, APIC welcomes the opportunity to work with the Food and Drug Administration (FDA) to ensure that devices and equipment used for necessary medical procedures and care are as safe as possible and do not cause increased risk to patients. What follows are our comments on the issues for consideration. Proposed Definitions of Third-Party and OEM Activities APIC agrees with the definitions proposed by the FDA. We also suggest adding the term reprocess, which generally refers to all operations performed (e.g. to cleaning, disinfecting, sterilizing, and validation) to render a used reusable or single-use device patient-ready or to allow an unused product that has been opened to be made patient ready. APIC recommends that this term be defined as, the treatment of medical devices or equipment using validated technology and/or methods described by the original equipment manufacturer (OEM) in order to make them safe for reuse. Evaluation of Risk Associated With These Third-Party and OEM Activities 1. Who are the different stakeholders involved with the medical device activities listed previously? What are their respective roles? APIC identifies the following as stakeholders in the medical device activities listed previously: Patients: Patients often do not completely understand that an item once thought to be single use can be safely restored or refurbished to the OEM s original specifications. Additionally, for

reusable patient care equipment, patients often do not understand the complexity of the processes required to adequately clean, disinfect, and/or sterilize reusable patient care equipment. Patients expect to be cared for safely and trust that healthcare providers and manufacturers take steps to ensure safety and quality of devices and procedures utilized to provide care. Education of patients and the concerned general public regarding the definitions listed previously and the safety and efficacy of reprocessing is important in order to support patient-centered healthcare decisions. Licensed Independent Practitioners: Fit and function of the refurbished, reprocessed, or reconditioned device (such as a surgical instrument or reusable device) is an important concern for direct patient care stakeholders involved in these activities. Ensuring that reprocessed, refurbished, or reconditioned devices are returned to the OEM s original specifications is important to physicians. Concerns regarding perceived inferior quality of reprocessed, refurbished, or reconditioned devices are often voiced by physicians to other stakeholders during efforts to evaluate implementation of reprocessing initiatives in hospitals and other healthcare settings. Additionally, device availability and its impact on operating room or clinic turn-around time are important concerns for physician stakeholders involved in these medical device activities. Infection prevention and control: Patient safety and mitigation of the risk of transferring an infection from one patient to another patient via contaminated device(s) is of utmost concern for the infection preventionist stakeholder. Infection preventionists work to ensure that manufacturer instructions for use regarding care and cleaning of reusable patient devices and equipment are followed. Infection preventionists also work to ensure healthcare settings that contract with third parties or OEMs to reprocess, refurbish, repair, or remanufacture devices do so only if the FDA guidance (including required 510K clearances) for safety and efficacy are met. Sterile processing: As stakeholders, sterile processing professionals must follow all manufacturer instructions for use regarding care and cleaning of reusable patient care devices. Additionally, when devices demonstrate wear and tear, fail quality assurance testing, or are identified by users as demonstrating the need for repair, sterile processing staff are most commonly tasked with the responsibility for sending the device out for repair or refurbishing and returning the item into circulation once repair or refurbishing has been completed. Unlike nurses and physicians, certification and/or standardized education requirements for sterile processing professionals are not required in all states. Other stakeholder groups involved that are important to note include: Biomedical engineering Risk management and legal professionals affiliated with hospitals and health care settings Perioperative services Hospital Administrators and operational leadership

Manufacturers Third party reprocessing companies Instrument and equipment repair companies 2. What evidence exists regarding actual problems with the safety and/or performance of devices that result from these activities? Specific examples should be submitted. Many are already aware of the redesign of the endoscopic retrograde cholangiopancreatography (ERCP) duodenoscopes which affected the ability of end users to effectively clean the scopes, resulting in outbreaks at several institutions. Recent epidemiological evidence linked non-tuberculous mycobacterium infections and deaths due to the use of heater-cooler units during cardiac surgery in the United States and Europe. The mechanism of spread was proven by Sommerstein et al. 1 to be airborne transmission in the operating room. 3. What are the potential risks (patients/users) and failure modes (devices) introduced as a result of performing the previously defined activities on medical devices? Please speak to issues common to all devices as well as specific risks with specific devices. Failures involving reprocessing, reconditioning, or remanufacturing reusable patient care devices can result in catastrophic patient outcomes. Recent infections involving Carbapenem-resistant Enterobacteriaceae (CRE) transmitted by Olympus TJF-180V (ERCP) duodenoscopes highlight this. Changes made by the OEM to the scope resulted in hard to reach crevices that increased difficulty in cleaning. The reprocessing of reusable patient care devices is a complex process with many steps. This is especially true in the case of duodenoscopes and endoscopic ultrasound scopes, which are highly complicated devices that are exceedingly difficult to clean effectively. In a prospective multisite study it was noted that all steps in manual high level disinfection are completed only 1.4% of the time. 2 The same study reflected that all steps in high level disinfection with an automated endoscope reprocessor (AER) are completed only 75.4% of the time. 2 For these reasons, accepting a manufacturer s high level disinfection process that assumes 100% compliance with the process does not reflect the reality in the field, and an assumption that it will be followed puts patients at risk. 4. These activities are performed by OEMs and various third-party entities, including hospitals and humanitarian organizations. Are the risks different depending on who performs the previously mentioned activities? Key to risk mitigation is education of the individuals performing the tasks. They must have a clear understanding of the importance to detail in following the OEM guidance and raising concerns if they encounter difficulty complying with that guidance. Risk mitigation needs to occur with whoever is performing these activities, but a vulnerability for third-party entities is that to perform activities outside the OEMs guidelines may in some unpredictable manner alter the intended use and performance. 5. We are interested in knowing if these activities are more difficult or riskier to perform on certain devices versus others. Please cite specific examples in your response, along with an explanation of the source of this particular complexity.

Devices that must be high level disinfected pose a greater risk due to the complexity of the task. Endoscopes are classified as semi-critical items because they come into contact with an intact mucosal barrier that is naturally colonized with microorganisms and therefore not considered to be sterile. However this same exposure makes endoscopes far more likely to incur higher levels of contamination than items that are typically considered critical and require sterilization. In addition, devices which use water baths or water reservoirs pose a unique challenge as organisms known to contaminate water may become pathogens in the right environment and with the right host. Medical devices are becoming more and more complex and manufacturer instructions for cleaning are likewise more complex and often impractical or inadequate in the clinical setting. Manufacturers also test a limited number of products to clean and disinfect their devices, thereby posing challenges for end users. Using a different hospital-approved disinfectant can nullify the product warranty, requiring institutions to have several different cleaning products in stock, more complicated training for staff, and thereby challenges with staff compliance. 6. What information do third-party entities need in order to perform these activities in a way that results in safe and effective operation of the medical device? Please provide specific examples. Collaboration with manufacturers is essential to determine the appropriate guidelines. A validation approach is necessary to determine when the process has been successful at eliminating the risk of transmission of microorganisms and/or infection. The approach should take into consideration both the financial and resource cost to the institution. As an example, routine microbiological surveillance to ensure equipment is not contaminated can incur a significant cost to an institution that does not have the capability to do environmental sampling in-house. Manufacturers reprocessing instructions should state specific timetables for the reprocessing steps and avoid imprecise words such as immediate. For devices that have a defined number of times it can be reprocessed, simple mechanisms that clearly delineate to users and reprocessors when the limit is reached is necessary. Instructions for scopes should also include data to show what process timeframes resulted in successful reprocessing. In addition, manufacturers need to validate the total number of complete reprocessing cycles a scope should be expected to tolerate during its usable lifespan. For low level disinfection, manufacturers need to provide simple cleaning instructions with a broader range of cleaning products, including new technologies such as hydrogen peroxide. It may be helpful to understand which chemicals are compatible with a device instead of brand names of products. APIC appreciates the opportunity to share our views with the FDA, and we look forward to working with manufacturers, and third party reprocessors to ensure the safest possible care for patients. Sincerely, Susan Dolan, RN, MS, CIC 2016 APIC President

1 Sommerstein R, Rüegg C, Kohler P, Bloemberg G, Kuster SP, Sax H. Transmission of Mycobacterium chimaera from heater cooler units during cardiac surgery despite an ultraclean air ventilation system. Emerg Infect Dis. 2016 Jun Available at: http://wwwnc.cdc.gov/eid/article/22/6/16-0045_article. Accessed March 29, 2016. 2 Ofstead CL, Wetzler HP, Snyder AK, Horton RA. Endoscope reprocessing methods: a prospective study on the impact of human factors and automation. Gastroenterol Nurs 2010 Jul-Aug;33(4):304-11.