University of Pennsylvania Environmental Health and Radiation Safety. Diagnostic Energized Equipment Radiation Safety Manual

Similar documents
RADIATION PROTECTION

REGULATORY GUIDE 4.3 TEXAS DEPARTMENT OF STATE HEALTH SERVICES RADIATION SAFETY LICENSING BRANCH (RSLB) P.O. Box Austin, Texas

King Abdulaziz University Faculty of Dentistry. Radiology Policy & Procedure Clinical Manual

Allied Health Department. Radiation Protection Program (RPP) Policies & Procedures

Compliance Guidance for DENTAL CONE BEAM COMPUTED TOMOGRAPHY (CBCT) QUALITY ASSURANCE MANUAL (1st Edition)

University of Cincinnati

University of Cincinnati

RADIATION POLICY Page 1 of 5 Reviewed: August 2017

Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition)

Compliance Guidance for QUALITY ASSURANCE MANUAL (3 rd Edition)

RADIATION PROTECTION PROGRAM GUIDANCE

RADIATION PROTECTION PROGRAM GUIDANCE

2018 LACC Clinical Obligations & Grading System

RADIATION PROTECTION PROGRAM FOR USE OF RADIATION GENERATING MACHINES IN THE HEALING ARTS, RESEARCH AND EDUCATION

2 Quality Assurance In A Diagnostic Radiology Department. 1.1 Aim. 1.2 Introduction. 1.3 Key Elements of Quality assurance

RADIATION EXPOSURE CONTROL PLAN

Effective Date: 6/15/77. Date Reviewed:

Proper Care and Handling of Personnel Radiation Monitors W.L.R.& Associates

Competency Evaluation - Level III

Chapter 4732 Modifications Summary SEPTEMBER 30, 2016

DEPARTMENT OF HEALTH DIRECTORATE: RADIATION CONTROL CODE OF PRACTICE FOR USERS OF MEDICAL X-RAY EQUIPMENT

Revision Date June X-Ray Safety Program. Environmental Health and Safety Radiation Safety

Name: Date: Contact Information:

Radiologic Technology Program. Radiation Safety and Protection Program

RADIATION SAFETY: IS. E. Vano (Madrid/ES) Monday 28 th Sept :30 12:45 MY CATH LAB DOING ENOUGH? Radiation Protection Pavilion

Mission Statement. What we do 2/22/2016. Inspections, Digital Imaging, and Continuing Education, oh, my! What to expect when you are inspected

We are very excited to provide this update for your Radiation for Dental Safety Manual.

RADIATION PRODUCING MACHINES SAFETY MANUAL

MEDICAL UNIVERSITY OF SOUTH CAROLINA RADIATION MONITOR AND ITS USE

Michigan Department of Licensing and Regulatory Affairs Part 15 Computed Tomography Installations Guidance for CT Rules

THE UNIVERSITY OF AKRON

Radiation Protection Procedures for the Cincinnati Proton Therapy Center (CPTC)

Mandatory Licensure for Radiologic Personnel. Christopher Jason Tien

Walter L. Robinson & Associates Presents..

UNIVERSITY OF VICTORIA X-RAY EQUIPMENT SAFETY POLICIES AND PROCEDURES

Yale University ALARA (AS LOW AS REASONABLY ACHIEVABLE) PROGRAM

Radiologic technologists take x rays and administer nonradioactive materials into patients bloodstreams for diagnostic purposes.

Mobile X-Ray Machines

The Quality Management Examination

Radiation Safety Manual

STANDARD OPERATING PROCEDURE FOR COMPUTED TOMOGRAPHY (CT) ALBURY WODONGA HEALTH WODONGA CAMPUS

Implementing ALARA in the medical sector

Diagnostic X-Ray Imaging Quality Assurance: An Overview

Effective Date: 5-July Table of Contents

COMPETENCY BASED CLINICAL EDUCATION STANDARD

STANDARD OPERATING PROCEDURE FOR MAMMOGRAPHY EXAMINATIONS ALBURY WODONGA HEALTH WODONGA CAMPUS

Local Government Records Control Schedule

Doing Business As name (if applicable): 2. Mailing Address: (Street Address/City/State/Zip) 3. Physical Location: (Street Address/City/State/Zip)

RADIATION CONTROL - REGISTRATION OF RADIATION MACHINES, FACILITIES AND SERVICES REGISTRATION OF RADIATION MACHINES, FACILITIES AND SERVICES

Inspections and investigations.

Radiation Control Chapter RADIATION SAFETY REQUIREMENTS FOR INDUSTRIAL RADIOGRAPHIC OPERATIONS

Teaching Current Radiography Topics

XAVIER UNIVERSITY TASK ANALYSIS

COMMUNITY COLLEGE OF PHILADELPHIA

Radiation Producing Machines Safety Manual

10/8/12. Radiation Reduction and Monitoring Program: What the FDA and Other Regulatory Agencies Want. Disclosures. Two Principles of Radiation Safety

The Practice Standards for Medical Imaging and Radiation Therapy. Radiography Practice Standards

Radiation Safety Initial Training Module 3 Policies and Procedures

STANDARD ITEM. 1.1 Title: Prevention of Radiographic-Inspection Ionizing-Radiation Hazard; accomplish

Order no. 173/2003 of16/10/2003

RULES OF TENNESSEE BOARD OF MEDICAL EXAMINERS DIVISION OF HEALTH RELATED BOARDS

MONITORING HEALTH CARE INDUSTRY REPRESENTATIVES EXPOSED TO IONIZING RADIATION

The Practice Standards for Medical Imaging and Radiation Therapy. Computed Tomography Practice Standards

Notification Form for. Veterinary Clinics having. X-Ray Equipment

IR(ME)R Inspection (Announced) Abertawe Bro Morgannwg University Health Board Princess of Wales Hospital Radiology Department

RESPIRATORY PROTECTION PROGRAM

NAVSEA STANDARD ITEM. 1.1 Title: Prevention of Radiographic-Inspection Ionizing-Radiation Hazards; accomplish

The Practice Standards for Medical Imaging and Radiation Therapy. Limited X-Ray Machine Operator Practice Standards

University of Maryland Baltimore. Radiation Safety Procedure

University of Maryland Baltimore. Radiation Safety Procedure

NRC INSPECTION MANUAL

Radiation Dose Management Requirements from MACRA and Joint Commission, Potential Effects on Reimbursement

Standards of Practice, College of Medical Radiation Technologists of Ontario

UNIVERSITY OF SOUTH ALABAMA RADIATION SAFETY PROCEDURES MANUAL

Dose Limits. Trevor Boal Radiation Protection Unit RSM-NSRW

Our MISSION is to ensure that the benefits outweigh the risks for all medical radiation exposures in Africa.

MINNESOTA DEPARTMENT OF HEALTH

a. The Commanding Officer (CO) is responsible for maintaining an effective radiation safety program as outlined in reference (a) and shall:

Recommendations to Health Quality Ontario

Position No. Job Title Supervisor s Position RADIOLOGY TECHNOLOGIST Manager Medical Technology Systems

APEx Program Standards

RADIATION SAFETY MANUAL

Image Gently and Image Wisely. Priscilla F. Butler, MS, FAAPM, FACR Senior Director and Medical Physicist American College of Radiology

HERCA Position Paper. Justification of Individual Medical Exposures for Diagnosis

PROCEDURES FOR PERSONAL DOSIMETRY

MEASURES TO STRENGTHEN INTERNATIONAL CO-OPERATION IN NUCLEAR, RADIATION, TRANSPORT AND WASTE SAFETY

The Practice Standards for Medical Imaging and Radiation Therapy. Quality Management Practice Standards

Department of Defense INSTRUCTION. Occupational Ionizing Radiation Protection Program

Department of Defense INSTRUCTION

Radiation Licensure and Management (RS100) Course

RADIATION ONCOLOGY RESIDENCY SUPERVISION POLICY

U.S. Nuclear Regulatory Commission

COMPLIANCE WITH THIS PUBLICATION IS MANDATORY

Northumbria Healthcare NHS Foundation Trust. CT (Computerised Tomography) Issued by Radiology Department

RADIOLOGIC TECHNOLOGY PROGRAM STUDENT HANDBOOK. and POLICY MANUAL

Radiation Safety Code of Practice

The Practice Standards for Medical Imaging and Radiation Therapy. Radiation Therapy Practice Standards

1. Terms. For definition of the terms used in this instruction, see AFI , Air Force Nuclear Weapons Surety Program (formerly AFR 122-1).

Safely Working with Radioactive Materials

SCOPE OF PRACTICE FOR CANADIAN CERTIFIED MEDICAL PHYSICISTS

Transcription:

University of Pennsylvania Environmental Health and Radiation Safety Diagnostic Energized Equipment Radiation Safety Manual (Reviewed: September 2012) I. Proper Operating Procedures A. Radiographic Units B. Fluoroscopic Units C. Mobile Diagnostic Units D. Computed Tomography Units E. Dental units II. III. Techniques of External Radiation Protection Shielding A. Patient Shielding B. Patient Gonadal Shielding C. Personnel Shielding D. Structural Shielding IV. Pregnant Patient and Pregnant Worker Policy A. Patient B. Personnel V. Badging & Dosimetry Policy A. Personnel Dosimeter Policy B. Dosimeter Placement C. Dose Reports D. Pick-up & Drop-off of dosimeters E. Regulatory Limits VI. Quality Assurance Program A. Reject-repeat analysis program B. Film processors C. X-Ray Equipment D. Inspection of protective garments E. Response to reported problems F. Visual inspection PAGE i

VII. Diagnostic Equipment A. Registration of diagnostic units B. Acquisition of new diagnostic units C. Disposal of diagnostic units VIII. Important Phone Number PAGE ii

I. Proper Operating Procedures A. Proper Operating Procedures for Radiographic Units 1. Limit the primary x-ray beam to the smallest area possible consistent with the objectives of the clinical examination. 2. Align the x-ray beam properly with the patient and the CR or digital image receptor. 3. Remain behind a protective barrier (i.e., a leaded glass wall or a leaded door) during the entire radiographic exposure, unless adequate performance of the examination requires your presence. 4. Observe the patient during the exposure from the protected area. 5. No person should routinely hold patients during diagnostic examinations. When a patient must be held in position for radiography, a mechanical supporting or restraining device should be used. Pregnant women or persons under the age of 18 years should not be permitted to hold patients. If a patient must be held by someone, that individual shall be protected with appropriate shielding devices such as protective aprons and gloves. Positioning should be arranged so that no part of the holder's torso, even if covered by protective clothing, will be struck by the useful beam and so that the holder's body is as far as possible from the useful beam. [Reference: NCRP Report 102, 2.4(h)] 6. Provide protective garments (lead aprons and/or shielding) for all individuals whose presence in the room is necessary during the radiographic exposure. B. Proper Operating Procedures for Fluoroscopic Units 1. Only persons required for the fluoroscopic procedure should be in the room during the procedure. 2. As in a radiographic procedure, use the smallest possible beam area to reduce patient exposure and scatter radiation. 3. Perform visual observation of the alignment of the image intensifier or flat panel detector, x-ray tube, and the patient prior to the initiation of a fluoroscopy procedure. 4. Minimize fluoroscopic doses by reducing the fluoroscopic time used. Fluoroscopic time, of course, varies with different patients, the type of the examination, and the complexity of the clinical study. 5. Operators should use the timing device to indicate an audible preset time, which will serve as a reminder to keep procedures as short as possible. According to State regulations, the predetermined time may not exceed 5 minutes [PA 221.41a]. 6. Use the shortest possible distance from the image intensifier or flat panel detector to the patient. The Automatic Exposure Control (AEC) on the fluoroscope will automatically increase the radiation output of the fluoroscope when longer distances are used. 7. The fluoroscopist should wear a thyroid shield, leaded gloves, and glasses, as necessary, to reduce exposure to the thyroid, extremities, and eyes. 8. Uses of Low Dose and Pediatric modes are recommended when available.

9. Patient dose rates are reduced with the use of pulsed fluoroscopy. It is recommended that pulsed fluoroscopy be utilized when its use is consistent with the clinical objectives of the procedure. 10. If the operator has questions regarding the use of various operating modes, they may be directed to the Medical Physicist. Please contact (215) 898-7187 for further information. C. Proper Operating Procedures for Mobile Diagnostic Units If proper care is not taken, mobile equipment has a greater potential than standard diagnostic equipment for unnecessary radiation exposure of personnel and patients. 1. If possible, stand at least 2 meters (~6.5 ft.) away from the tube head and the patient. Distance is often the best possible protection from radiation. [PA 221.34a.(f)(2)] 2. It is important that only individuals necessary for the diagnostic examination be in the vicinity. Individuals who are required to remain in the room should wear protective clothing, or should be located behind a protective shield. 3. Wear protective garments of 0.5 mm lead equivalence when at a distance less than 120 cm (~4 ft.) from the useful beam, and a minimum of 0.25 mm lead equivalence at distances greater than 120 cm. [PA 221.11.(e)] D. Proper Operating Procedures for Computed Tomography Units 1. Use a scan plane position device, such as a light field or laser, to indicate directly or indirectly the position of the slice plane(s) on the patient within 2 mm. [NCRP 3.9.1(e)] 2. Observe the patient during the CT exposures either directly from the control area or by use of a viewing system. [PA221.203.(b)] 3. Only individuals whose presence is necessary should be in the CT x-ray room during exposures. All such individuals should be protected with leaded aprons and/or portable shields. 4. Operators should evaluate the size of the patient, and adjust the operating parameters (kvp, mas, etc.) to take the patient size into account. Dose modulation systems (i.e. CareDose 4D, Smart ma, Z-DOM, etc.) should be used when applicable to reduce unnecessary radiation to the patient. E. Proper Operating Procedures for Dental Units 1. Only persons required for a radiographic procedure should be in the radiographic room during exposure. All persons must be adequately protected by protective garments. [References: NCRP 35, 4.4; PA221.11.(e)]. 2. Align the x-ray beam and the film or image sensor very carefully with the area to be radiographed. 3. Stand behind protective barrier and observe the patient during the dental exposure. 4. Neither the operator nor the assistant shall hold the film or digital image receptor in place for the patient during the exposure. Use the film or digital image receptor holder devices during the exposure.

II. Techniques of External Radiation Protection Control radiation exposure levels via four basic methods: A. Maximize the distance from radiation source. B. Minimize the radiation exposure time. C. Shield the radiation source properly. D. Shield patients and personnel. III. Shielding A. Patient Shielding "Sensitive body organs (e.g., lens of eye, gonads) should be shielded whenever they are likely to be exposed to the useful beam provided that such shielding does not eliminate useful diagnostic information or proper treatment. Shielding should never be used as a substitute for beam collimation."[reference: NCRP 102, 2.2 (d)] B. Patient Gonadal Shielding Shield the gonads with at least 0.5 mm of lead equivalence during diagnostic procedures in which gonads are in the useful beam, except for cases in which this would interfere with the diagnostic procedure. C. Personnel Shielding Personnel who remain in the room during examinations must be protected by proper shielding. 1. All personnel in the room during an exposure should wear protective garments These garments are to be of 0.5 mm lead equivalence when at a distance less than 120 cm (~4 ft.) from the useful beam, and a minimum of 0.25 mm lead equivalence at distances greater than 120 cm. [PA 221.11.(e)] 2. Personnel who are likely to be exposed to high levels of scattered radiation to the thyroid during any procedure should wear thyroid shields as well. 3. Leaded glasses can greatly reduce the exposure of the eye lenses to scattered radiation in fluoroscopy, especially for physicians. 4. Any person who must have his or her hand near the primary beam (as in cases in which no other means are available to immobilize a patient) should wear leaded gloves to reduce exposure of the extremities. D. Structural Shielding Rooms which house stationary x-ray equipment have been designed with sufficient shielding in the walls to provide protection to anyone outside of the room. Do not tamper with the integrity of the shielded walls. If any personnel notice structural changes, such as holes drilled into walls, Environmental Health and Radiation Safety (EHRS) should be notified as soon as possible.

IV. Pregnant Patient and Pregnant Worker Policy A. Patient Special consideration must be given to the protection of the embryo or fetus of women known to be, or potentially, pregnant. A patient of childbearing age should be questioned to ascertain the likelihood of pregnancy. If the patient is found to be pregnant or likely to be pregnant, the physician or radiologist should be consulted to decide whether this radiation dose to the patient is justified. B. Personnel An employee should contact the EHRS either directly or through a supervisor when she knows or suspects that she is pregnant. If for personal reasons an employee does not wish to disclose pregnancy to her supervisor, confidential disclosures can be made directly to the EHRS. V. Badging and Dosimetry Policy A. Personnel Dosimeter Policy EHRS uses personnel monitoring to identify inadequate or improper radiation safety practices and potentially serious radiation exposure situations. EHRS will issue proper personnel dosimeters when evaluation of equipment reveals that the radiation dose to personnel could potentially be larger than ALARA limits per calendar quarter to the whole body (125 mrem). Radiation workers are monitored by EHRS. All radiation workers must conscientiously wear the radiation monitoring devices provided by the EHRS. These devices may include body and/or ring dosimeters. B. Dosimeter Placement Interpretation of the measured dose depends on the placement of the dosimeter. All personnel must wear their dosimeters correctly. The following list indicates where the dosimeters are to be worn: 1. Film or Luxel Body Badges are to be worn above any protective clothing at collar level. 2. Ring Dosimeters are to be worn so that the employee s name is facing the source of radiation. For x-ray equipment, the name would typically be facing out. Personnel must return all monitoring devices promptly at the end of each predetermined wear period, so that the radiation dose can be evaluated. Do not expose personnel monitoring devices to extreme heat or humidity. If any dosimeter has received a dose higher than the ALARA trigger level (Table 1), the employee will be notified and the reason for the high reading will be investigated. Measures will be taken to keep radiation doses below these trigger levels whenever possible: Table1. ALARA Investigation Trigger Levels Radiology Cardiac Catheterization Whole Body 125 mrem/calendar quarter 250 mrem/calendar quarter Extremities 5000 mrem/calendar quarter 5000 mrem/calendar quarter

C. Dose Reports EHRS sends dose summary reports for each wear period and on an annual basis. These reports will be available in the Radiology Manager s office. Personnel dosimetry information may also be obtained by contacting EHRS directly. D. Pick-up and Drop-off of Dosimeters EHRS delivers new dosimeters to each department on the last Wednesday of the final month of your wear period. Each group should have one person who is responsible for the distribution of dosimeters. The old dosimeters should be returned to the designated personnel. EHRS will pick these up by the first Wednesday of the following month for analysis. E. Regulatory Limits Federal and State regulations require radiation exposures of staff and members of the general public to be below certain regulatory limits. In practice, radiation exposures are only a fraction of these limits. These regulatory limits are displayed in Table 2. Table 2. Regulatory Limits on the Exposure of Staff and the General Public to Radiation. Staff General Public Whole Body 5000 mrem/year 100 mrem/year Lens of the Eye 15,000 mrem/year N/A Extremities & Skin 50,000 mrem/year N/A VI. Quality Assurance Program The quality assurance program will ensure that doses to patients are in accordance with the standards of good practice set forth by the American College of Radiology A. Reject-repeat analysis program The analysis of the rejected radiographs provides information about the different aspects of radiological imaging. B. Film processors Test each film processor daily for temperature, contrast, density, and speed. A graph of results of daily measurements will show deviation from normal behavior. C. X-Ray Equipment All equipment is tested annually by EHRS to ensure that equipment not only meets regulatory requirements, but also provides good image quality with an appropriate radiation dose to the patient. D. Inspection of protective garments Perform an annual inspection of the shielding garments, such as lead aprons and lead gloves, to ensure the integrity of these items.

E. Response to reported problems Make all repairs on the units as soon as possible. Send repair documentation to authorized personnel upon completion of the repairs. F. Visual inspection Report all conspicuous problems with energized equipment or with shielded rooms, as well as any other safety problems observed by personnel to EHRS immediately. VII. Diagnostic Equipment A. Registration of Diagnostic x-ray Units Pennsylvania regulations require that all radiation-producing equipment be registered with the Department of Environmental Protection. This registration is performed by EHRS on an annual basis. EHRS also maintains a listing of all units currently registered. B. Acquisition of New Diagnostic x-ray Units It is the responsibility of clinical personnel to notify the EHRS upon acquisition of any new diagnostic equipment. Authorized EHRS personnel will conduct a radiation safety survey on all new units prior to patient use. EHRS will also make the necessary shielding determinations for stationary x-ray producing units prior to their installation. C. Disposal or Transfer of Diagnostic x-ray Units Clinical personnel must notify the EHRS of any diagnostic equipment intended for disposal or transfer to another facility. EHRS will ensure that the proper notifications to State Agencies are made. VIII. Important Phone Numbers University of Pennsylvania, Environmental Health and Radiation Safety: Monday Friday during business hours (215) 898-7187 On-Call Physicist for assistance after hours (215) 573-6626 CHOP Radiology Physics and Engineering Director: Winnie Zhu (215) 590-4320