The Newcastle upon Tyne Hospitals NHS Foundation Trust. Decontamination of Healthcare Equipment following Patient Use and Prior to Service or Repair

Similar documents
Decontamination of Medical and Laboratory Equipment Prior to Maintenance or Transportation

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Animals on Hospital Premises Policy

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Water Safety Policy

The Newcastle upon Tyne NHS Hospitals Foundation Trust. Latex Operational Policy

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Ventilation Policy

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Patients Wills Policy

The Newcastle Upon Tyne Hospitals NHS Foundation Trust. Standard Precautions Policy

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Code of Practice for Wound Care Company Representatives and Staff with whom they interact

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Safe and Effective Use of Bedrails

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Strong Potassium Solutions Safe Handling and Storage

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Employment Policies and Procedures. Breastfeeding Supporting Staff Policy

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Implementation Policy for NICE Guidelines

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Key Control Operational Policy

The Newcastle Upon Tyne Hospitals NHS Foundation Trust. Use of Patients Own Drugs (PODs)

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Advice and Guidance on Workplace Temperatures for all Trust Employees

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Access to Drugs Policy

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Introduction and Development of New Clinical Interventional Procedures

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Named Key Worker for Cancer Patients Policy

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Procedure for Monitoring of Delayed Transfers of Care

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Patient Choice Directive Policy & Guidance

The Newcastle Upon Tyne Hospitals NHS Foundation Trust. Unlicensed Medicines Policy

The Newcastle upon Tyne Hospitals NHS Foundation Trust

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Injectable Medicines Policy

The Newcastle Upon Tyne Hospitals NHS Foundation Trust. Strategy for Non-Medical Prescribing

POLICY FOR TAKING BLOOD CULTURES

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Protected Mealtime Policy

Medical Devices Management Policy

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Strategy for the Prevention of Slips, Trips and Falls

Clinical staff undertaking Endoscopy and Nasendoscope interventions

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Central Alert System (CAS) Policy and Procedure

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Medicines Reconciliation Policy and Procedure for Adult and Paediatric Patients

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Pre-Operative Marking

Decontamination of equipment

The Newcastle upon Tyne Hospitals NHS Foundation Trust. First Aid Policy

NHS Lewisham CCG Health & Safety Policy

The Newcastle Upon Tyne Hospitals NHS Foundation Trust Radiation Safety Policy

Administration of urinary catheter maintenance solution by a carer

Medical Devices Management Policy

BLOOD AND BODILY FLUID GUIDELINES

The Newcastle upon Tyne Hospitals NHS Foundation Trust

your hospitals, your health, our priority CATHETERISATION Urethral/ supra-pubic POLICY NAME: VERSION NUMBER : 1 PROFESSIONAL ADVISORY BOARD (PAB)

The Newcastle upon Tyne Hospitals NHS Foundation Trust. Visitors Policy

WARD CLOSURE POLICY V

Bare Below the Elbow Supplementary Policy for Hand Hygiene

Health and Safety Policy

Other (please specify): Note: This document has been assessed for any equality, diversity or human rights implications

Diagnostic Testing Procedures in Urodynamics V3.0

A list of authorised referrers will be retained by the Colposcopy team and the Clinical Imaging Department.

Sharps Policy Safe Use and Disposal

Version: 2. Date adopted: 17 May publication: Review date: September Expiry date: March 2019

Health and Safety Policy and Arrangements

Date ratified November Review Date November This Policy supersedes the following document which must now be destroyed:

Infection Prevention and Control Policy

Health Care Support Worker. Job description

Shetland NHS Board Standard Operating Procedure for Cleaning, Maintenance, Audit and Replacement of Mattresses

Hepatitis B Immunisation procedure SOP

Clinical Guideline for Post-Operative Nausea and Vomiting 1. Aim/Purpose of this Guideline

Spillage of Blood and Other Body Fluids

Other (please specify): Note: This policy has been assessed for any equality, diversity or human rights implications

Decontamination of Medical Devices:

NHS GREATER GLASGOW & CLYDE CONTROL OF INFECTION COMMITTEE POLICY

Sharps Safety Policy

New Clinical Interventional Procedures Policy

Section G - Aseptic Technique. Version 5

Agenda item 3.3 Appendix 4 MANDATORY TRAINING POLICY

Clinical Guideline for Clinical Imaging Referral Protocol for Upper & Lower GI Non medical Endoscopist within RCHT. 1. Aim/Purpose of this Guideline

1.3 Referrer: in the context of this protocol the term referrer refers to a health care worker who is authorised to refer individuals for X-rays.

Wandsworth CCG. Continuing Healthcare Commissioning Policy

Hoist and Sling for Safer Patient Use Policy

Version: Date Adopted: 20 October Name of responsible Committee: Date issue for publication: Review Date: March 2018

Healthcare Associated Infection (HAI) inspection tool

HAZARDOUS SUBSTANCES POLICY Page 1 of 5 Reviewed: May 2017

CLINICAL GUIDELINE FOR THE USE OF RECTUS SHEATH CATHETERS IN CHILDREN. 1. Aim/Purpose of this Guideline

The Newcastle Upon Tyne Hospitals NHS Foundation Trust. Mandatory Training Policy

PROCEDURE FOR TAKING A WOUND SWAB

Infection Prevention and Control: Audit Policy

Policy for Handling the Spillage of Cytotoxic and Anti-Cancer Drugs

HANDLING OF LAUNDRY POLICY

Policy for the Control and Management of patients Colonised or Infected with Vancomycin resistant enterococci (VRE)

CORPORATE SAFETY MANUAL

Lincolnshire Partnership NHS Foundation Trust (LPFT) Title of Policy

Medical Devices Policy

Diagnostic Test Reporting & Acknowledgement Procedures. - Pathology & Clinical Imaging

Standard Precautions for Infection Control

Skin Care and the Management of Work Related Dermatitis

Unannounced Theatre Inspection Report

Policy for the Management of Safety Alerts issued via the Central Alerting System (CAS)

Occupational Health Surveillance Policy V2.1

GCP Training for Research Staff. Document Number: 005

NHS GREATER GLASGOW & CLYDE STANDARD OPERATING PROCEDURE (SOP)

Section F - Decontamination Policy

CLINICAL GUIDELINE FOR CLINICAL IMAGING REFERRAL PROTOCOL FOR NURSE SPECIALISTS IN HEART FUNCTION WITHIN RCHT Summary. Start

Executive Director of Nursing and Chief Operating Officer

Safe Bathing Policy V1.3

CLINICAL GUIDELINE FOR USE OF BED AND CHAIR SENSOR ALARM MATS FOR PREVENTING FALLS IN ADULT PATIENTS

Policy for the authorising of blood components by the Haematology Clinical Nurse Specialist V1.0

Isolation Care of Patients in Isolation due to Infection or Disease

ASEPTIC TECHNIQUE POLICY

Aseptic Non-Touch Technique Policy

This guideline is for nursing staff within the Pain Services assisting with the administration of botulinum toxin.

Transcription:

The Newcastle upon Tyne Hospitals NHS Foundation Trust Decontamination of Healthcare Equipment following Patient Use and Prior to Service or Repair Version No.: 5.0 Effective From: 27 December 2017 Expiry Date: 27 December 2020 Date Ratified: 14 December 2017 Ratified By: Infection Prevention and Control Committee 1 Introduction Safe systems of work must be implemented to protect all staff and patients from the transmission of organisms from medical devices, associated consumables and materials used in the treatment, diagnosis and care of patients. In order to comply with the duty required by the Health and Safety at Work Act 1974 and associated regulations, e.g. Control of Substances Hazardous to Health Regulations (2002), risk to health and safety must be reduced as far as reasonably practicable. It is essential that all medical devices and other healthcare equipment undergo an appropriate decontamination process following patient use and prior to inspection, service or repair on or off site. Contamination may be by microbiological or other biological agents; blood, body fluids or chemicals which are corrosive, irritant, toxic, cytotoxic or radioactive. A safe system of work, in compliance with MHRA: Managing Medical Devices will minimise the risk to personnel. As contamination is not always visible, the standard must be that all equipment is cleaned following each patient use and prior to inspection, service or repair. The Health Act (2008): Code of Practice on the prevention and control of infections and related guidance (revised 2015) clearly defines the responsibility of the organisation to protect, patients, staff and others from acquiring Health Care Associated Infections. The decontamination programme identified within the Act is reflected in this policy. 2 Policy scope This policy applies to all healthcare professionals working across acute and community services within NuTH. This includes medical staff, nurses, allied health professionals, students, temporary clinical staff, those working in the Trust from other organisations and contractors. 3 Aim of policy The aim of this policy is to ensure that all healthcare equipment is appropriately decontaminated to remove or minimise the risk of infection/other hazards prior to Page 1 of 8

transport of equipment for inspection, servicing or repair, either on hospital premises or elsewhere. 4 Duties 4.1 The Chief Executive has overall responsibility for implementation, monitoring and review of this policy. This responsibility is delegated to the Director of Infection Prevention and Control (DIPC). 4.2 The Infection Prevention and Control Committee (IPCC) will review the policy and any new evidence base within the time frame set out in the policy. 4.3 The Infection Prevention and Control (IPC) and Medical Electronics Team are responsible for providing expert advice in accordance with this policy and for supporting staff in its implementation. 4.4 Ward/Department Managers are responsible for ensuring that all staff are trained in the safe use and handling of medical devices within their area. Managers must ensure that staff are aware of how to effectively decontaminate equipment used within their department. Managers must therefore ensure that all staff comply with this policy. It is the responsibility of all staff to ensure equipment is stored clean and ready for use and effectively decontaminated prior to transfer for service or repair. 4.5 Manufacturers and suppliers have a responsibility to provide information on the compatibility of their particular medical devices or equipment with methods and agents for effective decontamination. Refer to manufacturers instructions for use. Guidance on decontamination has been issued by the MHRA: Managing Medical Devices. Such general guidance will require to be interpreted in the light of particular local situations. Where manufacturer s guidance differs from Trust Policy guidance must be sought from the IPC Team prior to purchase. 5 Definitions Medical devices - refers to all products, except medicines, used in healthcare for diagnosis, prevention, monitoring or treatment. The range of products is very wide and includes contact lenses, condoms, heart valves, hospital beds, resuscitators, radiotherapy machines, surgical instruments and syringes, wheelchairs and walking frames. 6 Decontamination of healthcare equipment 6.1 Classification of infection risk associated with the decontamination of medical devices and equipment The following guidance will determine whether equipment needs sterilisation, disinfection or cleaning following patient use and/or prior to inspection, service or repair. For further guidance on methods of sterilisation, disinfection or cleaning please refer to the Cleaning and Disinfection Procedure. Page 2 of 8

Classification of infection risk associated with the decontamination of medical devices (MHRA Managing Medical Devices 2015): Risk Application of Item Recommendation High In close contact with broken skin or mucous Cleaning followed by sterilisation. membrane Introduced into sterile body areas Medium In contact with mucous membranes contaminated with particularly virulent or readily transmissible organisms Prior to use on immunocompromised patients Cleaning followed by sterilisation or disinfection required. Where sterilisation may damage equipment, cleaning followed by high level disinfection may be used as an alternative*. Low In contact with healthy skin Not in contact with patient Cleaning only. Other factors to be taken into consideration when choosing a method of decontamination include the nature of the contamination, the time required for processing, the heat, pressure, moisture and chemical tolerance of the device, the availability of the processing equipment and the quality and risks associated with the decontamination. Note: * Where sterilisation will damage the equipment, cleaning followed by high-level disinfection may be used as an alternative. This must be in accordance with manufacturers guidance and advice sought from the Infection Prevention and Control Team. Potential contamination of Healthcare Equipment or Medical Devices with Prion Proteins (CJD/vCJD) will result in the need for decontamination, quarantining or disposal of the device. Refer to Control of Transmissible Spongiform Encephalopathies (TSEs), including Creutzfeldt-Jacob Disease (CJD), in hospital patients; contact the Infection Prevention and Control Team for advice. The Trust policy for ensuring a clean patient environment, Decontamination of Patient Environment (including Terminal Cleaning), provides further explicit guidance on cleaning responsibilities. The Decontamination of Reusable Medical Devices Policy focuses on reusable surgical instruments and flexible endoscopes. A comprehensive list of recommended methods of decontamination can be found in the Trust Cleaning and Disinfection Procedure; these lists are not exhaustive. Page 3 of 8

6.2 Procedure following patient use and when sending medical devices for service or repair Following use on a patient or when requiring inspection or service, all medical devices must be checked for visible evidence of contamination by the user/clinician; however, as contamination is not always visible, all equipment must be cleaned following patient use. Every attempt must be made by the user to adequately decontaminate the equipment prior to transfer for repair or servicing. If it is not possible to decontaminate, then the equipment must be safely contained and clearly identified as contaminated until advice is obtained from the Infection Prevention and Control Team and the Medical Electronics Department. All equipment MUST be accompanied by the Trust Declaration of Decontamination Status of Healthcare Equipment Following Use and Prior to Service or Repair form, (see Appendix 1). If items are dispatched to suppliers, or presented for service or inspection on the Trust s premises without a declaration of decontamination, the receiver will refuse to accept the item and it will be returned to the ward/department until it is accompanied by the aforementioned form. In some instances total decontamination may not be possible at source i.e. point of use, due to internal contamination of the equipment, requiring additional tools to gain access to the affected parts. The equipment must be removed to a suitable designated area for appropriate decontamination prior to inspection, service or repair. In this instance, the nature of contamination must be clearly communicated to the receiving organisation using the Trust Declaration of Decontamination Status of Healthcare Equipment Following Patient Use and Prior to Service or Repair form. 6.3 Items that have been involved in an incident In particular situations, for example when an item of equipment has been involved in an incident, its condition may be altered or influenced by the decontamination process. In such situations, advice must be sought from those investigating the incident, the Infection Prevention and Control Team and Medical Electronics prior to undertaking any decontamination processes. 6.4 Transporting items of equipment internally and externally to Trust Any packaging must be sufficiently robust to withstand transport and if possible packaging specifically designed for the item of equipment must be used in accordance with the Carriage of Dangerous Goods Regulations 2012. The condition of the item must be clearly labelled indicating content and contamination status. This is so that it can be clearly determined prior to opening the package. E.g. biohazard label if required and Trust Declaration of Decontamination Status of Healthcare Equipment Following Patient Use and Prior to Service or Repair form. Page 4 of 8

Transport of contaminated equipment within the Trust must be in a suitable container via internal hospital transport. Where appropriate all external parts of large items of mobile equipment should be covered in orange clinical waste bags and suitably labelled. It is illegal to send contaminated items by any form of postal system e.g. Royal Mail, or courier. It must be transported via an approved contractor who is licensed to transport such goods. It is the responsibility of the ward/department manager to ensure a safe system of packaging and transport is in place for all equipment moved within and external to the Trust. Contact Health and Safety for advice. 6.5 Management of Macerators When a macerator fails, it is the responsibility of the nursing staff in that area to report the fault immediately to the Estates Department or Interserve FM via Helpdesk 21000. Nursing staff responsibilities:- Label the macerator as out of commission as soon as the fault is identified On consultation with Estates/Interserve FM staff, remove any whole bed pans stacked in the macerator, wearing PPE in accordance with Trust Standard Precautions Policy Staff must not attempt to use the macerator until the fault has been fixed Seek advice from IPC Nurses if required, regarding alternative disposal of pulp products during this time Estates/Interserve FM responsibilities:- Isolate water and electrical supply to prevent further use and inadvertent powering of the unit Remove any pulp/waste material at blade level wearing PPE in accordance with Trust Standard Precautions Policy. This task must be risk assessed and control measures adhered to Please follow Estates guidance as agreed with Trust Health & Safety Repair the macerator Source temporary replacement macerator if the problem cannot be rectified It is accepted nursing staff will be unable to effectively decontaminate the macerator prior to repair and therefore all personnel working on this equipment must assume it is contaminated. For this reason a Declaration of Decontamination Status of Healthcare Equipment following Patient use and Prior to Service or Repair for is not required. Page 5 of 8

7 Training All staff working on Trust premises, including Trust employed staff; agency and locum staff are responsible for accessing IPC Policies via the intranet in order to assist in maintaining safe practice. 8 Equality and Diversity The Trust is committed to ensuring that, as far as is reasonably practicable, the way we provide services to the public and the way we treat our staff reflects their individual needs and does not discriminate against individuals or groups on any grounds. This policy has been appropriately assessed. 9 Process for monitoring compliance Standard/process/ issue Correct completion of Declaration of Decontamination Status of Healthcare Equipment Following Patient Use and Prior to Service or Repair form Monitoring and audit Method By Committee Frequency Audit Medical Trust Annually Electronics Decontamination Department Group 10 Consultation and review Consultation of this policy was undertaken by members of IPCC and IPC Nurses. This policy will be reviewed every two years by IPCC or when significant changes make earlier review necessary. 11 Implementation of policy (including raising awareness) This policy is a revision of a previous Decontamination of Healthcare Equipment following Patient Use and Prior to Service or Repair Policy. Clinical Directors/Matrons/Sisters/Charge Nurses and Clinical Leads should ensure that staff are aware of this policy. This policy is available for staff to access via NuTH intranet. 12 References and associated documentation Great Britain, Department of Health (2006). Essential Steps to Safe, Clean Care: Reducing health care associate infection. London: DH, 2006 Great Britain, Department of Health (2008). Health and Social Care Act 2008: Code of Practice on the prevention and control of infections and related guidance (revised 2015). Health and Safety at Work Act (1974) Management of Health and Safety at Work Regulations 1999 Approved Code of Practice and guidance MHRA Managing Medical Devices (2015) Page 6 of 8

Microbiology Advisory Committee (MAC) to Department of Health. Sterilisation, Disinfection and Cleaning of Medical Equipment Guidance on Decontamination The Carriage of Dangerous Goods Regulations, 2012 90/679/eec Council Directive on the protection of workers from risks related to exposure to biological agents at work 13 Associated Documents Cleaning and Disinfection Procedure Control of Transmissible Spongiform Encephalopathies (TSEs), including Creutzfeldt-Jacob Disease (CJD), in hospital patients Decontamination of Patient Environment (including Terminal Cleaning) Standard Precautions Decontamination of Reusable Medical Devices Policy Author: EME Services Officer, Medical Electronics Matron Infection Prevention and Control Page 7 of 8

Appendix 1 Declaration of Decontamination Status of Healthcare Equipment Following Patient Use and Prior to Service or Repair (Form must be completed for all medical devices including: e.g. Patient Monitors, Infusion Devices, Anaesthetic Machines, Ventilators and patient equipment including: e.g. Armchair, Commode, Wheelchair, Low Level Bed, Mattress) DEVICE NAME & MODEL: ASSET ID: (If applicable) FAULT REPORT REF. No: 0025000 IF REPORTING A FAULT (This section only to be completed if reporting a fault) DESCRIPTION OF FAULT : (Report the last settings and what you consider to be the problem) Was this device involved in an adverse incident? YES / NO If YES, please supply the Datix Incident Report No. Declaration of Contamination Status (This section to be completed for all medical devices and patient equipment) This equipment has not been used in any invasive procedure or been in contact with blood, other body fluids, expired gases, or pathological samples. It has been cleaned in preparation for inspection, servicing, repair or transportation, or This equipment has been exposed internally or externally to hazardous materials as indicated below (*delete as appropriate). *BLOOD - BODY FLUIDS - RESPIRED GASES - PATHOLOGICAL SAMPLES - CHEMICAL - RADIOACTIVE CONTAMINATION OR OTHER SUBSTANCES HAZARDOUS TO HEALTH (Please state):.... Please give details of the decontamination methods used:... If the equipment could not be decontaminated please indicate why. (Such equipment must not be returned without prior agreement of the recipient).. I DECLARE THAT I HAVE TAKEN ALL REASONABLE STEPS TO ENSURE THE ACCURACY OF THE ABOVE INFORMATION, IN ACCORDANCE WITH THE TRUST POLICY ON DECONTAMINATION OF HEALTHCARE EQUIPMENT FOLLOWING PATIENT USE AND PRIOR TO SERVICE OR REPAIR. Authorised Signature:... Print Name:...... Department:....Position:....... Date:....Tel. No:..... NOTE: IT IS ILLEGAL TO SEND CONTAMINATED ITEMS BY POST Devices returned for service/repair without a completed Fault Reporting and Contamination Status Form will not be accepted by the receiving department. For assistance contact the Electronics & Medical Engineering Department, the Health and Safety Team or the Infection Control Team. White copy to be attached to device Yellow copy to be retained by department Page 1 of 8

The Newcastle upon Tyne Hospitals NHS Foundation Trust Equality Analysis Form A This form must be completed and attached to any procedural document when submitted to the appropriate committee for consideration and approval. PART 1 1. Assessment Date: 09.11.17 2. Name of policy / strategy / service: Decontamination of Healthcare Equipment following Patient Use and Prior to Service or Repair 3. Name and designation of Author: Louise Hall Matron IPC and Dave Crawford EME services Officer, Medical Electronics 4. Names & designations of those involved in the impact analysis screening process: Louise Hall and Dave Crawford 5. Is this a: Policy X Strategy Service Is this: New Revised X Who is affected Employees X Service Users Wider Community X 6. What are the main aims, objectives of the policy, strategy, or service and the intended outcomes? (These can be cut and pasted from your policy) The aim of this policy is to ensure that all healthcare equipment is appropriately decontaminated to remove or minimise the risk of infection/other hazards prior to transport of equipment for inspection, servicing or repair, either on hospital premises or elsewhere. 7. Does this policy, strategy, or service have any equality implications? Yes No X If No, state reasons and the information used to make this decision, please refer to paragraph 2.3 of the Equality Analysis Guidance before providing reasons:

8. Summary of evidence related to protected characteristics Protected Characteristic Race / Ethnic origin (including gypsies and travellers) Evidence, i.e. What evidence do you have that the Trust is meeting the needs of people in various protected Groups Does evidence/engagement highlight areas of direct or indirect discrimination? If yes describe steps to be taken to address (by whom, completion date and review date) None relevant to this policy No No Sex (male/ female) None relevant to this policy No No Religion and Belief None relevant to this policy No No Sexual orientation including None relevant to this policy No No lesbian, gay and bisexual people Age None relevant to this policy No No Any reasonable No No adjustments that staff require are discussed with them. Disability learning difficulties, physical disability, sensory impairment and mental health. Consider the needs of carers in this section Gender Re-assignment None relevant to this policy No No Marriage and Civil Partnership None relevant to this policy No No Maternity / Pregnancy Staff who are pregnant or nursing mothers have a risk assessment of their duties. No No 9. Are there any gaps in the evidence outlined above? If yes how will these be rectified? No Does the evidence highlight any areas to advance opportunities or foster good relations. If yes what steps will be taken? (by whom, completion date and review date) 10. Engagement has taken place with people who have protected characteristics and will continue through the Equality Delivery System and the Equality Diversity and Human Rights Group. Please note you may require further engagement in respect of any significant changes to policies, new developments and or changes to service delivery. In such circumstances please contact the Equality and Diversity Lead or the Involvement and Equalities Officer. Do you require further engagement? Yes No X 11. Could the policy, strategy or service have a negative impact on human rights? (E.g. the right to respect for private and family life, the right to a fair hearing and the right to education?

No PART 2 Name: Louise Hall Date of completion: 09.11.17 (If any reader of this procedural document identifies a potential discriminatory impact that has not been identified, please refer to the Policy Author identified above, together with any suggestions for action required to avoid/reduce the impact.)