From: Risacher, Wendi Sent: Monday, April 13, :01 AM To: Hupp, Diane Subject: Corrected Alaris safety concerns. Hi Diane,

Similar documents
Clinical Skills Validation: Alaris Pump System

VAN WERT COUNTY HOSPITAL. Policy/Procedure: Interdepartmental No.: N Issue Date: 6-90 By: Nursing No. of Pages: 9

Improving the Safe Use of Multiple IV Infusions

Giving Intravenous (IV) Nutrition Through a Central Line with a CADD Pump

HomeMed Information. for the UMHS Cancer Center

Sharp HealthCare Safety Training 2015 Module 3, Lesson 2 Always Events: Line and Tube Reconciliation and Guardrails Use

Table of Contents. Nursing Skills. Page 2 of 8. Nursing School Made Simple Guaranteed 2014 SimpleNursing.com All Rights Reserved.

A AIRWAY Open the Airway B BREATHING Deliver two (2) Breaths. Code Blue Policy. Indications for Calling A Code Blue

Hospital-wide Lean Project:

To describe the process for the management of an infusion pump involved in an adverse event or close call.

Berinert Home Infusion Program. By: Loris Aro RN INNOMAR STRATEGIES

Please adjust your computer volume to a comfortable listening level. This is lesson 5 How to take medication properly.

INQUEST INTO THE DEATH OF: MARIE TANNER

UPMC PASSAVANT Policy Manual. TITLE/SUBJECT: IntraOsseous Device POLICY NO:

WYOMING STATE BOARD OF NURSING ADVISORY OPINION

INPATIENT Annual Core Competency Performance Stations (Nursing) 2010 (Unlicensed Staff Direct & Non-Direct Care Providers * )

Intravenous Epoprostenol (Flolan) Therapy

Plum 360 TM Infusion System with Full IV-EHR Interoperability

Bar Code Medication Administration and MAR Resource Manual

Staff Responsible Procedure Rationale/Reason

About the Critical Care Center

Information for Patients

STANDARDIZED PROCEDURE HEPATIC ARTERY INFUSION OF CHEMOTHERAPY (Adults, Peds)

WYOMING STATE BOARD OF NURSING ADVISORY OPINION INTRAVENOUS THERAPY BY LICENSED PRACTICAL NURSES

Home+ Home+ Home Infusion. Home Infusion. regionalhealth.org/home

Thopaz drain. Patient information. Thopaz Drain. Thoracic ward contact number: Reviewed: May 2015 Next review: May 2016 Version 1

The International Patient Safety Goals

This document is NOT FOR PROMOTIONAL USE. Do not copy, distribute, or share with physicians, staff, or patients. FOR INTERNAL USE ONLY.

What is a Mitrofanoff?

CARE OF THE PATIENT REQUIRING CONTINUOUS FLOLAN INFUSION GUIDELINE

Alaris Products. Protecting patients at the point of care

Considerations for Sterile Compounding of Parenteral Products for Pediatric Use: Part 2 PharMEDium Lunch and Learn Series LUNCH AND LEARN

Patient Safety Case Study. Clara K. Terral. Angelo State University

PROCESS FOR HANDLING ELASTOMERIC PAIN RELIEF BALLS (ON-Q PAINBUSTER AND OTHERS)

PATIENT CARE MANUAL PROCEDURE

Placement and Care of Your Gastrojejunostomy Tube (GJ Tube) Interventional Radiology

Hospira Sapphire PCA/Epidural Pump Handout

1. Infection Control, Centers for Disease Control and Prevention (CDC). (2002). Guideline for hand hygiene in health-care settings

ABSTRACT. dose", all steps in the setup of the secondary infusion must be conducted correctly.

DIALYSIS SAFETY. Dialysis Safety: What Patients Need To Know

Indications for Calling A Code Blue or Pediatric Medical Emergency

Patient Safety and Quality Measures for CRRT: The UAB Experience. Ashita Tolwani, M.D. University of Alabama at Birmingham CRRT 2012

Family/Caregiver Education Checklist Return Demonstration of Knowledge FIRST 24 HOURS

If viewing a printed copy of this policy, please note it could be expired. Got to to view current policies.

Professional Practice and Patient Safety Council

Just Culture Toolkit Scenarios

201 KAR 20:490. Licensed practical nurse intravenous therapy scope of practice.

STANDARDIZED PROCEDURE ALLOGENEIC /AUTOLOGOUS HEMATOPOIETIC STEM CELL INFUSION (Adult, Peds)

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

Electronic Medication Administration Process and Tips

CENTRAL IOWA HEALTHCARE Marshalltown, Iowa

McKinley T34 Ambulatory syringe pump Used in the provision of adult palliative and end of life care

ER ORIENTATION OUTLINE DAY 1

9/15/2017. Nursing Management Congress 2017 Interruptions in Clinical Practice. Interruptions in Clinical Practice. Review of the Literature

Medication Safety Way Beyond the 5 Rights

All about Your Implanted Venous Access Device (IVAD, Port )

End-to-end infusion safety. Safely manage infusions from order to administration

WHAT are medication errors?

Safety Innovations FOUNDATIONHTSI. Nine Recommendations To Prevent Multiple Line Infusion Medication Errors

ANTIBIOTIC ADMINISTRATION & MEDICATION ERROR AND REPORTING 12 th APRIL 2010

2016 SUMMER STUDENT NURSE EXTERNSHIP PROGRAM SKILLS CHECK LIST

DEPARTMENT OF HEALTH AND HUMAN SERVICES CENTERS FOR MEDICARE & MEDICAID SERVICES Medication Administration Observation

Alaris Guardrails Quick Overview for Staff Pharmacists

Anatomy of a Fatal Medication Error

Infusion Pumps: The Delivery Mechanism For Your Compounded Products PharMEDium Lunch and Learn Series

Policy Checklist. Working Group: Administration of Infusion of Intravenous Fluids & Medicines in Neonates (Chairperson: Dr Hogan) YES

Bedside Shift Reporting

Case Study from Parallon

Think proactively = prevent codes Elective intubation better than PEA arrest

St. Vincent s East Page 1 of 5

ADOPTED REGULATION OF THE STATE BOARD OF NURSING. LCB File No. R Effective April 4, 2016

Using Clinical Data Categories with the Pyxis MedStation

TUBE FEEDING WITH NUTRICIA CHOICE

Nasojejunal feeding tube

Comprehensive Aspiration Risk Management Plan (CARMP) Individual s Name: Case Manager: Date of CARMP: DOB:

What s in a Kit? A Supply Show and Tell to Facilitate Billing Accuracy. Speaker Disclosures. HCPCS History 3/9/2016

10/4/2012. Disclosure. Leading a Meaningful Event Investigation. Just Culture definition. Objectives. What we all have in common

NORTH SHORE MEDICAL CENTER NURSING PROCEDURE

SPONGE ACCOUNTing SYSTEM AUDIT TOOL

PURPOSE To establish a standardized process for the activity of an independent double check for medication administration.

Policy for use of the Royal Marsden Manual of Clinical Nursing Procedures (9th Edition)

Medication Management at Acme Medical Center

STANDARDIZED PROCEDURE LUMBAR DRAIN INSERTION (Adults, Peds)

5. returning the medication container to proper secured storage; and

Testing the Effectiveness of a New Device to Prevent Medical Line Entanglement in Pediatric Patients

Martin Health System Stuart, Florida Laboratory Services. Laboratory Services and Policies

Tube Feeding Status Critical Element Pathway

Pedi-CSI: Pediatric Clinical Safety Investigation Through Virtual Patient Safety Rounds

Medication Administration Using the Home Pump (Eclipse)

T34 Syringe Driver. Information for patients and carers. Palliative Care. Patient Information Leaflet

LEAN Transformation Storyboard 2015 to present

Equipment Cleaning Guidelines Template

GAMUT QI Collaborative Consensus Quality Metrics (v. 05/16/2016)

2000, SAFETY SOURCE PRODUCTIONS

patientinformation Peripherally inserted central catheter (PICC) Vascular Access The Rotherham NHS Foundation Trust

IV Interoperability: Smart Pump and BCMA Integration

Contributing to a culture of safety by increasing usage of the drug library on smart infusion pumps: A Quality Improvement Project

RIGHT HEMICOLECTOMY. Patient information Leaflet

Care of a Freka Percutaneous Endoscopic Gastrostomy (PEG)

SARASOTA MEMORIAL HOSPITAL NURSING DEPARTMENT POLICY

Transcription:

From: Risacher, Wendi Sent: Monday, April 13, 2015 8:01 AM To: Hupp, Diane Subject: Corrected Alaris safety concerns Hi Diane, I just again wanted to thank you for personally coming to address the Alaris pump issues and support the nursing staff. I am forwarding the list of concerns that the senior staff (Myself, Chris Madden. Karla Persia, Nicole Gargarella, Diane Hake, Brittany Tatro, Deanna Schneider, Mike Badach, Dawn Badach, Annette Fleck) compiled. I am not sure if you had ever seen the list so I just wanted to make sure that you had it. Most of the issues are product design and are not fixable. Your personal presence yesterday was appreciated by the staff. Thanks Wendi Risacher

Alaris Syringe Pump Concerns/Childrens PICU/CICU 1. Vital information is not displayed in one area. Need to navigate thru many screens to check dose, wgt. concentration. Too many steps to verify in an emergency. Old pumps had all info in one easy place with the syringe directly above in a horizontal manner so that you could compare the syringe to the pump. Even with the very easy to operate Medfusion pump, PSORS happened often with programming or placing wrong syringe on pump. 2. The most vital info (the dose) is not displayed constantly across the screen. Have to stand and watch it scroll to verify dose. Also have to bend over and read syringe vertically. The second row of pumps are practically on the floor and are difficult to see. 3. Unable to turn syringe in pump to read all details on label without syringes coming out and having to reprogram pump. Also pt gets bolused or decompensates with syringe manipulation. 4. There is no back button to the previous screen if you make an error. You must shut the channel down and start over. 5. Can't reprogram new concentration of infusion without shutting pump down. No option to just alter concentration. 6. Can not infuse fluid bolus on pump if programmed correctly for most patients. Get max volume error that can't be overridden. 7. The already addressed dose wt changing error potential. 8. Will allow you to choose the wrong syringe size. Accepted my choice of a 1cc when a 3cc was inserted. Had to shut whole channel off and start over instead of just changing it. 9. Med chambers have no independent battery. If something happens to main brain, all pressors will shut off. 10. Many more steps and buttons to push to do same tasks. We compared programming a dose of Vanco on both pumps and the Alaris took 14 steps/button pushes and 69 seconds vs the Medfusion 9 steps/button pushes and 39 seconds. we both did it at a normal pace, not racing. 11. Pumps constantly alarming disturbing patients and families. Very frustrating to nurses. Alarm silence time different for different things. Some only last 10 seconds for syringes/tubing changes. It is very distracting when you are changing life saving medication and we are focusing on stopping the alarms instead of our task at hand. Even when you put it on delay, when you take the syringe out it starts re-alarming. 12. You can not tell which pump is alarming. Nurses are telling me that they are very anxious when they have sick patients on pressors every time the pumps ring. 13. Medications take longer to infuse because it will only let you infuse the amount that is in the syringe after priming the tubing...not the whole dose. We already often run out of access when giving multiple medications. This leads to delays in getting meds into patient. Can interfere with peak and troughs. 14. Won't let you program total dose of med when it is greater than 60cc to run at your rate. Electrolytes come in a bag but must be infused via mini pump. We use a push pull system with a 60cc syringe but program 100cc to run over 2 hours. Old pump would allow this and just alarm when syringe was empty and we would refill syringe from the bag to complete the infusion.

15. I have witnessed many nurses just pushing the remaining volume of med in their tubing instead of infusing the flush because the pumps are so difficult. 16. On intermittent meds in 3,5,10cc syringes, you have to program both the dose and volume. You can't program the pump without the syringe in...but with the syringe in, you cant see both the dose and volume so you have to turn it or take it out...both of which make you start over from the beginning. If you type the wrong info in by accident, you have to shut it off and start all over again. With our other pumps we just programmed the med concentration, which was programmed in pump, and the dose. The volume was already figured out by the pump. 17. The physicality of the pumps are also unacceptable for both nursing and families. When they are at the head of the bed, we can not not reach our emergency resuscitation equipment (mapelson bag, suction, trachs, face masks.). We also need many more poles to use the same amt of pumps that we used to have one pole for. When the patients lines are on the L side of the bed, the parents can not get to their child or even see their child from the couch. The parents then have to come on the R side of the bed to see their child which is interfering with our ability to provide care leading to both parental and nursing frustration. 18. Transporting patients now takes extra hands to push multiple poles, and maneuvering through doorways/hallways is difficult. Pumps stick out and get knocked off of bed and walls leading to potential chamber disruption and purging of medication. They are also unstable risking accidental line removal. 19. Pumps stick out too far and tubing on far end of pump has very little slack leading to increased risk of accidental line removal. Parents often accidentally bump pumps and are afraid of harming their child. I had a parent come out of her room terrified because she tripped and bumped the pumps. 20. Old pumps showed volume infused on screen to know when specific volume reached the patient (Cris Madden brought this up and can speak to this issue better than I). 21. Nurse had to add second chamber to her pump and her Epi pump became dislodged on unstable patient. 22. Can't remove inner chambers if pt not using them when outer chambers are infusing meds. When other nurses need pumps we have many unused that we can not give them without interrupting our own patient's meds to remove chambers. 23. Old mini pumps were easy to transport by placing on patients bed. Alaris unable to do this. 24. Too bulky for CT and MRI. Tubing on outer chamber too short. 25. After checking main brain for med dose/concentration/wt we have to hit start or the pump will alarm making it seem as if the pump stopped working. Why do we have to hit start if we didn't stop it? Just more extra steps and nuisance alarms. 26. Plunger is sensitive and it is hard to place it without bolusing the med, or placing it on too loose so that the lag time is greater. 27. Takes way too long to change syringe on critical infusions. Several nurses have stated that they felt that their patient would have random abnormal blood pressure swings even when they did nothing to the pump. Even if we "double run" our pressors, we do not do this process with the flush or other meds in the same line so now those changes will affect the pt as well.

28. Nurses are programming their pumps in the basic infusion mode (which bypasses all of the safety features of the pump) because the program is too difficult/time consuming/frustrating to use. 29. Free flow risk is much greater with a vertical syringe style vs horizontal. Problems that should be easy to fix 1. Can near empty alarm be dc'd? 2. No option to piggyback meds when programmed in TPN mode. 3. Many meds not found in library. 4. Reference cards on pumps get in the way of the chambers attaching, and are in infections hazard. I stayed after work to do this so sorry it there are any grammar/typing errors. Wendi