Networked Medical Devices And The IEC80001 Standard: Are You Ready?
|
|
- Ella McKenzie
- 6 years ago
- Views:
Transcription
1 Networked Medical Devices And The IEC80001 Standard: Are You Ready? The Third Annual Medical Device Connectivity Conference & Exhibition Friday, September 9, 2011 Rick Hampton
2 Why are we here? In the past, networked medical devices required their own separate network. The HITECH Act and other business drivers are pushing healthcare institutions to converge the medical device network and the HIT network. Does converging medical devices safely require more than simply connecting the medical devices to the HIT network? What methodology do you use to address patient safety, continued efficacy of the medical device, and system security? Are there potential regulatory and legal responsibilities? How will your facility or company deal with the HIT network becoming a critical part of a regulated medical device?
3 The Joint Commission Sentinel Event Alert 42 ( ) Title Safely implementing health information and converging technologies As health information technology (HIT) and converging technologies the interrelationship between medical devices and HIT are increasingly adopted by health care organizations, users must be mindful of the safety risks and preventable adverse events that these implementations can create or perpetuate. Technology-related adverse events can be associated with all components of a comprehensive technology system and may involve errors of either commission or omission. This Alert focuses on how to safely implement HIT and converging health technologies.
4 The Joint Commission Sentinel Event Alert 42 Leadership standards LD.4.20 (LD *) and LD.4.40 (LD *) address designing new processes and establishing a safety program. In addition, since technology is prevalent in health care from patient admission to the surgical suite to the ordering and administration of medication and the use of equipment and medical devices any Joint Commission standard could potentially be tied to technology. Users should consider the use of any technology in relation to the standards and be aware of potential dangers to patients, as in any clinical situation. Environment of Care standards EC , EC , and EC address product recalls, selection, testing, and maintenance of medical equipment, and contingencies when equipment fails.
5 ECRI Institute Top 10 Technology Hazards For 2011 Alarm Hazards ranked #2 Alarm-related adverse incidents typically involve, alarms not being properly relayed to ancillary notification systems (e.g., paging system, wireless phones), potentially leading to a failure to notify relevant staff. Data Loss, System Incompatibilities, and Other Health IT Complications ranked #5 The convergence of medical technology and health information technology (HIT) is no longer a new phenomenon or purely a specialized concern. Technologies like medication management systems and processes like the incorporation of medical data from devices such as physiologic monitors and ventilators into electronic health records are just a few examples of how convergence is becoming more commonplace.
6 ECRI Institute Top 10 Technology Hazards For 2011 Data Loss, System Incompatibilities, and Other Health IT Complications ranked #5 While convergence presents many benefits, including improved standards of care and operational efficiencies, it also presents many risks. Ineffective convergence can adversely affect patient care in a wide variety of ways. It can, for example, lead to data being lost (e.g., overwritten, unsuccessfully transmitted) or being associated with the wrong patient, which in turn can lead to misdiagnosis, inappropriate treatment, or the need for repeat testing. To prevent hazards associated with convergence of medical technology and HIT, ECRI Institute recommends, [employing] good project management, change management, and risk management processes. To aid in this effort, hospitals should consider applying the International Electrotechnical Commission s new IEC standard, Application of Risk Management for IT- Networks Incorporating Medical Devices Part 1: Roles, Responsibilities and Activities.3
7 FDA NEWS RELEASE: February 14, 2011 Medical Device Data Systems are off-the-shelf or custom hardware or software products used alone or in combination that display unaltered medical device data, or transfer, store or convert medical device data for future use, in accordance with a preset specification. Examples of MDDS products include: devices that collect and store data from a glucose meter for future use or that transfer lab results to be displayed at a nursing station for future use. Information technology companies that design, install or market these systems, and hospitals that develop them in their facilities, must follow Class I requirements as well.
8 What is IEC 80001? IEC Ed.1: Application of risk management for it-networks incorporating medical devices Part 1: Roles, responsibilities and activities It is an international standard It is a voluntary standard unless formally adopted by regulatory authorities Specifically aimed at healthcare facilities of all sizes It is a process standard describing the risk management process Does not provide a checklist, but informs users how to create a program suitable for their facility
9 Future Documents Technical Reports (guidance documents) to be published soon: Guidance for Healthcare Delivery Organizations An implementation guide for HDOs of all sizes Step by Step Risk Management Still not a bullet-list of steps to accomplish, but rather a simple explanation of concepts from ISO and how they might be adopted by Responsible Organizations Security Use of Wireless Technologies With more to come
10 Origins of Core ideas drawn from: ISO Application of risk management to medical devices Applies ONLY to medical device manufacturers ISO IT Service Management Standards ITIL IT Infrastructure Library
11 The Key Properties defines three Key Properties to be managed. They are, in this order of priority: Safety (freedom from unacceptable risk of physical injury or damage to the health of people or damage to property or the environment) Effectiveness (ability to produce the intended result for the patient and the Responsible Organization) Data and System Security (an operational state of a Medical IT-Network in which information assets (data and systems) are reasonably protected from degradation of confidentiality, integrity, and availability) does not specify acceptable risk levels.
12 Role Relationships Medical Device Manufacturer Providers of Other Information Technology Responsible Organization Top Management Medical IT-Network Risk Manager
13 Contact Information Rick Hampton Wireless Communications Manager Partners HealthCare System One Constitution Center, Suite 200 Charlestown, MA Office: Cell:
14 Other References ISO/IEC : 2005 Medical Electrical Equipment requires manufactures to include some information in accompanying documents if medical equipment is to be connected to an IT network ISO/IEC 14971:2007 Application of risk management to medical devices ISO/IEC : 2010 Application of risk management for IT-networks incorporating medical devices Part 1: Roles, responsibilities & activities ISO/IEC :2005 IT Service Management System Information Technology Infrastructure Library (ITIL v3) HIMSS/NEMA HN Manufacturer s Disclosure Statement for Medical Device Security (MDS2) MIL-STD-882E DOD s Standard Practice for System Safety, ACCE ECRI Security Guide for Biomedical Technology, Systems Engineering Guide for Systems of Systems, Version 1.0, Office of the Deputy Under Secretary of Defense for Acquisition and Technology, Systems and Software Engineering. Washington, DC: ODUSD(A&T)SSE, DOD, Aug 2008 National Institute of Standards and Technology (NIST) standards (
15 Final Thoughts In flying I have learned that carelessness and overconfidence are usually far more dangerous than deliberately accepted risks. Wilbur Wright in a letter to his father, September 1900
QA offers significant economic benefits!
and Safety Systems in the USA J. Tobey Clark, MSEE, CCE, SASHE University of Vermont, USA Definitions Quality assurance Planned and systematic actions that can be demonstrated to provide confidence that
More informationFor some years, the automation of hospital administrative
An Introduction to IEC 80001: Aiming for Patient Safety in the Networked Healthcare Environment Sherman Eagles Editor s note: At press time, the second draft of IEC/CD2 80001, Application of risk management
More informationSTATEMENT. JEFFREY SHUREN, M.D., J.D. Director, Center for Devices and Radiological Health Food and Drug Administration
STATEMENT JEFFREY SHUREN, M.D., J.D. Director, Center for Devices and Radiological Health Food and Drug Administration Institute of Medicine Committee on Patient Safety and Health Information Technology
More informationU-M Hospitals and Health Centers Policies and Procedures
U-M Hospitals and Health Centers Policies and Procedures UMHHC Policy 05-02-006 Safe Medical Device Act Policy Issued: 4/00; Last Reviewed: 10/04; Last Revised: 10/04 Return to UMHHC Policies Table of
More informationAdministrative Policies and Procedures. Policy No.: N/A Title: Medical Equipment Management Plan
Administrative Policies and Procedures Originating Venue: Environment of Care Title: Medical Equipment Management Plan Cross Reference: Date Issued: 11/14 Date Reviewed: Date: Revised: Attachment: Page
More informationTop Ten Health Technology Hazards
Top Ten Health Technology Hazards MASHMM July 19, 2013 James P. Keller, M.S. Vice President, Health Technology Evaluation and Safety jkeller@ecri.org (610) 825-6000, ext. 5279 Presentation Overview ECRI
More informationSoftware Regulation and Validation
Software Regulation and Validation Keiichiro Ozawa FUJIFILM Corporation 12/07/2016 4th Joint Conference of Taiwan and Japan on Medical Products Regulation 1 Agenda 0. Introduction 1. Qualification and
More informationThe anesthesiologist switches the patient from the ventilator to the cardiopulmonary
Technology for surgery Integrating devices for patient safety The anesthesiologist switches the patient from the ventilator to the cardiopulmonary bypass machine but forgets to resume ventilation after
More informationChemical Biological Defense Materiel Reliability Program
Army Regulation 702 16 Product Assurance Chemical Biological Defense Materiel Reliability Program Headquarters Department of the Army Washington, DC 2 May 2016 UNCLASSIFIED SUMMARY of CHANGE AR 702 16
More informationMedical Device and Health Software
Medical Device and Health Software Standards and regulations now and in the future Justin McCarthy Introduction I was head of Clinical Engineering in Cardiff till 2009. I have been chair of BSI committee
More informationSentinel Events and S Patient Patient entinel Event Alerts Safety Act Safety Ac Revised: BW/September 2010
Sentinel Events Sentinel Events and Sentinel Event Alerts Revised: BW/September 2010 Patient Patient Safety Safety Act Act What is a Sentinel Event? 0 A sentinel event is an unexpected occurrence involving
More informationSafety Surveillance for Medical Devices
Safety Surveillance for Medical Devices May 10, 2012 James P. Keller, M.S. Vice President, Health Technology Evaluation and Safety jkeller@ecri.org (610) 825-6000, ext. 5279 Presentation Overview ECRI
More informationMidwest Alliance for Patient Safety Patient Safety Organization Getting Started with a PSO. An Illinois Hospital Association Company
Midwest Alliance for Patient Safety Patient Safety Organization Getting Started with a PSO An Illinois Hospital Association Company Today s Roadmap Objectives: 1. Explain the PSQIA and PSO Basics 2. Learn
More informationChoosing Regulatory-Compliant Power Strips for Every Area of a Healthcare Facility
www.tripplite.com Choosing Regulatory-Compliant Power Strips for Every Area of a Healthcare Facility WHITE PAPER 95-2883 Executive Summary The tangled web of regulatory compliance governing healthcare
More informationE.H.R. s and Improving Patient Safety - What Has Been the Real Impact?
E.H.R. s and Improving Patient Safety - What Has Been the Real Impact? Presented by: Mary Erickson, RN, HTS Accounting Manager HTS, a division of Mountain Pacific Quality Health Foundation 1 Understand
More informationEMERGING TRENDS WHAT I WILL COVER INCREASED INTEREST DEVICES ARE MIGRATING SAFE AND RELIABLE DEVICES LEAD TO LIVING WELL
SAFE AND RELIABLE DEVICES LEAD TO LIVING WELL THE HOME AND MEDICAL DEVICE TECHNOLOGY Presented by Mary Weick-Brady, MSN, RN April 20, 2012 Great Lakes Biomedical Conference on Home and Mobile Health WHAT
More informationCIO Legislative Brief
CIO Legislative Brief Comparison of Health IT Provisions in the Committee Print of the 21 st Century Cures Act (dated November 25, 2016), H.R. 6 (21 st Century Cures Act) and S. 2511 (Improving Health
More informationComponent Description Unit Topics 1. Introduction to Healthcare and Public Health in the U.S. 2. The Culture of Healthcare
Component Description (Each certification track is tailored for the exam and will only include certain components and units and you can find these on your suggested schedules) 1. Introduction to Healthcare
More informationGuide to Incident Reporting for In-vitro Diagnostic Medical Devices
Guide to Incident Reporting for In-vitro Diagnostic Medical Devices SUR-G0004-4 02 AUGUST 2012 This guide does not purport to be an interpretation of law and/or regulations and is for guidance purposes
More informationEligibility Introduction Practice Ethics and Patient Rights and Responsibilities (RI)... 6
Table of Contents Eligibility... 2 Introduction... 3 Practice Ethics and Patient Rights and Responsibilities (RI)... 6 Provision of Care, Treatment, and Services (PC)... 8 Medication Management (MM)...
More informationQC Explained Quality Control for Point of Care Testing
QC Explained 1.0 - Quality Control for Point of Care Testing Kee, Sarah., Adams, Lynsey., Whyte, Carla J., McVicker, Louise. Background Point of care testing (POCT) refers to testing that is performed
More informationGHTF SG2 Guidance: Group work output presentation
GHTF SG2 Guidance: Group work output presentation SG2 Post-market Surveillance & Vigilance SG2 is charged with the task of developing harmonized manufacturers adverse event reporting and other forms of
More informationWhich QMS Standard should be chosen for the structural quality of a medical laboratory? Matthias ORTH
Which QMS Standard should be chosen for the structural quality of a medical laboratory? Matthias ORTH IFCC Committee on Clinical Laboratory Management - http://www.ifcc.org/ifcc-education-division/emd-committees/c-clm/
More informationP2 Policies and Procedures for Institutions Working with PSOs
Working With Patient Safety Organizations (PSOs) Ronni P. Solomon ECRI Institute P2 Policies and Procedures for Institutions Working with PSOs Ronni P. Solomon, Executive Vice President and General Counsel,
More informationManager. 2. To establish procedures for selecting and acquiring biomedical equipment.
Page 1 of 8 CENTRAL STATE HOSPITAL POLICY SUBJECT: BIOMEDICAL EQUIPMENT MANAGEMENT ANNUAL REVIEW MONTH: RESPONSIBLE FOR REVIEW: October Regional Safety & Environmental Health Manager LAST REVISION DATE:
More informationUNCLASSIFIED. R-1 ITEM NOMENCLATURE PE A: Biometrics Enabled Intelligence FY 2012 OCO
Exhibit R-2, RDT&E Budget Item Justification: PB 2012 Army DATE: February 2011 COST ($ in Millions) FY 2010 FY 2011 FY 2013 FY 2014 FY 2015 FY 2016 To Program Element - 14.114 15.018-15.018 15.357 15.125
More informationTexas Tech University Health Sciences Center El Paso
Texas Tech University Health Sciences Center El Paso Medical Equipment Management Plan Medical Equipment Management Plan Contents I. Objective and Purpose II. III. IV. Selection and Acquisition Equipment
More informationPOCKET GUIDE TO THE ACCREDITATION STANDARDS (ISO 15189:2014)
(GLENMARIE BRANCH) POCKET GUIDE TO THE ACCREDITATION STANDARDS (ISO 15189:2014) Know the requirement!! Prepared by: Dr.Lily Manorammah Contents INTRODUCTION:... 3 OUR STRATEGY... 3 MANAGEMENT REQUIREMENTS...
More informationPRIVACY BREACH MANAGEMENT GUIDELINES. Ministry of Justice Access and Privacy Branch
Ministry of Justice Access and Privacy Branch December 2015 Table of Contents December 2015 What is a privacy breach? 3 Preventing privacy breaches 3 Responding to privacy breaches 4 Step 1 Contain the
More informationSKAC240 Manage health, safety, security and welfare in sport and active leisure
Manage health, safety, security and welfare in sport and active leisure Overview Health, safety, security and welfare are of vital importance in sport and active leisure. Incidents may occur that will
More informationLessons from Chicago
Lessons from Chicago Lela Holden, PhD, RN Patient Safety Officer Edward P. Lawrence Center for Quality & Safety Massachusetts General Hospital October 5, 2010 Let s hear from Catherine Zeta-Jones 2002
More informationFault Tree Analysis (FTA) Kim R. Fowler KSU ECE February 2013
Fault Tree Analysis (FTA) Kim R. Fowler KSU ECE February 2013 Purpose for FTA In the face of potential failures, determine if design must change to improve: Reliability Safety Operation Secondary purposes:
More informationPRIVACY IMPACT ASSESSMENT (PIA) For the
Aug 25, 2017 PRIVACY IMPACT ASSESSMENT (PIA) For the Business Continuity Planning System (BCPS) Defense Finance and Accounting Service SECTION 1: IS A PIA REQUIRED? a. Will this Department of Defense (DoD)
More informationWorking document QAS/ RESTRICTED September 2006
RESTRICTED September 2006 PREQUALIFICATION OF QUALITY CONTROL LABORATORIES Procedure for assessing the acceptability, in principle, of quality control laboratories for use by United Nations agencies The
More informationINSTRUCTION. Department of Defense. NUMBER August 19, 2009 Incorporating Change 1, October 10, 2017 USD(AT&L)
Department of Defense INSTRUCTION SUBJECT: Protecting Personnel from Electromagnetic Fields References: See Enclosure 1 NUMBER 6055.11 August 19, 2009 Incorporating Change 1, October 10, 2017 USD(AT&L)
More informationEconomic Burden of Counterfeit Medicine in Africa: Situation Analysis and Proposed Solution
Patient Access Economic Burden of Counterfeit Medicine in Africa: Situation Analysis and Proposed Solution Kasem S Akhras Head, Patient Access MENA ISPOR Annual European Meeting Africa Network Forum October
More informationSTANDARDS Point-of-Care Testing
STANDARDS Point-of-Care Testing For Surveys Starting After: January 1, 2018 Date Generated: January 12, 2017 Point-of-Care Testing Published by Accreditation Canada. All rights reserved. No part of this
More informationApril 17, Edition of the Joint Commission International Accreditation. SUBJECT: MITA Feedback on the 5 th Standards for Hospitals
1300 North 17 th Street Suite 1752 Arlington, Virginia 22209 Tel: 703.841.3200 Fax: 703.841.3392 www.medicalimaging.org April 17, 2013 Paul vanostenberg, DDS, MS Vice President Accreditation and Standards
More informationDEPARTMENT OF HEALTH AND HUMAN SERVICES. Permanent Certification Program for Health Information Technology; Revisions to
DEPARTMENT OF HEALTH AND HUMAN SERVICES Office of the Secretary 45 CFR Part 170 RIN 0991-AB77 Permanent Certification Program for Health Information Technology; Revisions to ONC-Approved Accreditor Processes
More informationClinical Medical Engineering New Paradigms New Challenges & AUB Experience
Bassam Tabshouri Clinical Medical Engineering New Paradigms New Challenges & AUB Experience The 3rd International Conference ON Biomedical & Clinical Engineering Outline 1. HealthCare Technology 2. New
More informationGuidance for Industry and FDA Staff Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use
Guidance for Industry and FDA Staff Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use Document issued on December 24, 2008 For questions regarding this document contact CDR
More informationManagement Standards. EHS Policy and Program
Global Environment, Health and Safety Standards Abbott has an established Environment, Health and Safety (EHS) Policy and set of management and technical standards that form the basis of our EHS management
More information510(k) Clinical Data Requirements: Current Status and Considerations for Clinical Studies
510(k) Clinical Data Requirements: Current Status and Considerations for Clinical Studies Sandra Maddock, RN, BSN, CCRA IMARC Research, Inc. Introduction In an effort to promote innovation while protecting
More informationThe Solution to Medical Device Security Also Could Save Tens of Thousands of Lives and Millions of Dollars
The Solution to Medical Device Security Also Could Save Tens of Thousands of Lives and Millions of Dollars February 24, 2017 Evolver, Inc. The Solution to Medical Device Security Could Save Tens of Thousands
More informationPersonal Protective Equipment Program. Risk Management Services
Personal Protective Equipment Program Services Table of Contents I. Program Goals and Objectives... 2 II. Scope and Application... 2 III. Responsibilities... 2 IV. Procedures... 3 V. Training... 5 VI.
More information12.01 Safety Management Plan UWHC Administrative Policies
Page 1 of 7 12.01 Safety Management Plan Category: UWHC Administrative Policy Policy Number: 12.01 Effective Date: October 8, 2013 Version: Revision Section: Environmental Safety (Hospital Administrative)
More informationTeaching the Smart Grid
Teaching the Smart Grid Why Data Management is Essential to the Future of Electricity Alexandra Nilles July 31, 2013 Alexandra Nilles (WISE) Teaching the Smart Grid July 31, 2013 1 / 17 Outline What is
More informationSJMHS SAFE MEDICAL DEVICE ACT (SMDA) 7/2017
SJMHS SAFE MEDICAL DEVICE ACT (SMDA) 7/2017 2 Objectives Upon completion of this course you should be able to: Describe the steps to take to report an adverse event with a medical device at your site Locate
More informationUnderstanding USP 797
Baxa Corporation Understanding USP 797 Technical Paper An Overview of USP General Chapter Pharmaceutical Compounding Sterile Preparations Mike Hurst, RPh, MBA 2004 Baxa Corporation Introduction USP
More information805A-36A-8005 Implement e-commerce Activities Status: Approved
Report Date: 02 May 2017 805A-36A-8005 Implement e-commerce Activities Status: Approved Distribution Restriction: Approved for public release; distribution is unlimited. Destruction Notice: Foreign Disclosure:
More information1. Workplace Violence Employee Survey 2010
1. Workplace Violence Employee Survey 2010 1. Do you feel safe at work? 2. Do you think you are prepared to handle a violent situation, threat, or responsive and escalating behaviours exhibited by clients
More informationGUIDELINES ON A MEDICAL DEVICES VIGILANCE SYSTEM
EUROPEAN COMMISSION DG ENTERPRISE AND INDUSTRY Directorate F-Consumer Good Unit F3- Cosmetic and Medical Devices MEDICAL DEVICES: Guidance document MEDDEV 2.12-1 rev 6 December 2009 GUIDELINES ON A MEDICAL
More informationDepartment of Defense INSTRUCTION. Non-Lethal Weapons (NLW) Human Effects Characterization
Department of Defense INSTRUCTION NUMBER 3200.19 May 17, 2012 Incorporating Change 1, September 13, 2017 USD(AT&L) SUBJECT: Non-Lethal Weapons (NLW) Human Effects Characterization References: See Enclosure
More informationContact Hours (CME version ONLY) Suggested Target Audience. all clinical and allied patient care staff. all clinical and allied patient care staff
1 Addressing Behaviors That Undermine a Culture of Safety PA CE CME FL 8/31/2016 2 2 7 3 43 1.0 1.0 1.0 all staff Sentinel Event Alert, Issue 40: Behaviors that undermine a culture of safety 2 Adverse
More informationAPEC Preliminary Workshop: Review of Drug Development in Clinical Trials
APEC Preliminary Workshop: Review of Drug Development in Clinical Trials Session 9 B Clinical Trial Assessment Patient Protection Informed Consent Susan D Amico Vice President and Global Head Clinical
More informationDOES TECHNOLOGY KEEP PATIENTS OUT OF HOSPITALS?
DOES TECHNOLOGY KEEP PATIENTS OUT OF HOSPITALS? Jeff Shuren, MD, JD Center for Devices and Radiological Health Food and Drug Administration November 6, 2014 1 Overview Challenges Medical Devices Used in
More informationReceived an RTA Deficiency List or AI Letter? Now What?
Received an RTA Deficiency List or AI Letter? Now What? Dealing with Unexpected Issues/Questions during the Submission Review Process Navigating Submission Challenges to reduce time & risk September 26,
More informationPatient Safety Course Descriptions
Adverse Events Antibiotic Resistance This course will teach you how to deal with adverse events at your facility. You will learn: What incidents are, and how to respond to them. What sentinel events are,
More informationPERFORMANCE WORK STATEMENT FOR. Sustainment/Patching Service U.S. AIR FORCE OWNED LAND MOBILE RADIO (LMR) EQUIPMENT. (ASTRO 25 Core System)
PERFORMANCE WORK STATEMENT FOR Sustainment/Patching Service U.S. AIR FORCE OWNED LAND MOBILE RADIO (LMR) EQUIPMENT (ASTRO 25 Core System) AT DYESS AFB, TEXAS PREPARED: July 2015 Nov 2017 TABLE OF CONTENTS
More informationSafety in the Pharmacy
Safety in the Pharmacy Course Practicum in Health Science - Pharmacology Unit I Preparation for Practicum Essential Question Why is safety in the pharmacy important not only to the patient, but the pharmacy
More informationHow the Industry Must Take in Stride New CMS and TJC Requirements
Two Steps Forward, One Step Back How the Industry Must Take in Stride New CMS and TJC Requirements Stephen L. Grimes About the Author Stephen L. Grimes, FACCE, FHIMSS, FAIMBE, is chief technology officer
More informationThis document describes the University s processes for reporting and investigating health and safety Incidents and Near Misses.
Health and Safety Guidelines: HSG 7.1 Incident Notification and Investigation 1. Purpose This document describes the University s processes for reporting and investigating health and safety Incidents and
More informationBIOMETRICS IN HEALTH CARE : A VALUE PROPOSITION FROM HEALTH CARE SECTOR
UMANICK TECHNOLOGIES, S.L. www.umanick.com info@umanick.com 1 / 7 Introduction In any country s health care system, many challenges have yet to be resolved. And patient identification is perhaps the greatest
More informationTo describe the process for the management of an infusion pump involved in an adverse event or close call.
TITLE INFUSION PUMPS FOR MEDICATION & PARENTERAL FLUID ADMINISTRATION SCOPE Provincial, Clinical DOCUMENT # PS-70-01 APPROVAL LEVEL Executive Leadership Team SPONSOR Provincial Medication Management Committee
More informationComprehensive Staff and Patient Communications: Emphasizing Patient-Centered Care
Nurse Call System Comprehensive Staff and Patient Communications: Emphasizing Patient-Centered Care Streamlining Communications T he needs of hospitalized patients are more complex than ever before, making
More informationNational Health Regulatory Authority Kingdom of Bahrain
National Health Regulatory Authority Kingdom of Bahrain THE NHRA GUIDANCE ON SERIOUS ADVERSE EVENT MANAGEMENT AND REPORTING THE PURPOSE OF THIS DOCUMENT IS TO OUTLINE SERIOUS ADVERSE EVENTS THAT SHOULD
More informationELECTRONICS TECHNICIAN I ELECTRONICS TECHNICIAN II
CITY OF ROSEVILLE ELECTRONICS TECHNICIAN I ELECTRONICS TECHNICIAN II DEFINITION To perform work in the installation, testing, maintenance, calibration, repair and modification of electrical and electronic
More informationMINIMUM CRITERIA FOR REACH AND CLP INSPECTIONS 1
FORUM FOR EXCHANGE OF INFORMATION ON ENFORCEMENT Adopted at the 9 th meeting of the Forum on 1-3 March 2011 MINIMUM CRITERIA FOR REACH AND CLP INSPECTIONS 1 MARCH 2011 1 First edition adopted at the 6
More informationTHE AMERICAN BOARD OF PATHOLOGY PATIENT SAFETY COURSE APPLICATION
THE AMERICAN BOARD OF PATHOLOGY PATIENT SAFETY COURSE APPLICATION Requirements: Component I Patient Safety Self-Assessment Program Programs must meet the following criteria to be an ABP approved Patient
More informationMay 8, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane, Room 1061 Rockville, MD
701 Pennsylvania Avenue, NW Suite 800 Washington, D.C. 20004 2654 Tel: 202 783 8700 Fax: 202 783 8750 www.advamed.org May 8, 2014 Division of Dockets Management (HFA-305) Food and Drug Administration 5630
More informationReview for Required Monitors
Review for Required Monitors The Joint Commission Hospital Accreditation Manual, 2009 Medicare Conditions of Participation, Hospitals Update: February 2009 Indicator / Monitor Restraint, Medical (non-specific
More informationPOLICY & PROCEDURE FOR INCIDENT REPORTING
POLICY & PROCEDURE FOR INCIDENT REPORTING APPROVED BY: South Gloucestershire Clinical Commissioning Group Quality and Governance Committee DATE February 2015 Date of Issue: 25 February 2015 Version No:
More informationWe are the regulator: Our job is to check whether hospitals, care homes and care services are meeting essential standards.
Inspection Report We are the regulator: Our job is to check whether hospitals, care homes and care services are meeting essential standards. Spire Wellesley Hospital Eastern Avenue, Southend-on-Sea, SS2
More informationFebruary New Zealand Health and Disability Services National Reportable Events Policy 2012
February 2012 New Zealand Health and Disability Services National Reportable Events Policy 2012 Table of Contents 1. Purpose 2. Treaty of Waitangi 3. Background 4. Scope 5. Policy 6. Review and Evaluation
More informationCenters for Medicare & Medicaid Pay for Performance Updates Jeff Flick Regional Administrator CMS, Region IX February 7, 2006
Centers for Medicare & Medicaid Pay for Performance Updates Jeff Flick Regional Administrator CMS, Region IX February 7, 2006 Slide -1 Big Changes in Medicare New orientation toward prevention Personalized
More informationGeneral Health and Safety Information for Victoria University. An outline of any further Health and Safety Training you may require.
Directed by: Health and Safety Officer Physical Plant Updated: March 2017 General Health and Safety Information for Victoria University. General Health and Safety Information for the task and work you
More informationONC Health IT Certification Program: Enhanced Oversight and Accountability
This document is scheduled to be published in the Federal Register on 10/19/2016 and available online at https://federalregister.gov/d/2016-24908, and on FDsys.gov DEPARTMENT OF HEALTH AND HUMAN SERVICES
More informationBurton Hospitals NHS Foundation Trust. On: 25 January Review Date: December Corporate / Directorate. Department Responsible for Review:
POLICY DOCUMENT Burton Hospitals NHS Foundation Trust MEDICAL DEVICES TRAINING POLICY Approved by: Trust Executive Committee On: 25 January 2017 Review Date: December 2019 Corporate / Directorate Clinical
More informationAvera Critical Lab Value Alarms. Candice Friestad, MSN, RN, MBA Director, Clinical Informatics
Avera Critical Lab Value Alarms Candice Friestad, MSN, RN, MBA Director, Clinical Informatics Avera Innovation Vision To integrate technology into the patient experience such that patient satisfaction,
More informationBed Rail Provision Guidance for Assessment and Provision. Clinical & Prescriber Support Special Interest Group
Bed Rail Provision Guidance for Assessment and Provision Clinical & Prescriber Support Special Interest Group Issue 1: June 2011 Index: Page: The NAEP Clinical & Prescriber Support Special Interest Group
More informationChanging Requirements for Devices//Device Constituent Parts in Combination Products
Changing Requirements for Devices//Device Constituent Parts in Combination Products Dan Wozinski, RPH, MBA Sanofi Disclaimer The content and viewpoints of this presentation are mine alone and not those
More informationSupporting The Joint Commission 2012 Standards and National Patient Safety Goals
Supporting The Joint Commission 01 Standards and National Patient Safety Goals for Pyxis technologies This document highlights select Joint Commission 01 Standards and National Patient Safety Goals mapped
More informationDATA PROTECTION POLICY
DATA PROTECTION POLICY Document Number 2010/35/V1 Document Title Data Protection Policy Author Nic McCullagh Author s Job Title Information Governance Manager Department IM&T Ratifying Committee Capacity
More informationPrepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
More informationUnderstanding Patient Choice Insights Patient Choice Insights Network
Quality health plans & benefits Healthier living Financial well-being Intelligent solutions Understanding Patient Choice Insights Patient Choice Insights Network SM www.aetna.com Helping consumers gain
More informationDOD MANUAL DOD ENVIRONMENTAL LABORATORY ACCREDITATION PROGRAM (ELAP)
DOD MANUAL 4715.25 DOD ENVIRONMENTAL LABORATORY ACCREDITATION PROGRAM (ELAP) Originating Component: Office of the Under Secretary of Defense for Acquisition, Technology, and Logistics Effective: April
More informationNursing Documentation 101
Nursing Documentation 101 Module 5: Applying Knowledge Part I Handout 2014 College of Licensed Practical Nurses of Alberta. All Rights Reserved. Nursing Documentation 101 Module 5: Applying Knowledge Part
More informationAcute Care Workflow Solutions
Acute Care Workflow Solutions 2016 North American General Acute Care Workflow Solutions Product Leadership Award The Philips IntelliVue Guardian solution provides general floor, medical-surgical units,
More informationAutomation and Information Technology
4 Automation and Information Technology Positions Automation and Information Technology Ensuring Patient Safety and Data Integrity During Cyber-attacks (1701) To advocate that healthcare organizations
More informationPOLICY NAME POLICY # Sentinel, Adverse Event and Near Miss. CSP Reporting and Investigation
Purpose To outline a reporting system that promotes client safety by learning from experiences and utilizing the results of investigations and data analysis to prepare and disseminate recommendations for
More informationRemote Patient Monitoring - How Are RM Systems Affecting Home and Hospital Ecosystems? Bill Betten Director of Business Solutions Devicix
Remote Patient Monitoring - How Are RM Systems Affecting Home and Hospital Ecosystems? Bill Betten Director of Business Solutions Devicix Key Take-Aways Explore what s new and emerging in remote patient
More informationChanges in practice and organisation surrounding blood transfusion in NHS trusts in England
See Commentary, p 236 1 National Blood Service, Birmingham, UK; 2 National Blood Service, Oxford, UK; 3 Clinical Evaluation and Effectiveness Unit, Royal College of Physicians, London, UK Correspondence
More informationPatient Safety is Everyone s Responsibility Tammy Brock, MSN RN CPHRM
Patient Safety is Everyone s Responsibility Tammy Brock, MSN RN CPHRM Objectives Know TJC 2016 National Patient Safety Goals Discuss human factors on patient safety What is your role in patient safety?
More informationReport No. D September 25, Controls Over Information Contained in BlackBerry Devices Used Within DoD
Report No. D-2009-111 September 25, 2009 Controls Over Information Contained in BlackBerry Devices Used Within DoD Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden for
More informationEnd-to-end infusion safety. Safely manage infusions from order to administration
End-to-end infusion safety Safely manage infusions from order to administration New demands and concerns 56% 7% of medication errors are IV-related. 1 of high-risk IVs are compounded in error. 2 $3.5B
More informationgeneral criteria New Zealand Code of Radiology Management Practice for accreditation
general criteria for accreditation New Zealand Code of Radiology Management Practice Radiology Services Particular requirements for quality and competence Developed from NZS/ISO 15189: 2007 general criteria
More informationSample Reportable Events
Sample Reportable Events This list serves as a guideline of event types typically reported through the ERS (Event Reporting System), online event reporting software. These examples come from hospitals
More informationThe Importance of Transfusion Error Surveillance This is step #1 in error management. Jeannie Callum, BA, MD, FRCPC, CTBS
The Importance of Transfusion Error Surveillance This is step #1 in error management Jeannie Callum, BA, MD, FRCPC, CTBS 6051 Clinical Errors 9083 Laboratory Errors 15134 Errors over 6 years I don t want
More informationRoles & Responsibilities of Investigator & IRB
Roles & Responsibilities of Investigator & IRB Jaranit Kaewkungwal Mahidol University Regulatory & Guidelines Regulatory & Guidelines GCP & Computer / Database Management Systems International Conference
More information(Type inside gray boxes, cells will expand) A. EIGHT POINT CRITERIA for IRB Review
Page 1 of 5 IRB Reviewers 8-Point Analysis Form Based on Federal Policy for the Protection of Human Subjects, Criteria for IRB Approval of Research (45 CFR 46.111) Protocol ID #/Title: Date of Review:
More information